  "results": [
    {
      "status": "Terminated",
      "city": "Boerne",
      "state": "TX",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88323",
      "recalling_firm": "Stanbio Laboratory, LP",
      "address_1": "1261 N Main St",
      "address_2": "",
      "postal_code": "78006-3014",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MI, MO, NC, NH, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, TN, TX, VA, VT, WA, WI, and WV. The countries of  Australia, Canada, Chile, Federal Republic of Germany, Hong Kong, Singapore, Taiwan, Thailand, United Kingdom, and Viet Nam.",
      "recall_number": "Z-2201-2021",
      "product_description": "Synchron CX/DX/LX Beta-Hydroxybutyrate .  in vitro diagnostic reagent.",
      "product_quantity": "100 units",
      "reason_for_recall": "Deterioration in the stability of the reagents which has resulted in lower than expected or out of range Quality Control values.",
      "recall_initiation_date": "20210712",
      "center_classification_date": "20210805",
      "termination_date": "20220329",
      "report_date": "20210811",
      "code_info": "Catalog Number/Product Description: B2440-180/Synchron CX/DX/LX ¿-Hydroxybutyrate LiquiColor Device Identifier:  00657498000526 Lot Numbers: 161020;",
      "more_code_info": ""
    }
  ]
}