  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88282",
      "recalling_firm": "Straumann USA LLC",
      "address_1": "60 Minuteman Rd",
      "address_2": "",
      "postal_code": "01810-1008",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-2198-2021",
      "product_description": "Straumann SCS Screwdriver, driver, for ratchet, short, L 21mm, stainless steel\t  Article Number: 046.401. dental surgical instrument.",
      "product_quantity": "130 pieces",
      "reason_for_recall": "Tip of the screwdriver does not have the correct configuration / profile therefore the screwdriver does not properly fit the opposing screw head and cannot pick up the screws, may result in injury",
      "recall_initiation_date": "20210527",
      "center_classification_date": "20210803",
      "termination_date": "20230321",
      "report_date": "20210811",
      "code_info": "Packaged lot number: EHG85 Lot Number engraved on part: CXV34",
      "more_code_info": ""
    }
  ]
}