      "status": "Ongoing",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94644",
      "recalling_firm": "Draeger Medical Systems, Inc.",
      "address_1": "6 Tech Dr",
      "address_2": "",
      "postal_code": "01810-2434",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution. International distribution to Afghanistan, Argentina, Armenia, Australia, Austria, Belarus, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-es-S, Bulgaria, Canada, Chile, China, Columbia, Costa Rica, Croatia, Czech Republic, Egypt, Estonia, Finland, France, French Guiana, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Liechtenstein, Lithuania, Malaysia, Malta, Martinique, Mexico, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Vietnam.",
      "recall_number": "Z-2189-2024",
      "product_description": "Infinity Acute Care System (IACS) Monitoring System, Infinity M500 with Power, Model No. MS20407. Rev 20, 21, and 23",
      "product_quantity": "34,102 units",
      "reason_for_recall": "Device is not in full compliance to Type CF requirements of IEC 60601-1 and IEC 60601-2-34 standards. Therefore, there is a potential risk of electric shock. Health consequences may include electrical shock, burns, and cardiac arrhythmia.",
      "recall_initiation_date": "20240513",
      "center_classification_date": "20240625",
      "report_date": "20240703",
      "code_info": "Model Number: MS20407. UDI-DI: 04049098054447. Rev 20, 21, and 23",
      "more_code_info": ""
    }
  ]
}