  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Irving",
      "address_1": "1915 Hurd Dr",
      "reason_for_recall": "There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results",
      "address_2": "",
      "product_quantity": "2,212 units",
      "code_info": "Lot Numbers / Expiration Date    84126UN20 / 09JUL2021  03676UN20 / 25AUG2021  10597UN20 / 29AUG2021  26275UN20 / 20OCT2021  77748UN20 / 08DEC2021  65958UN20 / 08DEC2021  81824UN21 / 16FEB2022",
      "center_classification_date": "20210802",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of VA, UT, TX, TN, SD,PA, NY, ME, ID, CA and the countries of AFGHANISTAN, AUSTRALIA, AUSTRIA, BAHAMAS, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVI, CANADA, CAYMAN ISLANDS, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FAROE ISLAND, FINLAND, FRANCE, GERMANY, GREENLAND, Hong Kong, INDIA, IRELAND, ISRAEL, ITALY, KUWAIT, LITHUANIA, LUXEMBOURG, MEXICO, MONTENEGRO, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PANAMA, PHILIPPINES, PORTUGAL, PR CHINA, REPUBLIC OF YEMEN, ROMANIA, RUSSIA, SAUDI ARABIA, SLOVAKIA, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, UNITED KINGDOM, VIETNAM.",
      "state": "TX",
      "product_description": "REF 8L92-42, Abbott Activated Alanine Aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740161866",
      "report_date": "20210811",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Abbott Laboratories",
      "recall_number": "Z-2187-2021",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "88123",
      "more_code_info": "",
      "recall_initiation_date": "20210610",
      "postal_code": "75038",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}