  "results": [
    {
      "status": "Ongoing",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97067",
      "recalling_firm": "CareFusion 303, Inc.",
      "address_1": "10020 Pacific Mesa Blvd",
      "address_2": "N/A",
      "postal_code": "92121-4386",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution to states of: AK\tAL\tAR\tAS\tAZ\tCA\tCO\tCT\tDC\tDE\tFL\tGA\tGU\tHI\tIA\tID\tIL\tIN\tKS\tKY\tLA\tMA\tMD\tME\tMI\tMN\tMO\tMP\tMS\tMT\tNC\tND\tNE\tNH\tNJ\tNM\tNV\tNY\tOH\tOK\tOR\tPA\tPR\tRI\tSC\tSD\tTN\tTX\tUT\tVA\tVT\tWA\tWI\tWV\tWY; and OUS (foreign) to countries of: Japan, Italy, Germany, Turkey",
      "recall_number": "Z-2179-2025",
      "product_description": "BD Alaris Pump Module Model, 8100  with Guardrails, Suite MX software versions",
      "product_quantity": "33,162,680 units total",
      "reason_for_recall": "Infusion pump module used with compatible  pump infusion sets  may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.",
      "recall_initiation_date": "20250708",
      "center_classification_date": "20250806",
      "report_date": "20250813",
      "code_info": "BD Alaris\" Pump Module Model 8100- All Serial Numbers/ UDI(s): 10885403222054, 10885403517723, 10885403810015, 10885403810039, 10885403810046",
      "more_code_info": ""
    }
  ]
}