  "results": [
    {
      "status": "Ongoing",
      "city": "Mounds View",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97130",
      "recalling_firm": "Medtronic, Inc.",
      "address_1": "8200 Coral Sea St Ne",
      "address_2": "N/A",
      "postal_code": "55112-4391",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Worldwide - US Nationwide distribution.",
      "recall_number": "Z-2168-2025",
      "product_description": "Medtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components:    1.\tCobalt Crome application, Software Model Number D00U005;    2.\tClaria Amplia Compia application, Software Model Number D00U009;    3.\tEvera MRI application, Software Model Number D00U010;    3.\tVisia AF application, Software Model Number D00U011;    5.\tViva Brava Evera application, Software Model Number D00U012;",
      "product_quantity": "22091 total devices",
      "reason_for_recall": "In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.",
      "recall_initiation_date": "20250618",
      "center_classification_date": "20250723",
      "report_date": "20250730",
      "code_info": "1. Software Model Number, D00U003, GTIN 00763000002039, Updated Software Version 6.5.5;  2. Software Model Number, D00U004, GTIN 00763000002046, Updated Software Version 6.5.5;  3. Software Model Number, D00U005, GTIN 00763000002053, Updated Software Version 9.5.2;  4. Software Model Number, D00U006, GTIN 00763000397852, Updated Software Version 4.4.6;  5. Software Model Number, D00U007, GTIN 00763000397869, Updated Software Version 4.4.6;  6. Software Model Number, D00U008, GTIN 00763000397876, Updated Software Version 3.5.5;  7. Software Model Number, D00U009, GTIN 00763000397883, Updated Software Version 3.4.2;  8. Software Model Number, D00U010, GTIN 00763000397890, Updated Software Version 3.4.2; 9. Software Model Number, D00U011, GTIN 00763000397906, Updated Software Version 3.4.2; 10. Software Model Number, D00U012, GTIN 00763000397913, Updated Software Version 3.4.2; 11. Software Model Number, D00U022, GTIN 00763000544300, Updated Software Version 3.7.5."
    }
  ]
}