  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Domat/Ems",
      "state": "",
      "country": "Switzerland",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97222",
      "recalling_firm": "Hamilton Medical AG",
      "address_1": "Parc Industrial Vial 10",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC, NJ, UT, TX, GA, SC, VA, IN, SD, MO, OR, TN, MA, AZ, NM.",
      "recall_number": "Z-2167-2025",
      "product_description": "Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.",
      "product_quantity": "2,560",
      "reason_for_recall": "Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.",
      "recall_initiation_date": "20250626",
      "center_classification_date": "20250730",
      "report_date": "20250806",
      "code_info": "UDI-DI: 07630002802963, Lots: 199675, 199676",
      "more_code_info": ""
    }
  ]
}