{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Morrisville",
      "state": "NC",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98745",
      "recalling_firm": "ARROW INTERNATIONAL, LLC",
      "address_1": "3015 Carrington Mill Blvd",
      "address_2": "",
      "postal_code": "27560-5437",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "Worldwide distribution: US nationwide and countries of:  AE, AL, AR, AT, AU, BE, BG, BS, BR, BW, CA, CH, CL, CN, CO, CZ, DE, EC, EE, ES, FR, GB, GR, GT, HU, ID, IE, IN, IT, JP, KW, KZ, LB, LT, LU, MN, MZ, NA, NL, NO, NP, NZ, PA, PE, PF, PH, PL, PT, QA, RO, RS, RU, SA, SG, SL, SK, SZ, TH, TN, TR, TT, UA, VN, ZA, ZM, and ZW",
      "recall_number": "Z-2159-2026",
      "product_description": "NextStep Antegrade Chronic Hemodialysis Catheter, REF: AC-15192-SFX, AC-15232-SFX, CS-15192-SFX, CS-15232-SFX, CS-15272-SFX, CS-15312-SFX, CS-15422-SFX, CS-15502-SFX;      NextStep Antegrade Hemodialysis Catheter with Arrow Simplicity Micro-Introducer, REF: CS-15192-SFXM, CS-15232-SFXM;    NextStep Retrograde Hemodialysis Catheter, REF: AC-15192-X, AC-15232-X, AC-15272-X, CS-15192-X, CS-15232-X, CS-15272-X, CS-15312-X;    NextStep Retrograde Hemodialysis Catheter with Arrow Simplicity Micro-Introducer, REF: CS-15192-XM, CS-15232-XM, CS-15272-XM, CS-15312-XM",
      "product_quantity": "19,687",
      "reason_for_recall": "16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.",
      "recall_initiation_date": "20260410",
      "center_classification_date": "20260513",
      "report_date": "20260520",
      "code_info": "REF/UDI-DI/Lots: AC-15192-SFX/30801902197758/AC-15192-SFX; AC-15232-SFX/00801902095909/33F23H1018;  CS-15192-SFX/30801902195297/33F23E0575, 33F23G0530, 33F23K0187, 33F23K1104, 33F23L0929, 33F24A0275, 33F24C0068, 33F24D1012, 33F25C0020, 33F25E1083, 33F25G0636, 33F25H0292, 33F25H0790, 33F25L0038;  CS-15232-SFX/30801902195549/33F23E0576, 33F23G0534, 33F23L0936, 33F24A0279, 33F24C0067, 33F24D1006, 33F24E0684, 33F24H0258, 33F24J0033, 33F24J0085, 33F24J0342, 33F24L0056, 33F24M0231, 33F25A0405, 33F25C0036, 33F25C0664, 33F25D0739, 33F25E0433, 33F25G0716, 33F25H0248; CS-15272-SFX/10801902195217/33F23B0321, 33F23D0859, 33F23H0500, 33F23K0177, 33F23L0943, 33F24A0280, 33F24C0083, 33F24H0042, 33F24H0260, 33F24J0022, 33F24J0038, 33F24J0088, 33F24L0177, 33F25C0028, 33F25E1090, 33F25F0301, 33F25F0514; CS-15312-SFX/30801902197529/33F23G0541, 33F23H0984, 33F23L0951, 33F24A0283; CS-15422-SFX/30801902195327/33F23K0179, 33F23L0956, 33F24A0290, 33F24E0814; CS-15502-SFX/30801902195563/33F23L0959, 33F24A0286, 33F24J0060; CS-15192-SFXM/10801902127157/33F25C0015, 33F25E1084; CS-15232-SFXM/10801902127232/33F25C0671;  AC-15192-X/ 00801902095879/33F23F0744, 33F24E0784, 33F25D0685; AC-15232-X/00801902095916/33F23H1013, 33F24J0068, 33F25C0675, 33F25J0589; AC-15272-X/00801902095954/33F23H1023;  CS-15192-X/30801902195303/33F23C0886, 33F23E0564, 33F23G0263, 33F23G0533, 33F23K0712, 33F23L0935, 33F24E0187, 33F24E0793, 33F24G0171, 33F24J0017, 33F24J0041, 33F25D0701, 33F25E0476, 33F25F0486, 33F25H0783, 33F25H0857; CS-15232-X/30801902195556/33F23C0862, 33F23E0566, 33F23F0722, 33F23K0171, 33F23L0940, 33F24A0916, 33F24C0061, 33F24D1037, 33F24E0800, 33F24J0043, 33F24L0176, 33F25A0028, 33F25C0029, 33F25C0644, 33F25D0096, 33F25E0352, 33F25F0300, 33F25F0513, 33F25E0477, 33F25J0745; CS-15272-X/10801902195224/33F23C0887, 33F23F0731, 33F23H0983, 33F23K0190, 33F23L0947, 33F24E0807, 33F24F0124, 33F24J0018, 33F24J0050, 33F24L0054, 33F24L0063, 33F24M0006, 33F25C0684, 33F25E0342, 33F25F0276, 33F25J0641; CS-15312-X/30801902195310/33F23C0863, 33F23D0858, 33F23G0544;  CS-15192-XM/10801902127225/33F25D0711; CS-15232-XM/10801902127300/33F25E0435; CS-15272-XM/10801902127409/33F25C0680, 33F25K0272; CS-15312-XM/10801902127508/33F25E0442",
      "more_code_info": ""
    }
  ]
}