  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northbrook",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97143",
      "recalling_firm": "Bridge to Life Ltd",
      "address_1": "707 Skokie Blvd Ste 340",
      "address_2": "N/A",
      "postal_code": "60062-2838",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "United States - New York, Maryland, Alabama, North Carolina, South Carolina, Ohio, Florida, Wisconsin, Utah, Tennessee, New Jersey, Georgia, Nebraska, Arizona, Minnesota, California, and Canada",
      "recall_number": "Z-2157-2025",
      "product_description": "Bridge to Life EasiSlush, (Sodium Chloride Solution for Sterile Slush Preparation) (0.9% Sodium Chloride Irrigation, USP),  Part Number REF BTLE-1250",
      "product_quantity": "4528 units",
      "reason_for_recall": "The firm issued a field safety notice after becoming aware of three lots of products not having quarterly dose audits being completed on time in the first quarter of 2024.  Routine lot by lot testing was completed properly however the quarterly dose audit (sampling/testing to monitor the gamma sterilization process and confirm the validated parameters remain effective to assure sterility and bioburden reduction per requirements) was not completed on time.",
      "recall_initiation_date": "20250612",
      "center_classification_date": "20250722",
      "report_date": "20250730",
      "code_info": "UDI 10860001991205-240206-260206-240372708 (case), 00860001991208-240206-260206-240372708 (bag), Lot Number 240372708;  UDI 10860001991205-240212-260212-240402836 (case), 00860001991208-240212-260212-240402836 (bag), Lot Number 240402836;  UDI 10860001991205-240214-260214-240402837 (case), 00860001991208-240214-260214-240402837 (bag), Lot Number 240402837."
    }
  ]
}