  "results": [
    {
      "status": "Terminated",
      "city": "Lugano",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85444",
      "recalling_firm": "Inpeco S.A.",
      "address_1": "Via San Gottardo 10",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide including in the states of IL, NY.",
      "recall_number": "Z-2141-2020",
      "product_description": "Accelerator a3600 Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Module (P/N FLX-226-01, FLX-226-10), or StaRRsed Interface Module (P/N FLX-268-00) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.",
      "product_quantity": "3 systems",
      "reason_for_recall": "The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of these Interface Modules is put off-line after a carrier in their secondary lane is physically returned on the main track and then put back on-line when the carrier is used to transport another tube.",
      "recall_initiation_date": "20200407",
      "center_classification_date": "20200528",
      "termination_date": "20240206",
      "report_date": "20200603",
      "code_info": "Serial Numbers: ACP.290  ACP.343  ACP.365"
    }
  ]
}