  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88285",
      "recalling_firm": "Maquet Medical Systems USA",
      "address_1": "45 Barbour Pond Dr",
      "address_2": "N/A",
      "postal_code": "07470-2094",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution to US states of AL, CA, FL, IL, KS, MI, MO, NC, NE & OH.",
      "recall_number": "Z-2140-2021",
      "product_description": "Heater Cooler Unit (HCU 30) The device supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers in extracorporeal circulation and for blankets with which patients can be warmed or cooled.",
      "product_quantity": "33",
      "reason_for_recall": "There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.",
      "recall_initiation_date": "20210714",
      "center_classification_date": "20210723",
      "termination_date": "20220803",
      "report_date": "20210804",
      "code_info": "HCU 30 Base Unit 100-120 V, Model Number 704629 I 70103.4371 HCU 30 Base Unit Canadian/US, Model Number 705501 I 70103.4653HCU 30 Base Unit 200-240 V, Model Number 939001 I 70102.8718 All serial numbers are affected."
    }
  ]
}