  "results": [
    {
      "status": "Ongoing",
      "city": "Gainesville",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88126",
      "recalling_firm": "Exactech, Inc.",
      "address_1": "2320 Nw 66th Ct",
      "address_2": "N/A",
      "postal_code": "32653-1630",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution US nationwide.",
      "recall_number": "Z-2128-2021",
      "product_description": "Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 136-36-XX, Novation GXL Liner, +5mm Lateralized, 36mm.  Orthopedic hip implant component.",
      "product_quantity": "89,050 total",
      "reason_for_recall": "Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.",
      "recall_initiation_date": "20210629",
      "center_classification_date": "20210722",
      "report_date": "20210728",
      "code_info": "All serial numbers.  UDI-DI numbers: 10885862024329, 10885862024336, 10885862024343, 10885862024350",
      "more_code_info": ""
    }
  ]
}