  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Gainesville",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88126",
      "recalling_firm": "Exactech, Inc.",
      "address_1": "2320 Nw 66th Ct",
      "address_2": "N/A",
      "postal_code": "32653-1630",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution US nationwide.",
      "recall_number": "Z-2123-2021",
      "product_description": "Exactech Connexion GXL acetabular polyethylene liners used in the following systems: Catalog numbers 132-40-XX, Novation GXL Liner, Lipped Ant, 40mm; Novation GXL Liner, Lipped, 40mm.  Orthopedic hip implant component.",
      "product_quantity": "89,050 total",
      "reason_for_recall": "Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.",
      "recall_initiation_date": "20210629",
      "center_classification_date": "20210722",
      "report_date": "20210728",
      "code_info": "All serial numbers.  UDI-DI numbers: 10885862207043, 10885862207050, 10885862207067, 10885862303349, 10885862303462, 10885862303363",
      "more_code_info": ""
    }
  ]
}