  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98642",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution.",
      "recall_number": "Z-2115-2026",
      "product_description": "Medline Convenience Kits:  1) KIT STC TRACHEOSTOMY, Model Number:DYKMBNDL55;   2) KIT STC TRACHEOSTOMY, Model Number:DYKMBNDL55A;   3) TRACH TOTE, Model Number:DYNJ85691",
      "product_quantity": "753 units",
      "reason_for_recall": "Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.",
      "recall_initiation_date": "20260107",
      "center_classification_date": "20260507",
      "report_date": "20260513",
      "code_info": "1) DYKMBNDL55, UDI-DI: 10193489478839(each), 40193489478830(case), Lot Number: 21ABK325; 2) DYKMBNDL55, UDI-DI: 10193489478839(each), 40193489478830(case), Lot Number: 21ABX057; 3) DYKMBNDL55, UDI-DI: 10193489478839(each), 40193489478830(case), Lot Number: 21CBH955; 4) DYKMBNDL55, UDI-DI: 10193489478839(each), 40193489478830(case), Lot Number: 21DBB713; 5) DYKMBNDL55, UDI-DI: 10193489478839(each), 40193489478830(case), Lot Number: 21FBA766; 6) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 21FBE305; 7) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 21FBS122; 8) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 21GBR275; 9) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 21GBR277; 10) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 21IBU128; 11) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 21JBJ227; 12) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 21WBA902; 13) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 22ABD970; 14) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 22ABM691; 15) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 22BBL555; 16) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 22BBM867; 17) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 22DBM224; 18) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 22DBR042; 19) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 22FBF305; 20) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 22FBX191; 21) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 22HBQ099; 22) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 22KBH787; 23) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 22LBK911; 24) DYKMBNDL55A, UDI-DI: 10193489897890(each), 40193489897891(case), Lot Number: 23CBN841; 25) DYNJ85691, UDI-DI: 10195327534141(each), 40195327534142(case), Lot Number: 24ABR853",
      "more_code_info": ""
    }
  ]
}