{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 Gbc Dr Ms 514",
      "reason_for_recall": "Siemens is conducting a field correction for Dimension Vista 500 and Dimension Vista 1500 systems using software versions 3.5.1 and 3.6. Under specific rare conditions which require a system reset, there is a potential for a sample to be dispensed into an aliquot well that has previously been used for a QC or calibrator product from a Vista vial. If two samples are dispensed into the same aliquot",
      "address_2": "PO BOX 6101",
      "product_quantity": "1937",
      "code_info": "Dimension Vista 500 Siemens material numbers: 10284473 (domestic), and 10488224 (international), Dimension Vista 1500 Siemens material numbers: 10444801 (domestic) and 10444802 (international).",
      "center_classification_date": "20130826",
      "distribution_pattern": "Worldwide distribution: US (nationwide) including PR and countries of: Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Malaysia, Netherlands, New Zealand, Norway, Portugal, Republic Korea, Saudia Arabia, Slovakia, Slovenia, South Africa, Spain, Switzerland, and United Kingdom.",
      "state": "DE",
      "product_description": "Dimension Vista 500 and Dimension Vista 1500  System Aliquot Well Double Dispense with software versions 3.5.1 and 3.6.     In vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids. Dimension Vista chemical and immunochemical applications use photometric, turbidimetric, chemiluminescence, nephelometric, and integrated ion-selective multisensory technology for clinical use.",
      "report_date": "20130904",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-2088-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65607",
      "termination_date": "20140827",
      "more_code_info": "",
      "recall_initiation_date": "20130619",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}