  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Abbott Park",
      "state": "IL",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83053",
      "recalling_firm": "Abbott Laboratories",
      "address_1": "100 Abbott Park Rd Bldg Ap8b",
      "address_2": "N/A",
      "postal_code": "60064-3502",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.    Distributed worldwide to ALGERIA, ANDORRA, AUSTRIA, AZERBAIJAN, BANGLADESH, BELARUS, BELGIUM, BOSNIA AND HERZEGOVI, BRUNEI, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, CROATIA, CURACAO, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GAZA & JERICHO, GEORGIA, GERMANY, GREECE, GREENLAND, HONDURAS, HUNGARY, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JORDAN, KAZAKHSTAN, KUWAIT, KYRGYZSTAN, LEBANON, LITHUANIA, LUXEMBOURG, MACEDONIA, MALDIVES, MALI, MAURITANIA, MONTENEGRO, MOROCCO, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SENEGAL, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWITZERLAND, TANZANIA, TRINIDAD & TOBAGO, TUNISIA, TURKEY, UAE, UGANDA, UNITED KINGDOM, VIETNAM, ZIMBABWE.",
      "recall_number": "Z-2078-2019",
      "product_description": "ARCHITECT BNP Calibrators    Ptoduct Usage:  The ARCHITECT BNP Calibrators are for the calibration of the ARCHITECT iSystem when used for the quantitative determination of human Btype natriuretic peptide (BNP) in human EDTA plasma.",
      "product_quantity": "5,483 units",
      "reason_for_recall": "Presence of a time dependent, stability drift in patient and control results returned from ARCHITECT BNP testing.",
      "recall_initiation_date": "20190524",
      "center_classification_date": "20190725",
      "termination_date": "20210401",
      "report_date": "20190731",
      "code_info": "a.  List Number 8K2802 - Lot Number (Expiration Date) UDI: 44K79318 (23JUL2019) (01)00380740017460(17)190723(10)44K79318, 44K80818 (05SEP2019) (01)00380740017460(17)190905(10)44K80818, 44K82118 (09OCT2019) (01)00380740017460(17)191009(10)44K82118, 44K82618 (15NOV2019) (01)00380740017460(17)191115(10)44K82618, 44K85119 (22JAN2020) (01)00380740017460(17)200122(10)44K85119;      b.  List Number 8K2803 - Lot Number (Expiration Date) UDI: 44K79418 (23Jul2019) (01)00380740145255(17)190723(10)44K79418, 44K81018 (5Sep2019) (01)00380740145255(17)190905(10)44K81018, 44K82018 (9Oct2019) (01)00380740145255(17)191009(10)44K82018, 44K85219 (22Jan2020) (01)00380740145255(17)200122(10)44K85219;      c.   List Number 8K2809, Lot Number 44K84119, Expiration Date 17Dec2019, No UDI"
    }
  ]
}