{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Billings",
      "state": "MT",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98741",
      "recalling_firm": "Windstone Medical Packaging, Inc.",
      "address_1": "1602 4th Ave N",
      "address_2": "",
      "postal_code": "59101-1521",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the state of Arizona.",
      "recall_number": "Z-2066-2026",
      "product_description": "Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Namic RA Syringes)    Angio Pack  Part numbers AMS6908E, AMS6908F",
      "product_quantity": "150 kits (US only)",
      "reason_for_recall": "Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full disconnection between syringe and manifold. If unwinding occurs, there is a potential for biohazard exposure, blood loss, infection. or air embolism.",
      "recall_initiation_date": "20260402",
      "center_classification_date": "20260508",
      "report_date": "20260520",
      "code_info": "Pack AMS6908E UDI-DI B098AMS6908E0 Lot 214875  Pack AMS6908F UDI-DI B098AMS6908F0 Lots 215436, 225001, 226706, 234088",
      "more_code_info": ""
    }
  ]
}