  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82940",
      "recalling_firm": "Medline Industries Inc",
      "address_1": "Three Lakes Drive",
      "address_2": "N/A",
      "postal_code": "60093",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "One distributor located in Illinois.",
      "recall_number": "Z-2066-2019",
      "product_description": "Medline Primary Warmer Pack, Product No. DYNJ52510A. The Medline Primary Warmer Pack contains the recalled Vyaire enFlow Disposable Cartridge, which is the subject of this recall, and the recalled BD Alaris Pump Infusion Set (RES#82951)     Product Usage:  The Vyaire enFlow Disposable Cartridges contain a heating element that is used to help raise or maintain a patient's body temperature.",
      "product_quantity": "4,680 kits",
      "reason_for_recall": "The Vyaire enFlow disposable cartridge was recalled due to the potential for unsafe aluminum levels to elute into the infusate. The Use of the BD Alaris Pump Infusion Set may lead to serious adverse events for patients due to the incomplete occlusion caused by variation in tubing wall thickness.",
      "recall_initiation_date": "20190503",
      "center_classification_date": "20190812",
      "termination_date": "20210301",
      "report_date": "20190821",
      "code_info": "Lots 2018090690  2018101190  2018101590  2018121190  2019013190  2019030490  2019040190"
    }
  ]
}