{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tempe",
      "state": "AZ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98674",
      "recalling_firm": "Erbe Medical, LLC",
      "address_1": "405 W Geneva Dr",
      "address_2": "",
      "postal_code": "85282-2003",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "US Nationwide distribution including in the states of IL, NY, CA, WA, UT, KY, PA, TX, AL, OH, FL, IN, NJ, NH, MI, CT, AR, MN, WI, AZ, CO, WV, MT, HI, TN, GA, VA, OK, RI, MO, NM, ID, SC, WY, KS, NC, MS, OR, MD, MA, DE, NE, AK, IA, DC, ME, LA, SD, ND, VT, PR, NV.",
      "recall_number": "Z-2062-2026",
      "product_description": "ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources, REF: 20325-239; Hybrid CO2 Tubing/Cap Set for Olympus Scopes & UCR, REF: 20325-240; Hybrid CO2 Tubing/Cap Set for Olympus Scopes, REF: 20325-206; Hybrid CO2 Tubing/Cap Set for Pentax Scopes, REF: 20325-207; ERBEFLO CleverCap Hybrid CO2 Tubing/Cap Set for Fujifilm Scopes, REF: 20325-248; Hybrid CO2 Tubing/Cap Set for Fujinon Scopes, REF: 20325-208",
      "product_quantity": "260,962",
      "reason_for_recall": "Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.",
      "recall_initiation_date": "20260331",
      "center_classification_date": "20260501",
      "report_date": "20260513",
      "code_info": "REF/UDI-DI/Lots:  20325-239/04065655000375/W4464506, W4464507, W4464512, W4464514, W4464515, W4464516, W4464517, W4465114, W4465115, W4465116, W4465702, W4465709, W4465991, W4466004, W4466266, W4466267, W4466268, W4466269, W4466271, W4466272, W4466273, W4466275, W4466278, W4466279, W4466813, W4466814, W4466815, W4466816, W4467057, W4467058, W4467059, W4467680, W4467681, W4468147, W4468148, W4468273, W4468274, W4468275, WO468776, WO468777, WO468778, WO468779, WO468780, WO469443, WO469444, WO469448, WO469449, WO469450, WO469451, WO469452, WO469933, WO469934, WO469947, WO469948, WO469949, WO469950, WO469951, WO469952, WO470768, WO470774, WO470775, WO470776, WO470777, WO470778, WO470896, WO470897, WO470913, WO470914, WO470915, WO470922, WO471236, WO471237, WO471238, WO471239, WO471240, WO471241, WO471242, WO473987, WO473988, WO473989, WO473991, WO475246, WO475247, WO475248, WO475249;  20325-240/04065655000405/W4466020, W4466021, WO468768, WO468769, WO468770, WO468771, WO469445, WO469446, WO469447, WO469936, WO469937, WO471244, WO471246, WO473170, WO473171, WO473172, WO473173, WO473174, WO473982, WO473983, WO473984, WO473985, WO474409, WO474410, WO474411, WO474412, WO474413, WO474414, WO475243;  20325-206/04065655000436/W4467039, W4467049, W4467678, W4467679, WO468582, WO470769, WO470770, WO470771, WO470772, WO470773, WO471076, WO471077, WO471078, WO471079, WO471390, WO478365, WO478366, WO478367;  20325-207/04065655000689/W4469939, W4469940, W4469941;  20325-248/04065655001006/W4467050, W4467051, W4467055, W4467056, W4467682, WO470898, WO470899, WO470900, WO470901, WO470902, WO470903, WO471252, WO471253, WO471254, WO471255, WO471256;  20325-208/04065655001099/W4467040, W4468271, WO468760",
      "more_code_info": ""
    }
  ]
}