  "results": [
    {
      "status": "Terminated",
      "city": "Newark",
      "state": "DE",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71315",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "500 Gbc Dr Ms 514",
      "address_2": "PO BOX 6101",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - Nationwide Distribution including Puerto Rico., and the countries of : Austria, Australia, Belgium, Canada, Denmark, France, Finland, Germany, Japan, Italy, New Zealand, Netherlands, Czech Republic, Portugal, Republic Korea,  Norway, Spain, Switzerland, Slovakia, Saudi Arabia, Slovenia,  and United Kingdom,",
      "recall_number": "Z-2054-2015",
      "product_description": "Siemens Healthcare Diagnostics Dimension Vista Systems, used with the Dimension Vista Intelligent Lab System software version 3.6.1.  In vitro diagnostic analyzer.",
      "product_quantity": "2315",
      "reason_for_recall": "Two software issues may occur in Vista software version 3.6.1. Issue #1: Samples may stop processing without notification.  Issue #2: Timing issue in Vista 1500 may cause wrong reagent or no reagent delivery. This may lead to unflagged, unexpected low results and results flagged with assay errors.",
      "recall_initiation_date": "20150521",
      "center_classification_date": "20150709",
      "termination_date": "20170310",
      "report_date": "20150715",