  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tustin",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94609",
      "recalling_firm": "Canon Medical System, USA, INC.",
      "address_1": "2441 Michelle Dr",
      "address_2": "N/A",
      "postal_code": "92780-7047",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: PA, SC, IA, AL, MI, AR, SD, TX, ID, KY, NY, MN, MD, OH, WI, NV, CT, AZ, IL, OK, CA, FL, GA, OR, IN, NJ, CO, MT, KS, MO, WY, VA, VT, LA, NH, WV, TN, PR",
      "recall_number": "Z-2041-2024",
      "product_description": "MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), Vantage Titan 3T MRT-3010 (MEXL-3010), Vantage Galan 3T MRT-3020 (MEXL-3020)",
      "product_quantity": "153",
      "reason_for_recall": "The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.",
      "recall_initiation_date": "20240510",
      "center_classification_date": "20240609",
      "report_date": "20240619",
      "code_info": "Model/UDI-DI/Serial Numbers:  MRT-1550 (MEXL-1550)/04987670103360/POA21X2002, S5A1982001, S5B22Z2041, S5B2172018, S2C2252024, S5A20Z2009, S2C21Y2021, S2B2052010, S5B2272031, POA2262005, S5B2292036, S5B2272029, S5B2282034, S5B22Y2040, S5B2292035, S6B2252008, S5A20Z2010, POA2252004, POA2182001, S1C2152071, S5B2352047, S2C2252023, S5B2122013, S6B2262010, S2C2192020, S6B2292013, S5B2132014, S6B2322018, S5B2292038, S6B2312016, S2B2052011, S2C2162018, S4A1992013, S3A19Z2013, S3A1982009, S5B2182021, S6B2322017, S5A2092008, S6B2362019, S5B22Z2042, S6B2222005, S2B2042009, S2B20Z2015, S5B2162017, S6B2262011, S5B2272030, S6B22Z2014, S5B2282033, S8B2272009, S5A2062007, S2B2072012, S2B2072013, S6B2222006, S6B2272012, S2B2072014, S6A20Z2001, POA2272006, S5A19Z2003, S1C2142070, POA2212003, S2C2142017, S5B2362048, S2A1962005, S6B2182004, S5B2292037, S6B2252007, S5B2282032, S5B2142015, S5B2382050, S2A18Z2003, S6B2112002, S6B2382021, S1B2042051, S8B2212007, S1B2062057, S1A1922022, S2A18Z2001, S1B2092067, S1C2242080, S5B2172019, S6B2252009, S6B2372020, S2A18Z2002, S2A1942004, S2C2212022, S2C2272025, S6B2392022, S2C2162019, S2C2132016, S4B2062020, S1C2262081, PDA21Y2001, S6B22Z2015, S5B22Z2043, S5B2242027,  MRT-2020 (MEXL-1520)/04987670101632/S1F18X2580, S1B1492068, S1D17Y2461, S1F2052753, S1C1722390, S1C1682336, S1E1842517, S1E1842527, S1C1672334,  MRT-1504 (MEXL-1504)/04987670102400/U5B15Z2001,  MRT-3010 (MEXL-3010)/04987670101649/A5D14X2032, A5C12Y2017, A5D13Y2025, A5C1412028, S7A1642006, A5D13Y2026, A5C1212009, A5C12X2016, A5G1692038, A5G1642036, A5F1562034, A5B11Y2005, A5C1262013, A5E1522033, A5G1692039, A5D14X2031, A5B1192003, A5B11Y2006, A5D1312019, A5D1362021, A5C1252012, A5C1292015, A5D13X2024, A5C1422029, A5B11Z2007, A5C1212008,  MRT-3020 (MEXL-3020)/04987670102615/D5A2082005, D5A2092007, D5B2152013, S5B17X2034, D7B2182017, 5DB2152002, S5B18X2046, S5A17X2033, D5A2062003, S5A1742021, D5B2212019, D5B2192016, S5A1662001, S5C18Y2048, S5A1732019, S5A1732018, S5A17Y2035, S7B1892004, S5B1842043, D7B2152015, S5C1972054, 5DB2172003"
    }
  ]
}