      "status": "Ongoing",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96946",
      "recalling_firm": "Ortho-Clinical Diagnostics, Inc.",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) and OUS (International) to countries of:  Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom.",
      "recall_number": "Z-2039-2025",
      "product_description": "Brand Name: VITROS;  Product Name: VITROS XT 7600 Integrated System;  Model/Catalog Number: (1) 6844461, (2) 6272222;  Software Version: version 3.2 to version 3.8.3 (all available software versions);  Product Description: In vitro diagnostic chemistry analyzer.",
      "product_quantity": "1958 units (1019 US, 939 OUS)",
      "reason_for_recall": "a software anomaly allows test results to be reported using Micro Tip and Micro Slide  diluent packs that the system treats as expired even when the  Use Expired Reagents  setting is not enabled if  a shelf expiration date is omitted during manual loading. The issue was identified by a customer complaint, in  which a CREA Urine test was processed with an  RE  (Reagent Expired) code, even though the diluent pack was  not expired and the use-expired-reagents setting was disabled. This can result in erroneous test results and/or  delaying test results.",
      "recall_initiation_date": "20250509",
      "center_classification_date": "20250627",
      "report_date": "20250709",
      "code_info": "VITROS XT 7600 Integrated System; Product Code: 6844461; UDI: 10758750012343;  VITROS XT 7600 System - Certified; Product Code: 6272222; UDI: 10758750012343;  Range of installed serial numbers:76000109-76002327;  Total systems currently installed in the US: 1,019 Total systems currently installed outside the US: 939 Total systems installed worldwide (Product Quantity Distributed): 1,958   Note1: The catalogue number 6844461 is used to denote the VITROS XT 7600 Integrated System.  Catalogue number 6272222 is used to denote VITROS XT 7600 System   Certified.  A VITROS XT 7600 Integrated System maintains its original serial number through the certification process.  Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined.  Note2: When manufactured, the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems.  Therefore, the amount of each system manufactured may not equal the amount of each system installed.  Additionally, not all serial numbers listed are associated with an active installed analyzer in use today.",
      "more_code_info": ""
    }
  ]
}