  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mentor",
      "state": "OH",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68696",
      "recalling_firm": "Steris Corporation",
      "address_1": "5960 Heisley Rd",
      "address_2": "",
      "postal_code": "44060-1834",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "US Nationwide Distribution - in the states of AL, AR, CA, DE, FL, GA, IL, IA, LA, MA, MI, NJ, NY, OH, OK, OR, SC, TX, VA. and in the country of Canada.",
      "recall_number": "Z-2036-2014",
      "product_description": "Reliance 1227 Cart and Utensil Washer/Disinfector    Product Usage:  Usage:  The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the cleaning and low-level disinfection of bedpans and urinals, basins, carts, beds, theater shoes and other miscellaneous reusable items used in the care of patients.",
      "product_quantity": "37",
      "reason_for_recall": "STERIS has identified that deformation of the washer\u0019s door handle pin may occur when the door handle is rotated with sufficient force.  Should deformation and/or breakage of the door handle pin occur there is a potential for the washer door to become inoperable.",
      "recall_initiation_date": "20140630",
      "center_classification_date": "20140710",
      "termination_date": "20150701",
      "report_date": "20140716",
      "code_info": "Serial Number Range: 3632511020 - 3613214010",
      "more_code_info": ""
    }
  ]
}