  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83203",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-2034-2019",
      "product_description": "Artis zeego, Model Number 10280959",
      "product_quantity": "117",
      "reason_for_recall": "A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.",
      "recall_initiation_date": "20190612",
      "center_classification_date": "20190722",
      "termination_date": "20200918",
      "report_date": "20190731",
      "code_info": "Serial Numbers: 160395  160484  160502  160351  160392  160449  160463  160313  160411  160406  160106  160134  160136  160347  160312  160488  160448  160427  160455  160486  160435  160451  160346  160384  160100  160493  160391  160321  160376  160447  160471  160481  160401  160119  160403  160404  160501  160438  160388  160378  160419  160114  160450  160429  160149  160137  160152  160153  160154  160138  160446  160368  160394  160412  160387  160432  160147  160433  160483  160335  160402  160356  160425  160148  160337  160460  160485  160443  160461  160370  160381  160508  160122  160407  160325  160494  160390  160507  160144  160487  160418  160007  160001  160363  160341  160139  160430  160482  160006  160462  160520  160505  160445  160322  160323  160382  160421  160474  160116  160428  160131  160318  160357  160369  160498  160413  160414  160128  160373  160509  160551  160327  160338  160477  160519  160141  160469"
    }
  ]
}