      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98664",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-2030-2026",
      "product_description": "Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows:    1. MICRODISECTOMY PACK, Medline Kit Number/SKU DYNJ80585B;   2. MICRODISECTOMY PACK, Medline Kit Number/SKU DYNJ80585C;   3, MICRODISECTOMY PACK, Medline Kit Number/SKU DYNJ80585D.",
      "product_quantity": "924 kits",
      "reason_for_recall": "During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.",
      "recall_initiation_date": "20260319",
      "center_classification_date": "20260430",
      "report_date": "20260506",
      "code_info": "Medline Kit Number/SKU DYNJ80585B:  UDI/DI each 10195327211158, UDI/DI case 40195327211159, Lot Number: 22GBV101;   Medline Kit Number/SKU DYNJ80585B:  UDI/DI each 10195327211158, UDI/DI case 40195327211159, Lot Number: 24IBG473;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25EBT656;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25DBN421;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25CBN640;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25BBL172;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25BBD651;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25ABA455;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 24KBO858;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 24JBH693;   Medline Kit Number/SKU DYNJ80585D:  UDI/DI each 10198459389337, UDI/DI case 40198459389338, Lot Number: 26ABH754;   Medline Kit Number/SKU DYNJ80585D:  UDI/DI each 10198459389337, UDI/DI case 40198459389338, Lot Number: 25LBA242;   Medline Kit Number/SKU DYNJ80585D:  UDI/DI each 10198459389337, UDI/DI case 40198459389338, Lot Number: 25HBT362;   Medline Kit Number/SKU DYNJ80585D:  UDI/DI each 10198459389337, UDI/DI case 40198459389338, Lot Number: 25GBL430;   Medline Kit Number/SKU DYNJ80585D:  UDI/DI each 10198459389337, UDI/DI case 40198459389338, Lot Number: 25FBT402.",
      "more_code_info": ""
    }
  ]
}