  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83203",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-2030-2019",
      "product_description": "Artis zee multi, Model Number 10094139",
      "product_quantity": "89",
      "reason_for_recall": "A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.",
      "recall_initiation_date": "20190612",
      "center_classification_date": "20190722",
      "termination_date": "20200918",
      "report_date": "20190731",
      "code_info": "Serial Numbers: 157435  157509  157660  157710  157458  130017  157698  157532  157449  157657  157136  157690  157616  157560  157436  157677  157528  157503  157615  157448  157165  157700  157602  157168  157642  157213  157164  157559  157156  157524  157588  157750  157579  157573  157650  157671  157554  157555  157521  157696  157488  157644  157648  157169  157668  157508  157510  157601  157139  158058  157536  157625  157600  157604  157565  157634  157635  157146  157583  157476  157497  157570  157658  157505  157594  157706  157586  157674  157414  157215  157692  157707  157507  157708  157544  157184  157659  157661  157531  157409  157688  157702  157704  157466  157442  157667  157438  157440  157645"
    }
  ]
}