  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96952",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, Bahamas.",
      "recall_number": "Z-2027-2025",
      "product_description": "Product: Atellica CH Diluent - CONS 2 x 1.5L;  Siemens Material Number (SMN): 11099300;",
      "product_quantity": "1,625 units (1450 US, 175 OUS)",
      "reason_for_recall": "Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.",
      "recall_initiation_date": "20250522",
      "center_classification_date": "20250625",
      "report_date": "20250702",
      "code_info": "Siemens Material Number (SMN): 11099300; UDI-DI: 00630414595771; Lot Number: 0000195412;",
      "more_code_info": ""
    }
  ]
}