{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98664",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-2018-2026",
      "product_description": "Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows:    1. OSC COLEMAN FAT HARVEST KIT, Medline Kit Number/SKU DYNJ32854B;   2. RR-DR. CHHETRI MDL PACK, Medline Kit Number/SKU DYNJ62635B.",
      "product_quantity": "208 kits",
      "reason_for_recall": "During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.",
      "recall_initiation_date": "20260319",
      "center_classification_date": "20260430",
      "report_date": "20260506",
      "code_info": "Medline Kit Number/SKU DYNJ32854B:  UDI/DI each 10198459413414, UDI/DI case 40198459413415, Lot Number : 25GBU656;  Medline Kit Number/SKU DYNJ62635B:  UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25GMD214;  Medline Kit Number/SKU DYNJ62635B:  UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25FMC376;  Medline Kit Number/SKU DYNJ62635B:  UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25DMJ030;  Medline Kit Number/SKU DYNJ62635B:  UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25BME902;  Medline Kit Number/SKU DYNJ62635B:  UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25AMJ032;  Medline Kit Number/SKU DYNJ62635B:  UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 24IMH612;  Medline Kit Number/SKU DYNJ62635B:  UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 24EME486;  Medline Kit Number/SKU DYNJ62635B:  UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 24DMD315;  Medline Kit Number/SKU DYNJ62635C:  UDI/DI each 10198459485619, UDI/DI case 40198459485610, Lot Number : 26BMD607;  Medline Kit Number/SKU DYNJ62635C:  UDI/DI each 10198459485619, UDI/DI case 40198459485610, Lot Number : 25JMD196.",
      "more_code_info": ""
    }
  ]
}