{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98664",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-2008-2026",
      "product_description": "Medline Neuro Sponges, various dimensions, labeled as:    1. Neuro Sponges,  Eponges neuro,  Neuroesponjas,  size 0.25\" X 0.25\", XR, 10/PK, Medline SKU NEUROSPNG02;   2. Neuro Sponges,  Eponges neuro,  Neuroesponjas,  size 0.5\" X 0.5\", XR, 10/PK, Medline SKU NEUROSPNG05;   3. Neuro Sponges,  Eponges neuro,  Neuroesponjas,  size 0.5\" X 1\", XR, LF, 10/PK, Medline SKU NEUROSPNG06;   4. Neuro Sponges,  Eponges neuro,  Neuroesponjas,  size 0.5\" X 1.5\", XR, LF, 10/PK, Medline SKU NEUROSPNG07;   5. Neuro Sponges,  Eponges neuro,  Neuroesponjas,  size 0.5\" X 3\", XR, LF, 10/PK, Medline SKU NEUROSPNG09;   6. Neuro Sponges,  Eponges neuro,  Neuroesponjas,  size .75\" X .75\", XR, LF, 10/PK, Medline SKU NEUROSPNG11;   7. Neuro Sponges,  Eponges neuro,  Neuroesponjas,  size 1\" X 1\", XR, LF, 10/PK, Medline SKU NEUROSPNG13;   8. Neuro Sponges,  Eponges neuro,  Neuroesponjas,  size 1\" X 3\", XR, LF, 10/PK, Medline SKU NEUROSPNG14.",
      "product_quantity": "4225160 units",
      "reason_for_recall": "During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.",
      "recall_initiation_date": "20260319",
      "center_classification_date": "20260430",
      "report_date": "20260506",
      "code_info": "1. NEUROSPNG02:  UDI/DI each 10197344111572, UDI/DI case 20197344111579, ALL LOTS;  2. NEUROSPNG05:  UDI/DI each 20080196286659, UDI/DI case 30080196286656, ALL LOTS;  3. NEUROSPNG06:  UDI/DI each 20080196286505, UDI/DI case 30080196286502, ALL LOTS;  4. NEUROSPNG07:  UDI/DI each 10080196286485, UDI/DI case 40080196286486, ALL LOTS;  5. NEUROSPNG09:  UDI/DI each 20080196286536, UDI/DI case 30080196286533, ALL LOTS;  6. NEUROSPNG11:  UDI/DI each 20080196286550, UDI/DI case 30080196286557, ALL LOTS;  7. NEUROSPNG13:  UDI/DI each 10080196286577, UDI/DI case 40080196286578, ALL LOTS;  8. NEUROSPNG14:  UDI/DI each 10080196286584, UDI/DI case 40080196286585, ALL LOTS.",
      "more_code_info": ""
    }
  ]
}