{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Wehrheim",
      "state": "",
      "country": "Germany",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98608",
      "recalling_firm": "Heraeus Medical GmbH (Dental Division)",
      "address_1": "Philipp-Reis-Str. 8",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;",
      "recall_number": "Z-1997-2026",
      "product_description": "Heraeus, PALAMIX uno. Material Number: 66057893.",
      "product_quantity": "51,771 units",
      "reason_for_recall": "Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).",
      "recall_initiation_date": "20260325",
      "center_classification_date": "20260429",
      "report_date": "20260506",
      "code_info": "Material Number: 66057893; UDI-DI: 04260102133587; Batch: all batches with numbers below 223060;",
      "more_code_info": ""
    }
  ]
}