      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97050",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Nationwide distribution.",
      "recall_number": "Z-1982-2025",
      "product_description": "UROSKOP Omnia Max. Model Number: 10762473",
      "product_quantity": "25 units",
      "reason_for_recall": "A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.",
      "recall_initiation_date": "20250604",
      "center_classification_date": "20250617",
      "report_date": "20250625",
      "code_info": "Model Number: 10762473. UDI Numbers: (01)04056869008981(21)4014, (01)04056869008981(21)4056, (01)04056869008981(21)4049, (01)04056869008981(21)4015, (01)04056869008981(21)4063, (01)04056869008981(21)4022, (01)04056869008981(21)4036, (01)04056869008981(21)4043, (01)04056869008981(21)4024, (01)04056869008981(21)4059, (01)04056869008981(21)4020, (01)04056869008981(21)4032, (01)04056869008981(21)4053, (01)04056869008981(21)4061, (01)04056869008981(21)4060, (01)04056869008981(21)4051, (01)04056869008981(21)4025, (01)04056869008981(21)4037, (01)04056869008981(21)4054, (01)04056869008981(21)4042, (01)04056869008981(21)4033, (01)04056869008981(21)4055, (01)04056869008981(21)4052, (01)04056869008981(21)4047, (01)04056869008981(21)4062. Serial Numbers: 4014, 4056, 4049, 4015, 4063, 4022, 4036, 4043, 4024, 4059, 4020, 4032, 4053, 4061, 4060, 4051, 4025, 4037, 4054, 4042, 4033, 4055, 4052, 4047, 4062.",
      "more_code_info": ""
    }
  ]
}