      "status": "Ongoing",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87974",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "N/A",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, Mexico, AE,  AR, AT, AU, BD, BE, BR, CH, CL, CN, CO, CU, CY, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GE, GI, GL, GR, HK, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KG, KH, KR, LA, LB, LK, LT, LY, MM, MN, MU, MV, MY, NI, NL, NO, NZ, OM, PA, PF, PH, PK, PL, PS, PT, QA, RE, RU, RW, SA, SE, SG , SI, SK, TH, TN, TR, UZ, VN, XK, YE, ZA, ZW.",
      "recall_number": "Z-1982-2021",
      "product_description": "Continuous CT Fluoroscopy DVI Ceiling Mount   CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT systems.  Brilliance 64, Brilliance 40, Ingenuity Flex    728235 - Brilliance 40  728231/728232 - Brilliance 64  728317/728318 - Ingenuity Flex",
      "product_quantity": "1,301 Monitor Bracket Assemblies total",
      "reason_for_recall": "Significant wear of the Monitor Bracket Assembly shaft  goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the  cables resulting in patient, operator, bystander, or service person  to serious injury.",
      "recall_initiation_date": "20210501",
      "center_classification_date": "20210630",
      "report_date": "20210707",
      "code_info": "Model/ Serial Number/UDI: N/A  728235 90040  728235 90191  728235 90192  728235 90061  728235 29090  728235 90111  728235 29042  728235 90165  728235 29082  728235 29152  728235 90054  728235 90087  728235 90122  728235 90142  728235 90104  728235 90075  728235 9237  728235 90110  728235 90084  728235 90208  728235 9047  728235 90125  728235 90171  728235 90212  728235 90056  728235 90227  728235 90164  728235 90051  728235 90035    728231 9367  728231 95709  728231 10625  728231 10379  728231 9900  728231 10630  728231 10755  728231 10758  728231 90094..  728231 9524  728231 95443  728231 95463  728231 90046  728231 95706  728231 95305  728231 10139  728231 29086  728231 10273  728231 95046  728231 95044  728231 95324  728231 90180  728231 95062  728231 95155  728231 95206  728231 95122  728231 95159  728231 90199  728231 95434  728231 95717  728231 95328  728231 95401  728231 10393  728231 90154  728231 95365  728231 95457  728231 95054  728231 95267  728231 95156  728231 95515  728231 10659  728231 10635  728231 29056  728231 95601  728231 95712  728231 95404  728231 90141  728231 90153  728231 95087  728231 95029  728231 90045  728231 95038  728231 90100  728231 95023  728231 95089  728231 90152  728231 90151  728231 64081  728231 10060  728231 90042  728231 90077  728231 95345  728231 9755  728231 90098  728231 9898  728231 9621  728231 9859  728231 90106  728231 10067  728231 9670  728231 9937  728231 9066  728231 10733  728231 90128  728231 9703  728231 9716  728231 9151  728231 9655  728231 10698  728231 9805  728231 10111  728231 9862  728231 9045  728231 10221  728231 10773  728231 9584  728231 9948  728231 95036  728231 9960  728231 90158  728231 9074  728231 95160  728231 95695  728231 10470  728231 29049  728231 9166  728231 90033  728231 95721  728231 10628  728231 29135  728231 95480  728231 9793  728231 29044  728231 9168  728231 95633  728231 9554  728231 9882  728231 9875  728231 10381  728231 10000  728231 9215  728231 95315  728231 95204  728231 10203  728231 10742  728231 10596  728231 90221  728231 95238  728231 90186  728231 10651  728231 10503  728231 10407  728231 10409  728231 90058  728231 10686  728231 10580  728231 95374  728231 90139  728231 64051  728231 95020  728231 95058  728231 95096  728231 10136  728231 9761  728231 9757  728231 90081  728231 10316  728231 9802  728231 N/A  728231 90056  728231 95035  728231 95143  728231 3491  728231 10446  728231 90227  728231 95271  728231 10551  728231 90176  728231 9969  728231 9656  728231 90099  728231 95720  728231 95719  728231 95168  728231 95669  728231 10506  728231 10065  728231 9579  728231 95704  728231 10600  728231 9950  728231 95330  728231 10066  728231 90057    728232 122850    728317 345281 (01)00884838059528(21)345281  728317 345321 (01)00884838059528(21)345321  728317 345217 N/A  728317 345306 (01)00884838059528(21)345306  728317 345300 (01)00884838059528(21)345300  728317 345003 N/A  728317 345213 N/A  728317 345307 (01)00884838059528(21)345307  728317 345287 (01)00884838059528(21)345287  728317 345293 (01)00884838059528(21)345293  728317 345021 N/A  728317 345201 N/A  728317 345252 N/A  728232 122850 N/A  728317  N/A  728317 340003 N/A  728317 260003 N/A  728317 345210 N/A  728317 345291 (01)00884838059528(21)345291  728317 345311 (01)00884838059528(21)345311  728317 345277 (01)00884838059528(21)345277  728317 345165 (01)00884838059528(21)345165  728317 345317 (01)00884838059528(21)345317  728317 345024 N/A  728317 345002 N/A  728317 345318 (01)00884838059528(21)345318  728317 345279 (01)00884838059528(21)345279  728317 345205 N/A  728317 N/A N/A  728317 345206 N/A  728317 345023 N/A  728317 345253 N/A  728317 345220 N/A  728317 345275 (01)00884838059528(21)345275  728317 345257 N/A  728317 345005 N/A  728317 345260 N/A  728317 345269 N/A  728317 345212 N/A  728317 345241 N/A  728317 345231 N/A  728317 345288 (01)00884838059528(21)345288"
    }
  ]
}