{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tiffin",
      "state": "OH",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98720",
      "recalling_firm": "American Contract Systems Inc",
      "address_1": "85 Shaffer Park Dr",
      "address_2": "",
      "postal_code": "44883-9290",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US distribution to MN. No OUS distribution.",
      "recall_number": "Z-1981-2026",
      "product_description": "Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor    CORONARY ANGIO PACK  Model numbers ANCA80AP  ANCA80AQ",
      "product_quantity": "7,311 kits",
      "reason_for_recall": "Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.",
      "recall_initiation_date": "20260319",
      "center_classification_date": "20260506",
      "report_date": "20260513",
      "code_info": "ANCA80AP UDI-DI 191072232168 Lot 42-8146111  ANCA80AQ UDI-DI 191072236678 Lots  42-8207211 42-8267711 42-8311311 42-8337611 42-8353311 42-8379211 42-8407111 42-8510711 42-8575111 42-8611511 42-8718911 42-8718912 42-8722611 42-8805311 42-8820611 42-8870911",
      "more_code_info": ""
    }
  ]
}