  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Ottignies-Louvain-La-Neuve",
      "state": "N/A",
      "country": "Belgium",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94576",
      "recalling_firm": "Ion Beam Applications S.A.",
      "address_1": "Chemin Du Cyclotron 3",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the state of Virginia and the country of India.",
      "recall_number": "Z-1945-2024",
      "product_description": "IBA Proton Therapy System - PROTEUS 235 - Designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.",
      "product_quantity": "2 units",
      "reason_for_recall": "Irradiation is not interrupted when some C230 Accelerator Control Unit critical parameters are outside of the defined tolerances",
      "recall_initiation_date": "20240510",
      "center_classification_date": "20240603",
      "report_date": "20240612",
      "code_info": "UDI: SAT140: (01)05404013814015(11)190507(21)SAT140 SAT136: (01)05404013801138(11)201105(21)SAT136  Serial Numbers: SAT136, SAT140 PTS-11 versions from PTS-11.1 and PTS-12 versions, with C230 accelerator",
      "more_code_info": ""
    }
  ]
}