  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64992",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "Worldwide distribution: USA (nationwide) including states of: AL, AK, AZ, AR, CA, CO, CT, DC, FL, GA, GUAM, IL, IN, IA, KY, LA, ME, MD, MA, MI, MN, MS, MO, NV, NH, NH, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY; and countries of:  Albania, Australia, Bahrain, Belgium, Canada, Chile, Czech Republic, Dominican Republic, France, Germany, Great Britain, Honduras, Hungary, Ireland, Israel, Italy, Japan, Korea, Kuwait, Netherlands, Nigeria, Norway, Romania, Saudi Arabia, Spain, United Arab Emirates and Venezuela.",
      "recall_number": "Z-1942-2013",
      "product_description": "GE Healthcare, Optima XR220amx, Optima XR200amx, and Brivo XR285amx.    The series are indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. The systems are indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body pants with the patient sitting, standing, or lying in the prone or supine position. These devices are not intended for mammographic applications.",
      "product_quantity": "458 (306 US; 152 OUS)",
      "reason_for_recall": "GE Healthcare conducted a recall on various mobile x-ray units (Optima XR220amx, Optima XR200amx, and Brivo XR285amx systems)  because of an Erratic Drive Performance.",
      "recall_initiation_date": "20120807",
      "center_classification_date": "20130809",
      "termination_date": "20140204",
      "report_date": "20130821",
      "code_info": "Serial #  1028082  00000102856WK4  00001027044WK5  00001027046WK0  00001027160WK9  00001027162WK5  00001027164WK1  00001027174WK0  00001027446WK2  00001027449WK6  00001027526WK1  00001027570WK9  00001027739WK0  00001027881WK0  00001027897WK6  00001027900WK8  00001027901WK6  00001027902WK4  00001027903WK2  00001027904WK0  00001027942WK0  00001028045WK1  00001028046WK9  00001028047WK7  00001028048WK5  00001028063WK4  00001028082WK4  00001028219WK2  00001028300WK0  00001028301WK8  00001028304WK2  00001028454WK5  00001028460WK2  00001028461WK0  00001028564WK1  00001028590WK6  00001028591WK4  00001028592WK2  00001028593WK0  00001028594WK8  00001028616WK9  00001028648WK2  00001028649WK0  00001028650WK8  00001028651WK6  00001028653WK2  00001028655WK7  00001028656WK5  00001028657WK3  00001028861WK1  00001028867WK8  00001028931WK2  00001028102WK0  00000102866WK4  00001028918WK9  00001027502WK2  00001027935WK4  00001028206WK9  00001028079WK0  00001028230WK9  00001028277WK0  00001028913WK0  00001027984WK2  00001027349WK8  00001027217WK7  00001027292WK0  00001028037WK8  00001027352WK2  00001027382WK9  00001027345WK6  00001027879WK4  00001028044WK4  00001028075WK8  00001027768WK9  00001027152WK6  00001028025WK3  00001027820WK8  00001027262WK3  00001027774WK7  00001028080WK8  00001028078WK2  00001027501WK4  00001027505WK5  00001028294WK5  00001027361WK3  00001028041WK0  00001028065WK9  00001027322WK5  00001027387WK8  00001027362WK1  00001027344WK9  00001027359WK7  00001028401WK6  00001028491WK7  00001027371WK2  00001028279WK6  00001028276WK2  00001028231WK7  00001028267WK1  00001028076WK6  00001027824WK0  00001027450WK4  00001027992WK5  00001027895WK0  00001027775WK4  00001027421WK5  00001027894WK3  00001027503WK0  00001027693WK9  00001028235WK8  00001028040WK2  00001028039WK4  00001028337WK2  00001027169WK0  00001027207WK8  00001027376WK1  00001027214WK4  00001027356WK3  00001027383WK7  00001027490WK0  00001027773WK9  00001027369WK6  00001027380WK3  00001027583WK2  00001028292WK9  00001028341WK4  00001028291WK1  00001028213WK5  00001027317WK5  00001028284WK6  00001028278WK8  00001027296WK1  00001028115WK2  00001028021WK2  00001027466WK0  00001027996WK6  00001027994WK1  00001028035WK2  00001028034WK5  00001028365WK3  00001028094WK9  00001027390WK2  00001027167WK4  00001027294WK6  00001027822WK4  00001027544WK4  00001028092WK3  00001027354WK8  00001027825WK7  00001027271WK4  00001027541WK0  00001027506WK3  00001028465WK1  00001027385WK2  00001027052WK8  00001028273WK9  00001027355WK5  00001027299WK5  00001027892WK7  00001027407WK4  00001027422WK3  00001027406WK6  00001027353WK0  00001029030WK2  00001027350WK6  00001027939WK6  00001028288WK7  00001027416WK5  00001027737WK4  00001027991WK7  00001027279WK7  00001027327WK4  00001027989WK1  00001028033WK7  00001027389WK4  00001027465WK2  00001028210WK1  00001028061WK8  00001027736WK6  00001027272WK2  00001028237WK4  00001028363WK8  00001027303WK5  00001028114WK5  00001027988WK3  00001027987WK5  00001028478WK4  00001027748WK1  00001028077WK4  00001028050WK1  00001027278WK9  00001028064WK2  00001027582WK4  00001027891WK9  00001028003WK0  00001027765WK5  00001027326WK6  00001027153WK4  00001027444WK7  00001027934WK7  00001028113WK7  00001027203WK7  00001027295WK3  00001027451WK2  00001028482WK6  00001027779WK6  00001027700WK2  00001027386WK0  00001027438WK9  00001028227WK5  00001027887WK7  00001027888WK5  00001027280WK5  00001028398WK4  00001028339WK8  00001028352WK1  00001028354WK7  00001028207WK7  00001028282WK0  00001028332WK3  00001028281WK2  00001027318WK3  00001027301WK9  00001028038WK6  00001027384WK5  00001027417WK3  00001027699WK6  00001027553WK5  00001027508WK9  00001027766WK3  00001028042WK8  00001027540WK2  00001027584WK0  00001028053WK5  00001027995WK8  00001027489WK2  00001027168WK2  00001027298WK7  00001027360WK5  00001027370WK4  00001028238WK2  00001028490WK9  00001027896WK8  00001028036WK0  00001028043WK6  00001027267WK2  00001027878WK6  00001027464WK5  00001027762WK2  00001028283WK8  00001028334WK9  00001028333WK1  00001027372WK0  00001027551WK9  00001027488WK4  00001027374WK6  00001028051WK9  00001027877WK8  00001027302WK7  00000102789WK3  00001027269WK8  00001027325WK8  00001027507WK1  00001028353WK9  00001027297WK9  00001028665WK6  00001028927WK0  00001027175WK7  00001027256WK5  00001027277WK1  00001027314WK2  00001027377WK9  00001028215WK0  00001028217WK6  00001028214WK3  00000002797EN9  00000002799EN5  00000002798EN7  00000002796EN1  00001027193WK0  00001027276WK3  00001028119WK4  00001028476WK8  00001027324WK1  00001027375WK3  00001027273WK0  00001027365WK4  00001028049WK3  00001027293WK8  00001027373WK8  00001027351WK4  00001027323WK3  00001027347WK2  00001027151WK8  00001027880WK2  00001027581WK6  00001028280WK4  00001027058WK5  00001027827WK3  00001027453WK8  00001027348WK0  00001027388WK6  00001027487WK6  00001027199WK7  00001028002WK2  00001027368WK8  00001028120WK2  00001028289WK5  00001028232WK5  00001028287WK9  00001028218WK4  00001027819WK0  00001027542WK8  00001027543WK6  00001027504WK8  00001027932WK1  00001028295WK2  00001027163WK3  00001027219WK3  00001027379WK5  00001027448WK8  00001027925WK5  00001027926WK3  00001027929WK7  00001027930WK5  00001027931WK3  00001027941WK2  00001028458WK6  00001028462WK8  00001027986WK7  00001027885WK1  00001027884WK4  00001027528WK7  00001027886WK9  00001027671WK5  00001027061WK9  00001027263WK1  00001027568WK3  00001028225WK9  00001028270WK5  00001027672WK3  00001027968WK5  00001028001WK4  00001027195WK5  00001027826WK5  00001027893WK5  00001027970WK1  00001027985WK9  00001028000WK6  00001027527WK9  00001027999WK0  00001027997WK4  00001027321WK7  00001027346WK4  00001027300WK1  00001028024WK6  00001028116WK0  00001028344WK8  00001027363WK9  00001027159WK1  00001028568WK2  00001026555WK1  00001027161WK7  00001027198WK9  00001027220WK1  00001027261WK5  00001027266WK4  00001027282WK1  00001027378WK7  00001027586WK5  00001027943WK8  00001028652WK4  00001028865WK2  00001028587WK2  00001028224WK2  00001028234WK1  00001027539WK4  00001027265WK6  00001028469WK3  00001028052WK7  00001028111WK1  00001028299WK4  00001028455WK2  00001028658WK1  00001028397WK6  00001027763WK0  00001028298WK6  00001028456WK0  00001028459WK4  00001028798WK5  00001028866WK0  00001027268WK0  00001027937WK0  00001027525WK3  00001027983WK4  00001027882WK8  00001028236WK6  00001028274WK7  00001028062WK6  00001027270WK6  00001028369WK5  00001027204WK5  00001027176WK5  00001027691WK3  00001027200WK3  00000102726WK9  00001027154WK2  00001027260WK7  00001028222WK6  00001028223WK4  00001028302WK6  00001028457WK8  00001028463WK6  00001028566WK6  00000002800EN1  00000002801EN9  00001027692WK1  00001028221WK8  00001028220WK0  00001028297WK8  00001028974WK2  00001027778WK8  00001027772WK1  00001027890WK1  00001028022WK0  00001028091WK5  00001027823WK2  00001027993WK3  00001028296WK0  00001028023WK8  00001027149WK2  00001027259WK9  00001027915WK6  00001027828WK1  00001028208WK5  00001027150WK0  00001027452WK0  00001027769WK7  00001027194WK8  00001027146WK8  00001027281WK3  00001027418WK1  00001028066WK7  00001028275WK4  00001028345WK5  00001028347WK1  00001027821WK6  00001027781WK2  00001027423WK1  00001028346WK3  00001028595WK5  00001027998WK2  00001028205WK1  00001027764WK8  00001027445WK4  00001028216WK8  00001027749WK9  00001028112WK9",
      "more_code_info": ""
    }
  ]
}