  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96999",
      "recalling_firm": "Philips North America",
      "address_1": "222 Jacobs St",
      "address_2": "",
      "postal_code": "02141-2296",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of AZ, CA, DE, MA, MI, MY, NE, NY, SD and the countries of Australia, Brazil, Ecuador, Germany, Iran, Italy, Japan, Lebanon, Libya,  Mexico, Poland, Portugal, Spain.",
      "recall_number": "Z-1930-2025",
      "product_description": "Intera 1.5T Achieva Nova  Product Number: 781172;",
      "product_quantity": "16 units (1 US, 15 OUS)",
      "reason_for_recall": "identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage.",
      "recall_initiation_date": "20250529",
      "center_classification_date": "20250609",
      "report_date": "20250618",
      "code_info": "Product Number: 781172; UDI-DI: no UDI; Serial Numbers: 21218, 21217, 21216, 21289, 21349, 12146, 12063, 12086, 20069, 21027, 21198, 12099, 18641, 21476, 18623, 12078;",
      "more_code_info": ""
    }
  ]
}