  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Chaska",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98591",
      "recalling_firm": "Beckman Coulter, Inc.",
      "address_1": "1000 Lake Hazeltine Dr",
      "address_2": "N/A",
      "postal_code": "55318-1037",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Bangladesh, Bolivia (Plurinational State of), Bosnia and Herzegovina, Brazil, Canada, China, C¿te d'Ivoire, Croatia, Czechia, Denmark, Egypt, El Salvador, Ethiopia, Georgia, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Kosovo, Lebanon, Libya, Malawi, Mexico, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, North Macedonia, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Somalia, Spain, Switzerland, Taiwan, Province of China, Thailand, Turkey, United Arab Emirates, Uzbekistan, Viet Nam, Yemen.",
      "recall_number": "Z-1922-2026",
      "product_description": "Access Total T4 Calibrator, Catalog No. 33805",
      "product_quantity": "2068 US, 2830 OUS",
      "reason_for_recall": "Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.",
      "recall_initiation_date": "20260316",
      "center_classification_date": "20260422",
      "report_date": "20260429",
      "code_info": "UDI-DI 15099590225872 Lots 538729, 538815, 538898, 539107, 539243",
      "more_code_info": ""
    }
  ]
}