      "status": "Terminated",
      "city": "Louisville",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85354",
      "recalling_firm": "Medtronic Navigation, Inc.",
      "address_1": "826 Coal Creek Cir",
      "address_2": "N/A",
      "postal_code": "80027-9710",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: AL, AZ, AR, CA, CO, CT, DC, FL, GA, IL, IN, KY, LA, ME, MD, MA, MI, MN, MO, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, WI    OUS: France, Germany, Greece, Iceland, Israel, Spain, Sweden, Switzerland, UK",
      "recall_number": "Z-1905-2020",
      "product_description": "Visualase Cooled Laser Applicator System Tubing Set kits    Product numbers:  VCLAS 9735559, 3mm tip/1.65mm catheter  VCLAS 9735560, 10mm tip/1.65mm catheter  VCLAS 9735561, 15mm tip/1.85mm catheter",
      "product_quantity": "4654 units",
      "reason_for_recall": "The firm received complaints on the saline tubing of Visualase Cooled Laser Applicator System (VCLAS) kit, including incorrect Luer connector assembly, incorrect drip-chamber assembly, and/or excessive adhesive causing tubing occlusion, all of which could result in disruption or prevention of saline flow.",
      "recall_initiation_date": "20190708",
      "center_classification_date": "20200505",
      "termination_date": "20241016",
      "report_date": "20200513",
      "code_info": "All kits with lot numbers from 0211041602 through 0217695790",
      "more_code_info": ""
    }
  ]
}