      "status": "Ongoing",
      "city": "Lexington",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94443",
      "recalling_firm": "FUJIFILM Healthcare Americas Corporation",
      "address_1": "81 Hartwell Ave Ste 300",
      "address_2": "",
      "postal_code": "02421-3160",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AK, CA, FL, IL, NE, NH, OH, TN, WI.",
      "recall_number": "Z-1898-2024",
      "product_description": "FDR Visionary Suite - Intended to generate digital or conventional radiographic  images of the skull, spinal column, chest, abdomen, extremities, and other body parts of  human anatomies in all routine radiography examinations.    Model/Catalog Number:  566-16130-23  566-16130-33",
      "product_quantity": "18 units",
      "reason_for_recall": "Bolts on the CH-200 tube may rupture due to fatigue failure, causing the tube support to fall, could result in patient injury",
      "recall_initiation_date": "20240411",
      "center_classification_date": "20240524",
      "report_date": "20240605",
      "code_info": "UDI-DI: 04540217052226 04540217057450  S/N: MP95A9F6A001, MP95A9F74002, MP95A9F74001, MP95AA983002, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA993001, MP95AA996002, MP95AA996001, MP95AA997001, MP95A9F6B001, MP95A9F76001, MP95AA983001, MP95AA988002, MP95AA98A001, MP95AA993002",
      "more_code_info": ""
    }
  ]
}