      "country": "United States",
      "city": "Tempe",
      "address_1": "1625 W 3rd St Bldg 1",
      "reason_for_recall": "Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.",
      "address_2": "",
      "product_quantity": "Total Devices=57,515 (U.S.=46,878; O.U.S.=10,637)",
      "code_info": "Product Code/GTIN:  VENEL16040 /00801741102431;  VENEL16060 /00801741102448;  VENEL16080 /00801741102455;  VENEL16100 /00801741102462;  VENEL16120 /00801741102479;  VENEL16140 /00801741102486;  VENEL16160 /00801741102493;  VENEL18040 /00801741102578;  VENEL18060 /00801741102585;  VENEL18080 /00801741102592;  VENEL18100 /00801741102608;  VENEL18120 /00801741102967;  VENEL18140 /00801741102974;  VENEL18160 /00801741102981;  VENEL20040 /00801741103063;  VENEL20060 /00801741103070;  VENEL20080 /00801741103087;  VENEL20100 /00801741103094;  VENEL20120 /00801741103100;  VENEL20140 /00801741103117;  VENEL20160 /00801741103124;  VENEM16040 /00801741102363;  VENEM16060 /00801741102370;  VENEM16080 /00801741102387;  VENEM16100 /00801741102394;  VENEM16120 /00801741102400;  VENEM16140 /00801741102417;  VENEM16160 /00801741102424;  VENEM18040 /00801741102509;  VENEM18060 /00801741102516;  VENEM18080 /00801741102523;  VENEM18100 /00801741102530;  VENEM18120 /00801741102547;  VENEM18140 /00801741102554;  VENEM18160 /00801741102561;  VENEM20040 /00801741102998;  VENEM20060 /00801741103001;  VENEM20080 /00801741103018;  VENEM20100 /00801741103025;  VENEM20120 /00801741103032;  VENEM20140 /00801741103049;  VENEM20160 /00801741103056;  VENUL16040 /00801741103629;  VENUL16060 /00801741103636;  VENUL16080 /00801741103643;  VENUL16100 /00801741103650;  VENUL16120 /00801741103667;  VENUL16140 /00801741103674;  VENUL16160 /00801741103681;  VENUL18040 /00801741103766;  VENUL18060 /00801741103773;  VENUL18080 /00801741103780;  VENUL18100 /00801741103797;  VENUL18120 /00801741103803;  VENUL18140 /00801741103810;  VENUL18160 /00801741103827;  VENUL20040 /00801741103902;  VENUL20060 /00801741103919;  VENUL20080 /00801741103926;  VENUL20100 /00801741103933;  VENUL20120 /00801741103940;  VENUL20140 /00801741103957;  VENUL20160 /00801741103964;  VENUM16040 /00801741103551;  VENUM16060 /00801741103568;  VENUM16080 /00801741103575;  VENUM16100 /00801741103582;  VENUM16120 /00801741103599;  VENUM16140 /00801741103605;  VENUM16160 /00801741103612;  VENUM18040 /00801741103698;  VENUM18060 /00801741103704;  VENUM18080 /00801741103711;  VENUM18100 /00801741103728;  VENUM18120 /00801741103735;  VENUM18140 /00801741103742;  VENUM18160 /00801741103759;  VENUM20040 /00801741103834;  VENUM20060 /00801741103841;  VENUM20080 /00801741103858;  VENUM20100 /00801741103865;  VENUM20120 /00801741103872;  VENUM20140 /00801741103889;  VENUM20160 /00801741103896;    U.S. Devices Serial Numbers:  ANEQ5007;  ANEQ5008;  ANEQ5009;  ANEQ5010;  ANEQ5011;  ANEQ5012;  ANEQ5013;  ANEQ5014;  ANEQ5015;  ANEQ5016;  ANEQ5017;  ANEQ5018;  ANEQ5019;  ANEQ5020;  ANEQ5021;  ANEQ5022;  ANER0482;  ANER0483;  ANER0484;  ANER0485;  ANER0486;  ANER0487;  ANER0488;  ANER0489;  ANER0494;  ANER0495;  ANER0496;  ANER0497;  ANER0498;  ANER0499;  ANER0500;  ANER0501;  ANER0502;  ANER0503;  ANER0504;  ANER0511;  ANER0512;  ANER0513;  ANER0514;  ANER0530;  ANER0531;  ANER1498;  ANER1499;  ANER1500;  ANER1501;  ANER1504;  ANER1505;  ANER1506;  ANER1531;  ANER1532;  ANER2801;  ANER2802;  ANER2803;  ANER2804;  ANER2805;  ANER2806;  ANER2807;  ANER2808;  ANER2809;  ANER2810;  ANER2811;  ANER2820;  ANER2821;  ANER2822;  ANER2823;  ANER2824;  ANER2825;  ANER2826;  ANER2827;  ANER2828;  ANER2829;  ANER2830;  ANER2831;  ANER2832;  ANER2833;  ANER2834;  ANER2835;  ANER2836;  ANER2839;  ANER2840;  ANER2842;  ANER2843;  ANER2844;  ANER2846;  ANER2847;  ANER2848;  ANER2849;  ANER2850;  ANER2856;  ANER2857;  ANER4199;  ANER4200;  ANER4201;  ANER4203;;  ANER4204;  ANER4206;  ANER4207;  ANER4208;  ANER4209;  ANER4212;  ANER4214;  ANER4215;  ANER4218;  ANER4220;  ANER4222;  ANER4223;  ANER4224;  ANER4226;  ANER4227;  ANER4229;  ANER4230;  ANER4231;  ANER4233;  ANER4234;  ANER4235;  ANER4237;  ANER4240;  ANER4241;  ANER4242;  ANER4243;  ANER5123;  ANER5124;  ANER5125;  ANER5129;  ANER5130;  ANER5131;  ANER5132;  ANER5133;  ANER5139;  ANER5140;  ANER5144;  ANER5145;  ANER5146;  ANER5147;  ANER5148;  ANER5155;  ANER5156;  ANER5157;  ANER5158;  ANER5162;  ANER5163;  ANER5164;  ANER5171;  ANER5172;  ANER5173;  ANER5174;  ANER5179;  ANER5180;  ANER5181;  ANES1051;  ANES1052;  ANES1053;  ANES1054;  ANES1055;  ANES1056;  ANES1057;  ANES1058;  ANES1059;  ANES2227;  ANES2228;  ANES2229;  ANES2230;  ANES2231;  ANES2232;  ANES2238;  ANES3025;  ANES3026;  ANES3027;  ANES3028;  ANES3029;  ANES3030;  ANES3031;  ANES3032;  ANES3033;  ANES3034;  ANES3035;  ANES3036;  ANES3037;  ANES3038;  ANES3041;  ANES3042;  ANES3043;  ANES3044;  ANES3046;  ANES3047;  ANES3048;  ANES3049;  ANES3053;  ANES3054;  ANES3055;  ANES3056;  ANES3059;  ANES3060;  ANES3065;  ANES3066;  ANES3068;  ANES4105;  ANES4106;  ANES4107;  ANES4108;  ANES4109;  ANES4110;  ANES4120;  ANES4121;  ANES4122;  ANES4123;  ANES4124;  ANES4125;  ANES4126;  ANES4127;  ANES4128;  ANET0562;  ANET0563;  ANET0564;  ANET0565;  ANET0575;  ANET0576;  ANET0577;  ANET0578;  ANET0579;  ANET0580;  ANET0581;  ANET0582;  ANET0583;  ANET1080;  ANET1081;  ANET1082;  ANET1083;  ANET1084;  ANET1471;  ANET1472;  ANET1473;  ANET1474;  ANET1475;  ANET1476;  ANET1477;  ANET1478;  ANET1479;  ANET1480;  ANET1481;  ANET1482;  ANET1483;  ANET1484;  ANET1489;  ANET1491;  ANET1492;  ANET1493;  ANET1494;  ANET1495;  ANET1496;  ANET1497;  ANET1499;  ANET1500;  ANET1501;  ANET1506;  ANET1507;  ANET2539;  ANET2545;  ANET2546;  ANET2547;  ANET2548;  ANET2549;  ANET2550;  ANET2551;  ANET2552;  ANET2553;  ANET2554;  ANET2555;  ANET2556;  ANET2557;  ANET2558;  ANET2559;  ANET2561;  ANET2562;  ANET2563;  ANET2564;  ANET2565;  ANET2566;  ANET2567;  ANET3804;  ANET3805;  ANET3806;  ANET3807;  ANET3808;  ANET3809;  ANET3810;  ANET3811;  ANET3812;  ANET3813;  ANET3824;  ANET3825;  ANET3826;  ANET3827;  ANET3828;  ANET3829;  ANET3830;  ANET3831;  ANET3832;  ANET3833;  ANET3834;  ANET3835;  ANET3836;  ANET3837;  ANET3838;  ANET3839;  ANET3840;  ANET3841;  ANET3843;  ANET3881;  ANET3882;  ANEU0745;  ANEU0746;  ANEU0747;  ANEU0755;  ANEU0756;  ANEU0757;  ANEU0758;  ANEU0759;  ANEU0760;  ANEU0761;  ANEU0762;  ANEU0763;  ANEU0764;  ANEU0765;  ANEU0766;  ANEU0767;  ANEU0768;  ANEU0769;  ANEU0770;  ANEU0771;  ANEU0773;  ANEU0774;  ANEU1596;  ANEU1597;  ANEU1598;  ANEU1599;  ANEU1600;  ANEU1611;  ANEU1612;  ANEU1613;  ANEU1614;  ANEU1615;  ANEU1616;  ANEU1617;  ANEU1618;  ANEU1619;  ANEU1620;  ANEU1621;  ANEU1622;  ANEU1623;  ANEU1624;  ANEU1625;  ANEU1626;  ANEU1628;  ANEU1629;  ANEU2437;  ANEU2438;  ANEU2439;  ANEU2440;  ANEU2441;  ANEU2449;  ANEU2450;  ANEU2451;  ANEU2452;  ANEU2453;  ANEU2454;  ANEU2455;  ANEU2456;  ANEU2457;  ANEU2458;  ANEU2459;  ANEU2460;  ANEU2461;  ANEU2462;  ANEU2463;  ANEU2464;  ANEU2465;  ANEU2466;  ANEU3280;  ANEU3281;  ANEU3282;  ANEU3283;  ANEU3295;  ANEU3296;  ANEU3297;  ANEU3298;  ANEU3299;  ANEU3300;  ANEU3301;  ANEU3302;  ANEU3305;  ANEU3307;  ANEU3308;  ANEU4229;  ANEU4230;  ANEU4235;  ANEU4236;  ANEU4237;  ANEU4243;  ANEU4244;  ANEU4245;  ANEU4250;  ANEU4251;  ANEU4254;  ANEU4255;  ANEU4258;  ANEU4259;  ANEU4262;  ANEU4263;  ANEU4266;  ANEU4267;  ANEV0588;  ANEV0590;  ANEV0592;  ANEV0593;  ANEV0595;  ANEV0596;  ANEV0601;  ANEV0602;  ANEV0605;  ANEV0607;  ANEV0608;  ANEV0610;  ANEV0612;  ANEV0613;  ANEV0616;  ANEV1723;  ANEV1724;  ANEV1725;  ANEV1726;  ANEV1727;  ANEV1728;  ANEV1729;  ANEV1730;  ANEV1731;  ANEV1732;  ANEV1733;  ANEV1734;  ANEV1735;  ANEV1736;  ANEV1746;  ANEV1747;  ANEV1748;  ANEV1749;  ANEV1750;  ANEV1751;  ANEV1752;  ANEV1753;  ANEV1754;  ANEV1755;  ANEV1756;  ANEV1757;  ANEV1758;  ANEV1759;  ANEV1760;  ANEV1761;  ANEV1762;  ANEV1763;  ANEV1764;  ANEV1765;  ANEV1766;  ANEV2620;  ANEV2624;  ANEV2628;  ANEV3073;  ANEV3074;  ANEV3075;  ANEV3076;  ANEV3077;  ANEV3086;  ANEV3087;  ANEV3088;  ANEV3089;  ANEV3090;  ANEV3091;  ANEV3092;  ANEV3093;  ANEV3094;  ANEV3095;  ANEV3096;  ANEV3097;  ANEV3498;  ANEV3499;  ANEV4031;  ANEV4032;  ANEV4033;  ANEV4034;  ANEV4035;  ANEV4036;  ANEV4037;  ANEV4038;  ANEV4039;  ANEV4040;  ANEV4041;  ANEV4042;  ANEV4052;  ANEV4053;  ANEV4054;  ANEV4055;  ANEV4056;  ANEV4057;  ANEV4058;  ANEV4059;  ANEV4060;  ANEV4061;  ANEV4062;  ANEV4063;  ANEV4064;  ANEV4065;  ANEV4066;  ANEV4067;  ANEV4068;  ANEV4069;  ANEV4070;  ANEV4072;  ANEV4073;  ANEV4075;  ANEW0556;  ANEW0557;  ANEW0558;  ANEW0559;  ANEW0575;  ANEW0576;  ANEW0577;  ANEW0578;  ANEW0579;  ANEW0580;  ANEW0581;  ANEW0582;  ANEW0583;  ANEW0584;  ANEW0585;  ANEW0586;  ANEW0587;  ANEW0588;  ANEW0589;  ANEW0590;  ANEW0591;  ANEW0592;  ANEW0593;  ANEW0594;  ANEW0595;  ANEW0598;  ANEW0601;  ANEW0602;  ANEW1520;  ANEW1536;  ANEW1537;  ANEW1538;  ANEW1539;  ANEW1580;  ANEW1581;  ANEW1582;  ANEW1583;  ANEW1584;  ANEW1585;  ANEW2012;  ANEW2013;  ANEW2014;  ANEW2015;  ANEW2016;  ANEW2017;  ANEW2018;  ANEW2019;  ANEW2020;  ANEW2021;  ANEW2022;  ANEW2023;  ANEW2024;  ANEW2025;  ANEW2026;  ANEW2027;  ANEW2037;  ANEW2038;  ANEW2039;  ANEW2040;  ANEW2041;  ANEW2042;  ANEW2043;  ANEW2044;  ANEW2238;  ANEW2240;  ANEW2531;  ANEW2532;  ANEW2768;  ANEW2769;  ANEW2770;  ANEW2771;  ANEW2775;  ANEW3386;  ANEW3387;  ANEW3388;  ANEW3389;  ANEW3390;  ANEW3391;  ANEW3392;  ANEW3393;  ANEW3401;  ANEW3402;  ANEW3403;  ANEW3404;  ANEW3405;  ANEW3406;  ANEW3407;  ANEW3408;  ANEW3424;  ANEW3428;  ANEW3429;  ANEW3431;  ANEW3432;  ANEW3433;  ANEW3434;  ANEW3437;  ANEW4235;  ANEW4236;  ANEW4237;  ANEW4238;  ANEW4239;  ANEW4240;  ANEW4241;  ANEW4255;  ANEW4256;  ANEW4257;  ANEW4258;  ANEW4259;  ANEW4262;  ANEW4263;  ANEW4264;  ANEW4266;  ANEW4267;  ANEW4268;  ANEW4269;  ANEW4273;  ANEW4274;  ANEW4275;  ANEW4277;  ANEW4278;  ANEW4279;  ANEW4281;  ANEW4283;  ANEW4284;  ANEW4285;  ANEW4286;  ANEW4287;  ANEW4288;  ANEW4289;  ANEW4290;  ANEW4296;  ANEW4299;  ANEX0681;  ANEX0682;  ANEX0683;  ANEX0684;  ANEX0685;  ANEX0686;  ANEX0687;  ANEX0688;  ANEX0689;  ANEX0690;  ANEX0691;  ANEX0692;  ANEX0695;  ANEX0696;  ANEX0699;  ANEX0700;  ANEX0702;  ANEX0703;  ANEX0706;  ANEX0708;  ANEX0710;  ANEX0715;  ANEX0716;  ANEX0717;  ANEX0719;  ANEX0721;  ANEX0722;  ANEX0723;  ANEX0724;  ANEX0726;  ANEX0727;  ANEX0728;  ANEX0730;  ANEX0731;  ANEX0733;  ANEX0734;  ANEX0737;  ANEX1712;  ANEX1713;  ANEX1714;  ANEX1715;  ANEX1716;  ANEX1717;  ANEX1720;  ANEX1721;  ANEX1722;  ANEX1723;  ANEX1724;  ANEX1725;  ANEX1726;  ANEX1727;  ANEX1730;  ANEX1734;  ANEX1736;  ANEX1739;  ANEX1740;  ANEX1742;  ANEX1748;  ANEX1756;  ANEX3098;  ANEX3099;  ANEX3100;  ANEX3101;  ANEX3104;  ANEX3105;  ANEX3112;  ANEX3113;  ANEX3115;  ANEX3116;  ANEX3118;  ANEX3122;  ANEX3124;  ANEX3126;  ANEX3127;  ANEX3129;  ANEX3130;  ANEX3132;  ANEX3138;  ANEX3139;  ANEX3141;  ANEX3142;  ANEX3143;  ANEX3146;  ANEX3150;  ANEX3152;  ANEX4721;  ANEX4722;  ANEX4725;  ANEX4726;  ANEX4731;  ANEX4732;  ANEX4733;  ANEX4735;  ANEX4736;  ANEX4738;  ANEX4742;  ANEX4746;  ANEX4747;  ANEX4748;  ANEX4750;  ANEX4752;  ANEX4755;  ANEX4756;  ANEX4758;  ANEX4759;  ANEX4760;  ANEX4763;  ANEX4765;  ANEX4766;  ANEX4767;  ANEX4768;  ANEX4769;  ANEX4770;  ANEX4772;  ANEX4775;  ANEX4777;  ANEX4778;  ANEX4781;  ANEX6025;  ANEX6026;  ANEX6027;  ANEX6028;  ANEX6032;  ANEX6033;  ANEX6035;  ANEX6037;  ANEX6038;  ANEX6039;  ANEX6045;  ANEX6048;  ANEX6052;  ANEX6053;  ANEX6054;  ANEX6055;  ANEX6059;  ANEX6060;  ANEX6061;  ANEX6062;  ANEX6064;  ANEX6067;  ANEX6070;  ANEX6071;  ANEX6072;  ANEY1229;  ANEY1230;  ANEY1231;  ANEY1232;  ANEY1233;  ANEY1234;  ANEY1235;  ANEY1236;  ANEY1237;  ANEY1240;  ANEY1241;  ANEY1242;  ANEY1243;  ANEY1244;  ANEY1248;  ANEY1249;  ANEY1250;  ANEY1254;  ANEY1255;  ANEY1256;  ANEY1272;  ANEY1273;  ANEY1274;  ANEY1275;  ANEY1276;  ANEY1277;  ANEY1282;  ANEY1283;  ANEY1284;  ANEY1285;  ANEY2037;  ANEY2038;  ANEY2039;  ANEY2040;  ANEY2041;  ANEY2042;  ANEY2043;  ANEY2044;  ANEY2045;  ANEY2046;  ANEY2047;  ANEY2048;  ANEY2064;  ANEY2065;  ANEY2066;  ANEY2067;  ANEY2068;  ANEY2069;  ANEY2070;  ANEY2071;  ANEY2072;  ANEY2073;  ANEY2074;  ANEY2075;  ANEY2076;  ANEY2077;  ANEY2078;  ANEY2079;  ANEY2080;  ANEY2081;  ANEY3217;  ANEY3218;  ANEY3225;  ANEY3226;  ANEY3231;  ANEY3232;  ANEY3235;  ANEY3236;  ANEY3237;  ANEY3238;  ANEY3239;  ANEY3240;  ANEY3242;  ANEY3244;  ANEY3246;  ANEY3247;  ANEY3248;  ANEY3250;  ANEY3251;  ANEY3252;  ANEY3253;  ANEY3254;  ANEY3260;  ANEY4344;  ANEY4345;  ANEY4347;  ANEY4349;  ANEY4350;  ANEY4351;  ANEY4352;  ANEY4353;  ANEY4359;  ANEY4360;  ANEY4361;  ANEY4362;  ANEY4560;  ANEZ0937;  ANEZ0944;  ANEZ0945;  ANEZ0954;  ANEZ0955;  ANEZ0956;  ANEZ0960;  ANEZ0961;  ANEZ0962;  ANEZ0963;  ANEZ0964;  ANEZ0965;  ANEZ0966;  ANEZ0968;  ANEZ0969;  ANEZ0970;  ANEZ0971;  ANEZ0975;  ANEZ0976;  ANEZ0977;  ANEZ0978;  ANEZ0979;  ANEZ0981;  ANEZ0986;  ANEZ0987;  ANEZ0988;  ANEZ0989;  ANEZ0990;  ANEZ0991;  ANEZ0992;  ANEZ0993;  ANEZ0994;  ANEZ0995;  ANEZ0998;  ANEZ0999;  ANEZ2109;  ANEZ2110;  ANEZ2111;  ANEZ2118;  ANEZ2119;  ANEZ2120;  ANEZ2131;  ANEZ2132;  ANEZ2133;  ANEZ2134;  ANEZ2135;  ANEZ2136;  ANEZ2147;  ANEZ2148;  ANEZ2149;  ANEZ2150;  ANEZ2151;  ANEZ2152;  ANEZ2153;  ANEZ2154;  ANEZ2155;  ANEZ2156;  ANEZ2157;  ANEZ2158;  ANEZ2159;  ANEZ2160;  ANEZ2161;  ANEZ2162;  ANEZ2163;  ANEZ2164;  ANEZ2165;  ANEZ3442;  ANEZ3443;  ANEZ3444;  ANEZ3445;  ANEZ3446;  ANEZ3447;  ANEZ3448;  ANEZ3449;  ANEZ3450;  ANEZ3451;  ANEZ3452;  ANEZ3464;  ANEZ3465;  ANEZ3466;  ANEZ3467;  ANEZ3468;  ANEZ3469;  ANEZ3470;  ANEZ3471;  ANEZ3472;  ANEZ3473;  ANEZ3474;  ANEZ3475;  ANEZ3476;  ANEZ3477;  ANEZ3478;  ANEZ3479;  ANEZ3480;  ANEZ3481;  ANEZ3482;  ANEZ3483;  ANEZ3484;  ANEZ3485;  ANEZ3486;  ANEZ3487;  ANEZ3488;  ANEZ3489;  ANEZ3490;  ANEZ3491;  ANEZ3492;  ANEZ3493;  ANEZ3494;  ANEZ3495;  ANEZ3496;  ANEZ3497;  ANEZ3498;  ANEZ3500;  ANEZ3503;  ANEZ3504;  ANEZ3505;  ANEZ3506;  ANEZ3840;  ANEZ3841;  ANEZ3935;  ANEZ3936;  ANEZ3940;  ANEZ3941;  ANEZ3942;  ANEZ3943;  ANEZ3944;  ANEZ3945;  ANEZ3946;  ANEZ3947;  ANEZ3948;  ANEZ3950;  ANEZ3951;  ANEZ3953;  ANEZ3954;  ANEZ3955;  ANFN1454;  ANFN1455;  ANFN1456;  ANFN1457;  ANFN1458;  ANFN1459;  ANFN1460;  ANFN1475;  ANFN1478;  ANFN1480;  ANFN1481;  ANFN1489;  ANFN1491;  ANFN2405;  ANFN2800;  ANFN2801;  ANFN2802;  ANFN2803;  ANFN2804;  ANFN2805;  ANFN2806;  ANFN2808;  ANFN2810;  ANFN2812;  ANFN2814;  ANFN2816;  ANFN2818;  ANFN2819;  ANFN2820;  ANFN2823;  ANFN2829;  ANFN2831;  ANFN2832;  ANFN2852;  ANFN2853;  ANFN2854;  ANFN2855;  ANFN3996;  ANFN3997;  ANFN3998;  ANFN4007;  ANFN4008;  ANFP0892;  ANFP0893;  ANFP0894;  ANFP0895;  ANFP0901;  ANFP0902;  ANFP0903;  ANFP0904;  ANFP0905;  ANFP0906;  ANFP0907;  ANFP0917;  ANFP0918;  ANFP0919;  ANFP0920;  ANFP0922;  ANFP0925;  ANFP0926;  ANFP0927;  ANFP0928;  ANFP0929;  ANFP0930;  ANFP1520;  ANFP2643;  ANDS0737;  ANDS0738;  ANDS1294;  ANDS0727;  ANDS0739;  ANDS0726;  ANDS0747;  ANDS0723;  ANDS0745;  ANDS0741;  ANDS0724;  ANDS0735;  ANDS1293;  ANDS0734;  ANDS0749;  ANDS0746;  ANDS0725;  ANDS0742;  ANDS0733;  ANDS0744;  ANDS0743;  ANDS0736;  ANDS0748;  ANDS1281;  ANDS1327;  ANDS1273;  ANDS1314;  ANDS1292;  ANDS1291;  ANDS1279;  ANDS1309;  ANDS1299;  ANDS1274;  ANDS1280;  ANDS1301;  ANDS0728;  ANDS1295;  ANDS1300;  ANDS1306;  ANDS1286;  ANDS1289;  ANDS0740;  ANDS0750;  ANDS1307;  ANDS1277;  ANDS1287;  ANDS1305;  ANDS1278;  ANDS1288;  ANDT1236;  ANDS1319;  ANDS1320;  ANDT1241;  ANDT1244;  ANDT1235;  ANDS1313;  ANDS1304;  ANDS1318;  ANDT1232;  ANDT1234;  ANDS2456;  ANDS1328;  ANDT1243;  ANDS1326;  ANDS1333;  ANDS1330;  ANDS1332;  ANDS2467;  ANDS2453;  ANDS2466;  ANDT1245;  ANDS2464;  ANDS1331;  ANDT1242;  ANDT1240;  ANDS2457;  ANDS2465;  ANDS2468;  ANDS1322;  ANDS1334;  ANDS2458;  ANDT1233;  ANDS4811;  ANDS2474;  ANDS2484;  ANDS2486;  ANDS2472;  ANDS2473;  ANDS4859;  ANDS4877;  ANDS3955;  ANDS3958;  ANDS3961;  ANDS2485;  ANDS2488;  ANDS2478;  ANDS3956;  ANDS3962;  ANDS3965;  ANDS3964;  ANDS4878;  ANDS3959;  ANDS4792;  ANDS4860;  ANDS2475;  ANDS4847;  ANDS4865;  ANDS2487;  ANDS3957;  ANDT2451;  ANDS4816;  ANDS4800;  ANDS4866;  ANDS3967;  ANDS4867;  ANDS2483;  ANDS2455;  ANDS4879;  ANDS4856;  ANDS3966;  ANDS4849;  ANDS3968;  ANDS1317;  ANDS4857;  ANDS4803;  ANDT2452;  ANDT2449;  ANDS2459;  ANDS4797;  ANDS3969;  ANDS2477;  ANDS2476;  ANDS3972;  ANDS3963;  ANDS4819;  ANDT2453;  ANDT2450;  ANDS3960;  ANDS4868;  ANDS4848;  ANDS4805;  ANDS4870;  ANDS4872;  ANDS4815;  ANDS4862;  ANDS4801;  ANDS4843;  ANDS4806;  ANDS4795;  ANDS4844;  ANDS4850;  ANDS4818;  ANDS4796;  ANDS4820;  ANDS4863;  ANDS4802;  ANDS4798;  ANDS4810;  ANDS4809;  ANDS4846;  ANDS4858;  ANDS4793;  ANDS4874;  ANDS4861;  ANDS4873;  ANDS4794;  ANDS4812;  ANDS4814;  ANDS4804;  ANDS4799;  ANDS4807;  ANDS4875;  ANDS4871;  ANDS4821;  ANDT1069;  ANDS4813;  ANDT1089;  ANDS4869;  ANDT1085;  ANDT1065;  ANDS4845;  ANDS4864;  ANDT1084;  ANDT1080;  ANDS4808;  ANDT1070;  ANDT1077;  ANDT1064;  ANDT1067;  ANDS4817;  ANDT1100;  ANDT1098;  ANDT1079;  ANDT1092;  ANDT1096;  ANDT1066;  ANDT1061;  ANDT1088;  ANDT1086;  ANDT1099;  ANDT1082;  ANDT1728;  ANDT1688;  ANDT1090;  ANDT1097;  ANDT1093;  ANDT1709;  ANDT1707;  ANDS4876;  ANDT1078;  ANDT1691;  ANDT1692;  ANDT1698;  ANDT1689;  ANDT1722;  ANDT1063;  ANDT1094;  ANDT1071;  ANDT1712;  ANDT1062;  ANDT1693;  ANDT1091;  ANDT1717;  ANDT1718;  ANDS4822;  ANDT1695;  ANDT1711;  ANDT1701;  ANDT1694;  ANDT1706;  ANDT1705;  ANDT1697;  ANDT1083;  ANDT1095;  ANDT1068;  ANDT1081;  ANDT1699;  ANDU2249;  ANDT1710;  ANDT1696;  ANDT1087;  ANDT1720;  ANDT1727;  ANDT1723;  ANDT1072;  ANDT1714;  ANDT2898;  ANDT2924;  ANDT2933;  ANDT2918;  ANDT2939;  ANDT1704;  ANDT1700;  ANDT2904;  ANDT1726;  ANDT1703;  ANDT2938;  ANDT2920;  ANDT2919;  ANDT2900;  ANDT2935;  ANDT2903;  ANDT2929;  ANDT2942;  ANDT2934;  ANDT1702;  ANDT2928;  ANDT2899;  ANDT1715;  ANDT2897;  ANDT2925;  ANDT2930;  ANDT2932;  ANDT2921;  ANDT2923;  ANDT2940;  ANDU0143;  ANDU3354;  ANDT1719;  ANDT2922;  ANDT2931;  ANDU0140;  ANDU0105;  ANDT1724;  ANDU0142;  ANDU0130;  ANDT2901;  ANDU0108;  ANDU0115;  ANDT2902;  ANDU0107;  ANDU0111;  ANDU0145;  ANDU0113;  ANDU0110;  ANDT2927;  ANDT2936;  ANDT2926;  ANDU0135;  ANDT1725;  ANDT1716;  ANDU0139;  ANDU0118;  ANDT2941;  ANDU0131;  ANDU0133;  ANDU0104;  ANDU0112;  ANDT1708;  ANDU0120;  ANDU0128;  ANDU0137;  ANDU0117;  ANDT2937;  ANDU0116;  ANDU0141;  ANDU0136;  ANDU0138;  ANDU0102;  ANDU0103;  ANDU0114;  ANDU0106;  ANDU0144;  ANDU0132;  ANDU0129;  ANDU0101;  ANDU0109;  ANDU1122;  ANDU1157;  ANDU1152;  ANDU1151;  ANDU1156;  ANDU1131;  ANDU1121;  ANDU1116;  ANDU1141;  ANDU1146;  ANDU1148;  ANDU1143;  ANDU0134;  ANDU1142;  ANDU1118;  ANDU1132;  ANDU1153;  ANDU1137;  ANDU1150;  ANDU1133;  ANDU1154;  ANDU0119;  ANDU1147;  ANDU1120;  ANDU1123;  ANDU1117;  ANDU1136;  ANDU1134;  ANDU1140;  ANDU1130;  ANDU1144;  ANDU1145;  ANDU1155;  ANDU1139;  ANDU1138;  ANDU1869;  ANDU1858;  ANDU1876;  ANDU1887;  ANDU1119;  ANDV0715;  ANDU1875;  ANDU1874;  ANDU1871;  ANDU1880;  ANDU1888;  ANDU1135;  ANDU1891;  ANDU1896;  ANDU1893;  ANDU1879;  ANDU1909;  ANDU1910;  ANDU1149;  ANDU1865;  ANDU1859;  ANDU1882;  ANDU1892;  ANDU1911;  ANDU1897;  ANDU1899;  ANDU1903;  ANDU1860;  ANDU1902;  ANDV0719;  ANDU3095;  ANDU3106;  ANDU3107;  ANDV0717;  ANDV1986;  ANDV0718;  ANDV1983;  ANDU3108;  ANDU3100;  ANDU3119;  ANDU3131;  ANDU3134;  ANDV0720;  ANDV0721;  ANDU3148;  ANDU3143;  ANDU1898;  ANDU3102;  ANDV0716;  ANDU3147;  ANDU3114;  ANDV2135;  ANDV2137;  ANDU3118;  ANDU3121;  ANDU3120;  ANDU1904;  ANDV1987;  ANDU3133;  ANDU3132;  ANDU3112;  ANDU3130;  ANDV2725;  ANDU3149;  ANDV1982;  ANDU4031;  ANDV0723;  ANDU4055;  ANDU4032;  ANDU4042;  ANDU4053;  ANDU4059;  ANDV3044;  ANDU4038;  ANDU4036;  ANDU4046;  ANDU4068;  ANDU4078;  ANDU4066;  ANDU4062;  ANDU4080;  ANDU4045;  ANDU4082;  ANDU4047;  ANDV0726;  ANDU4057;  ANDV0727;  ANDU4069;  ANDU4079;  ANDU4072;  ANDU4063;  ANDU4067;  ANDV3957;  ANDU4065;  ANDV3779;  ANDU4070;  ANDU4064;  ANDU4052;  ANDV3782;  ANDV0260;  ANDV0258;  ANDV0261;  ANDV0271;  ANDV0259;  ANDV0256;  ANDU4058;  ANDV0272;  ANDV0269;  ANDV0274;  ANDU3150;  ANDV0262;  ANDV0296;  ANDV0295;  ANDV0293;  ANDV0292;  ANDV0290;  ANDV0285;  ANDU4049;  ANDV0294;  ANDV0283;  ANDV0280;  ANDV0278;  ANDV0273;  ANDV0281;  ANDV0282;  ANDV0284;  ANDV0270;  ANDV0291;  ANDV0289;  ANDV0268;  ANDV0308;  ANDV0307;  ANDV0306;  ANDV0305;  ANDV0304;  ANDV0303;  ANDV0309;  ANDV0279;  ANDV1219;  ANDV1218;  ANDV1217;  ANDV1216;  ANDV1214;  ANDV1237;  ANDV1231;  ANDV1229;  ANDV1220;  ANDV1223;  ANDV1226;  ANDV1230;  ANDV1263;  ANDV1225;  ANDV1228;  ANDV1266;  ANDV1224;  ANDV1239;  ANDV1235;  ANDV1236;  ANDV1240;  ANDV1253;  ANDV1249;  ANDV1252;  ANDV0297;  ANDV1254;  ANDV1243;  ANDV1267;  ANDV2548;  ANDV2554;  ANDV2528;  ANDV2530;  ANDV1265;  ANDV1264;  ANDV2533;  ANDV2535;  ANDV2534;  ANDV2545;  ANDV2549;  ANDV2553;  ANDV2550;  ANDV2555;  ANDV2529;  ANDV2559;  ANDV2531;  ANDV2546;  ANDV2556;  ANDV2536;  ANDV2526;  ANDV2551;  ANDV2537;  ANDV3618;  ANDV3622;  ANDV3579;  ANDV2557;  ANDV3627;  ANDV3587;  ANDV3623;  ANDV3614;  ANDV2552;  ANDV3594;  ANDV3590;  ANDV3592;  ANDV3582;  ANDV3583;  ANDV3615;  ANDV3619;  ANDV3588;  ANDV3591;  ANDV3626;  ANDV3586;  ANDV2558;  ANDV3595;  ANDV3621;  ANDV3617;  ANDV3593;  ANDV3584;  ANDV3597;  ANDV3581;  ANDV3599;  ANDV3585;  ANDV3596;  ANDV3624;  ANDV3628;  ANDV2532;  ANDV2527;  ANDV3598;  ANDV3589;  ANDV3625;  ANDV3613;  ANDV3580;  ANDV3620;  ANDV3616;  ANDV3629;  ANDV4582;  ANDV4579;  ANDV4578;  ANDV4590;  ANDV4558;  ANDV4563;  ANDV4562;  ANDV4595;  ANDV4594;  ANDV4583;  ANDV4591;  ANDV4559;  ANDV4560;  ANDV4586;  ANDV4587;  ANDV4588;  ANDV4584;  ANDV4589;  ANDV4585;  ANDW1510;  ANDV4580;  ANDV4581;  ANDV4577;  ANDW1502;  ANDW1508;  ANDW1509;  ANDV4561;  ANDV4592;  ANDV4576;  ANDV4593;  ANDV4564;  ANDV1241;  ANDV4565;  ANDW1498;  ANDW1504;  ANDW1503;  ANDW1500;  ANDW1499;  ANDW1501;  ANDW1507;  ANDV3630;  ANDW1506;  ANDW1505;  ANDW1511;  ANDV2547;  ANDW2064;  ANDW2055;  ANDW2067;  ANDW2059;  ANDW2060;  ANDW2061;  ANDW2062;  ANDW2070;  ANDW2058;  ANDW2068;  ANDW2057;  ANDW2072;  ANDW2069;  ANDW2071;  ANDX1320;  ANDX1318;  ANDX1317;  ANDW2056;  ANDX1319;  ANDW2063;  ANDX1392;  ANDW2066;  ANDX1395;  ANDX1390;  ANDX1397;  ANDX1391;  ANDW2073;  ANDX1396;  ANDX1398;  ANDX1393;  ANDX1394;  ANDW2065;  ANDX1400;  ANDX1447;  ANDW3631;  ANDW3636;  ANDW3629;  ANDW3637;  ANDX1442;  ANDX1441;  ANDX1446;  ANDX1456;  ANDX1452;  ANDW3638;  ANDW3633;  ANDW3632;  ANDW3628;  ANDX1458;  ANDX1457;  ANDX1399;  ANDX1455;  ANDX2787;  ANDX1453;  ANDX1443;  ANDX1448;  ANDX1450;  ANDX1449;  ANDW3639;  ANDX1440;  ANDX1451;  ANDW3635;  ANDX1445;  ANDX1444;  ANDX1459;  ANDW3630;  ANDW4349;  ANDX1454;  ANDW4368;  ANDW4365;  ANDW4374;  ANDW4385;  ANDW4382;  ANDW4371;  ANDW4401;  ANDX6141;  ANDW4403;  ANDX6142;  ANDW4391;  ANDW4399;  ANDW4388;  ANDW4386;  ANDW4384;  ANDW4346;  ANDW4421;  ANDX6284;  ANDW4420;  ANDW3634;  ANDW4414;  ANDW4417;  ANDX0706;  ANDX0710;  ANDX0714;  ANDX0709;  ANDX0721;  ANDX0717;  ANDW4408;  ANDX0722;  ANDW4405;  ANDX0726;  ANDX0733;  ANDX0737;  ANDX0736;  ANDX0745;  ANDX0741;  ANDX0742;  ANDX0734;  ANDX0738;  ANDX0729;  ANDW4422;  ANDX0748;  ANDX0713;  ANDX0718;  ANDX0730;  ANDX0746;  ANDX0749;  ANDX2201;  ANDX0757;  ANDX0767;  ANDX0763;  ANDX0762;  ANDX2188;  ANDX2180;  ANDX2208;  ANDX2205;  ANDX2203;  ANDX2200;  ANDX2204;  ANDX0754;  ANDX2184;  ANDX0756;  ANDX2207;  ANDX2183;  ANDX2187;  ANDX2199;  ANDX2211;  ANDX0743;  ANDX2185;  ANDX2181;  ANDX2189;  ANDX2209;  ANDX2186;  ANDX2202;  ANDX2206;  ANDX0758;  ANDX2182;  ANDX2210;  ANDX2198;  ANDX3120;  ANDX3112;  ANDX3121;  ANDX3097;  ANDX3113;  ANDX3117;  ANDX3092;  ANDX3096;  ANDX3095;  ANDX3116;  ANDX3115;  ANDX3111;  ANDX3093;  ANDX3118;  ANDX3114;  ANDX3094;  ANDX3119;  ANDX5331;  ANDX5335;  ANDX5330;  ANDX5340;  ANDX5345;  ANDX5348;  ANDX5346;  ANDX5342;  ANDX5341;  ANDX3098;  ANDX3122;  ANDX5339;  ANDX5349;  ANDX5347;  ANDX5336;  ANDX5338;  ANDX5343;  ANDX5344;  ANDX5395;  ANDX5405;  ANDX5411;  ANDX5337;  ANDX5414;  ANDX5406;  ANDX5396;  ANDX5415;  ANDX5407;  ANDX5412;  ANDX5410;  ANDX5398;  ANDX5416;  ANDX5397;  ANDX5417;  ANDX5413;  ANDX5408;  ANDZ0246;  ANDX5409;  ANDZ0415;  ANDZ0245;  ANDX5418;  ANDX5394;  ANDZ0413;  ANDZ0416;  ANDZ0414;  ANDY1264;  ANDY1265;  ANDY1262;  ANDY1243;  ANDY1252;  ANDY1259;  ANDY1255;  ANDY1245;  ANDY1267;  ANDY1242;  ANDY1244;  ANDY1258;  ANDY1269;  ANDY1263;  ANDY1268;  ANDY1257;  ANDZ0785;  ANDY1256;  ANDZ0788;  ANDZ0786;  ANDY1266;  ANDY1254;  ANDY1240;  ANDZ0789;  ANDY1270;  ANDY1253;  ANDY1260;  ANDY1261;  ANDY1241;  ANDZ0787;  ANDY1246;  ANDY2483;  ANDY2503;  ANDY2482;  ANDY2484;  ANDY2478;  ANDY2493;  ANDY2479;  ANDY2477;  ANDY2499;  ANDY2498;  ANDY2497;  ANDY2476;  ANDY2502;  ANDY2501;  ANDY2481;  ANDY2475;  ANDY2496;  ANDY2504;  ANDZ1397;  ANDZ1390;  ANDY2480;  ANDZ1369;  ANDZ1361;  ANDY2494;  ANDY2495;  ANDZ1402;  ANDZ1406;  ANDZ1395;  ANDY2505;  ANDY2500;  ANDZ1407;  ANDZ2799;  ANDZ1366;  ANDZ1400;  ANDZ2801;  ANDZ2802;  ANDZ2800;  ANDZ1404;  ANDY3334;  ANDY3337;  ANDY3317;  ANDY3339;  ANDY3319;  ANDY3344;  ANDY3343;  ANDY3348;  ANDY3322;  ANDY2485;  ANDY3347;  ANDY3342;  ANDY3318;  ANDY3315;  ANDY3340;  ANDY3320;  ANDY3345;  ANDY3338;  ANDY3335;  ANDZ1272;  ANDY3341;  ANEN0317;  ANDZ1280;  ANDZ1276;  ANDY3336;  ANDY3346;  ANDZ1268;  ANDY3316;  ANDY3321;  ANDZ1279;  ANDZ1275;  ANDZ1245;  ANDZ1265;  ANDZ1246;  ANDZ1270;  ANDZ1273;  ANDZ1249;  ANDZ1253;  ANDZ1251;  ANEN0316;  ANDY3333;  ANDZ1247;  ANDZ1267;  ANDZ1269;  ANDZ1248;  ANDZ1274;  ANDZ1278;  ANDZ1266;  ANDZ1250;  ANDZ1277;  ANDZ1271;  ANDZ1281;  ANEN0315;  ANDZ1252;  ANDZ1401;  ANDZ1392;  ANDZ1371;  ANDZ1370;  ANDZ1391;  ANDZ1398;  ANDZ1362;  ANDZ1399;  ANDZ1387;  ANDZ1388;  ANDZ1368;  ANDZ1396;  ANDZ1386;  ANDZ1367;  ANEN1835;  ANDZ1364;  ANDZ1363;  ANEN1833;  ANDZ1408;  ANDZ1360;  ANDZ1365;  ANEN1832;  ANEN1831;  ANEN1830;  ANDZ1405;  ANDZ1403;  ANDZ1393;  ANEN1834;  ANDZ1389;  ANEN1829;  ANDZ2314;  ANDZ2351;  ANDZ2346;  ANDZ2349;  ANDZ2321;  ANDZ2324;  ANDZ2319;  ANDZ2335;  ANDZ2340;  ANDZ2320;  ANDZ2345;  ANDZ2315;  ANDZ2344;  ANDZ2347;  ANDZ2338;  ANDZ2318;  ANDZ2323;  ANDZ2336;  ANDZ2334;  ANEN3914;  ANEN3913;  ANEN3917;  ANDZ2322;  ANDZ2350;  ANDZ2343;  ANDZ2316;  ANEN3915;  ANDZ2317;  ANDZ2341;  ANDZ2337;  ANDZ2333;  ANDZ2342;  ANDZ3850;  ANDZ3863;  ANDZ3862;  ANDZ3843;  ANDZ3857;  ANDZ3872;  ANDZ3882;  ANDZ3874;  ANDZ3868;  ANDZ3853;  ANDZ3852;  ANDZ3881;  ANDZ3848;  ANDZ3847;  ANDZ3858;  ANDZ2348;  ANDZ3879;  ANDZ3855;  ANDZ1394;  ANDZ3859;  ANDZ2339;  ANDZ3875;  ANDZ3876;  ANDZ3878;  ANDZ3856;  ANDZ3880;  ANDZ3849;  ANDZ3845;  ANDZ3873;  ANDZ3864;  ANDZ3869;  ANDZ3851;  ANEN1091;  ANEN1120;  ANEN1101;  ANEN1106;  ANEN1100;  ANEN1092;  ANEN1105;  ANEN1102;  ANEN1099;  ANDZ3841;  ANEN1119;  ANEN1104;  ANEN1103;  ANEN1118;  ANDZ3867;  ANDZ3854;  ANEN2159;  ANEN2174;  ANEN2173;  ANEN2163;  ANEN2183;  ANEN2184;  ANEN2160;  ANEN2158;  ANEN2165;  ANEN2177;  ANEN2181;  ANEN2176;  ANEN2156;  ANEN1098;  ANEN2157;  ANEN2192;  ANEN3419;  ANEN3410;  ANEN3435;  ANEN3424;  ANEN3432;  ANEN3411;  ANEN2164;  ANEN3439;  ANEN3407;  ANEN3427;  ANEN3409;  ANEN3426;  ANEN3428;  ANEP4059;  ANEN3408;  ANEN2191;  ANEN3429;  ANEP4058;  ANEN4920;  ANEN4908;  ANEN4918;  ANEN4924;  ANEN4915;  ANEN4935;  ANEN4934;  ANEN4921;  ANEN4906;  ANEP0894;  ANEN4922;  ANEN4931;  ANEN2166;  ANEN3434;  ANEN3440;  ANEN4927;  ANEP0886;  ANEP0920;  ANEN3442;  ANEN4917;  ANEQ0212;  ANEP0901;  ANEN4909;  ANEN4933;  ANEN4919;  ANEN4911;  ANEP0885;  ANEP0895;  ANEP0900;  ANEP0913;  ANEN4932;  ANEP0889;  ANEP0888;  ANEN4929;  ANEP0911;  ANEP2448;  ANEP0887;  ANEP2445;  ANEP0897;  ANEP2471;  ANEP2451;  ANEP2444;  ANEP2481;  ANEP0908;  ANEP2473;  ANEP2488;  ANEP2478;  ANEP2460;  ANEP2464;  ANEP2463;  ANEP2480;  ANEP2486;  ANEP2489;  ANEP0917;  ANEP2483;  ANEP2477;  ANEP2476;  ANEP2475;  ANEP2456;  ANEP0890;  ANEP0910;  ANEP0916;  ANEP2479;  ANEQ3073;  ANEP3978;  ANEP3948;  ANEQ3074;  ANEP2469;  ANEP2449;  ANEP3975;  ANEP2452;  ANEP2447;  ANEP3987;  ANEP3977;  ANEP3947;  ANEP3983;  ANEP3981;  ANEP2457;  ANEP3946;  ANEP3951;  ANEP3970;  ANEP0903;  ANEP3960;  ANEP3980;  ANEP2446;  ANEP3961;  ANEP3971;  ANEP3945;  ANEP3950;  ANEP3976;  ANEP3943;  ANEP3955;  ANEP5339;  ANEP3952;  ANEP5379;  ANEP5353;  ANEP5376;  ANEP5369;  ANEP5348;  ANEP5388;  ANEP5349;  ANEP5383;  ANEQ4388;  ANEP5371;  ANEP5355;  ANEP5344;  ANEP5384;  ANEP3959;  ANEP5354;  ANEP3954;  ANEP5347;  ANEP3956;  ANEQ4751;  ANEP3953;  ANEP5382;  ANEP5336;  ANEP5337;  ANEP3958;  ANEP5342;  ANEP5345;  ANEP5351;  ANEP5377;  ANEP2453;  ANEQ4750;  ANEP5387;  ANEQ4746;  ANEP3965;  ANEP5350;  ANEP5343;  ANEP5346;  ANEP5372;  ANEQ1394;  ANEQ1413;  ANEQ1417;  ANEQ1391;  ANEQ1412;  ANEQ1389;  ANEQ1416;  ANEQ1414;  ANEQ1388;  ANEQ1386;  ANEQ1387;  ANEQ1385;  ANEQ2772;  ANEQ2765;  ANEQ2771;  ANEQ2760;  ANEQ2763;  ANEQ2769;  ANEQ1392;  ANEQ2762;  ANEQ2761;  ANEQ1395;  ANEQ1390;  ANEP5366;  ANEQ1415;  ANEP5360;  ANEQ2768;  ANEQ2764;  ANEQ1393;  ANEQ2766;  ANEQ2785;  ANEQ2783;  ANEQ2786;  ANEQ4098;  ANEQ4099;  ANEQ2780;  ANEQ4103;  ANEQ2779;  ANEP5381;  ANEQ2782;  ANEQ2784;  ANEQ4095;  ANEQ4089;  ANEQ4086;  ANEQ4085;  ANEQ4097;  ANEQ4088;  ANEQ2767;  ANEQ4110;  ANEQ2787;  ANEQ4102;  ANEQ4108;  ANEQ4100;  ANEQ4096;  ANEQ2781;  ANEQ2770;  ANEQ4087;  ANEQ4107;  ANEQ4084;  ANEQ4987;  ANEQ4993;  ANEQ4990;  ANEQ4992;  ANEQ4994;  ANEQ4995;  ANEQ4991;  ANEQ4988;  ANEQ4989;  ANEQ4986;  ANEQ5006;  ANEQ5004;  ANEQ5003;  ANEQ5005;    O.U.S. Devices Serial Numbers:  ANDQ5221;  ANDR0758;  ANDR0758;  ANDS3980;  ANDS3980;  ANDV0265;  ANDV2520;  ANDW4415;  ANDY3303;  ANDZ3832;  ANES1031;  ANEZ0929;  ANDQ5221;  ANDR0758;  ANDR0758;  ANDS3980;  ANDW4415;  ANDX5306;  ANDX5306;  ANDY2465;  ANDY3303;  ANDZ3832;  ANDZ3832;  ANEN3393;  ANES1031;  ANEW3409;  ANEZ0929;  ANEZ3417;  ANEZ3417;  ANFN2849;  ANDV0265;  ANEZ3417;  ANDX5306;  ANDP3970;  ANDS3980;  ANES1031;  ANDW4415;  ANDZ3832;  ANDS3980;  ANDV0265;  ANDQ3890;  ANEQ4969;  ANDS3980;  ANES1031;  ANDR0758;  ANDW4415;  ANEW3409;  ANDQ0330;  ANEQ1363;  ANDZ3832;  ANDR0758;  ANES1031;  ANES1031;  ANDV0265;  ANEQ4969;  ANFN2849;  ANEQ1363;  ANDS3980;  ANDX5306;  ANDY2465;  ANDY2465;  ANEP2433;  ANEQ4969;  ANES1031;  ANDR0758;  ANDW4415;  ANEZ0929;  ANDV0265;  ANDX5306;  ANEX6159;  ANDX5306;  ANEQ2743;  ANDR0758;  ANDV2520;  ANEQ1363;  ANEZ3417;  ANEZ3417;  ANES1031;  ANEZ3417;  ANDS3980;  ANDX5306;  ANES1031;  ANDR0758;  ANDS3980;  ANEN3393;  ANEQ1363;  ANEQ1363;  ANEQ4969;  ANEZ3417;  ANFP2367;  ANDQ3890;  ANDY2465;  ANDY2465;  ANEQ1363;  ANEQ4969;  ANDQ3890;  ANDQ3890;  ANDQ5221;  ANES1031;  ANES1031;  ANES1031;  ANDP3970;  ANDS3980;  ANDY2465;  ANEX0752;  ANEZ0929;  ANCX1871;  ANDR0758;  ANEZ3417;  ANDR0758;  ANDX5306;  ANEZ3417;  ANEP2433;  ANDW4415;  ANDR0758;  ANES1031;  ANES1031;  ANDZ3832;  ANDN0834;  ANDQ3890;  ANDQ5221;  ANDR0758;  ANDV0265;  ANDV2520;  ANDV2520;  ANDW4415;  ANDX5306;  ANDX5306;  ANEX6159;  ANEZ0929;  ANEZ3417;  ANEZ3417;  ANFN2849;  ANEN3393;  ANEP2433;  ANEQ1363;  ANEQ2743;  ANDQ3890;  ANDR0758;  ANDQ5221;  ANDW4415;  ANDY2465;  ANEP2433;  ANEQ1363;  ANEQ4969;  ANES1031;  ANES1031;  ANEX6159;  ANEZ0929;  ANEZ0929;  ANEZ0929;  ANDS3980;  ANEQ1363;  ANEP2433;  ANEQ1363;  ANEQ4969;  ANEZ3417;  ANES1031;  ANDR0758;  ANDR0758;  ANDS3980;  ANCQ4297;  ANDP3970;  ANDS3980;  ANES1031;  ANDY2465;  ANDQ0330;  ANDQ3890;  ANDQ3890;  ANDQ5221;  ANDX5306;  ANDZ1339;  ANDV0265;  ANDZ1339;  ANEN3393;  ANEP2433;  ANES1031;  ANDS3980;  ANDY2465;  ANEQ2743;  ANFP2367;  ANCU3715;  ANEX6159;  ANEQ2743;  ANES1031;  ANDQ5221;  ANDS3980;  ANEQ1363;  ANDV2520;  ANCX1871;  ANDQ5221;  ANDV0265;  ANDX5306;  ANDQ3890;  ANDZ3832;  ANEQ1363;  ANEQ2743;  ANEQ4969;  ANES1031;  ANES1031;  ANEX0752;  ANEZ3417;  ANDZ3832;  ANDQ3890;  ANDQ5221;  ANDR0758;  ANEQ1363;  ANES1031;  ANEX6159;  ANEW3409;  ANEX0752;  ANEX6159;  ANEX6159;  ANFN2849;  ANFP0883;  ANFP2367;  ANFP2367;  ANDY2465;  ANDQ3051;  ANDQ3051;  ANDQ5229;  ANDS4783;  ANDS4783;  ANDT2944;  ANDW4348;  ANDX0755;  ANDX0755;  ANDX0755;  ANDX5326;  ANDX5326;  ANDX5388;  ANDY1215;  ANDZ1355;  ANDZ3831;  ANEN1066;  ANEN4887;  ANEP2422;  ANEP2422;  ANEQ2744;  ANES2217;  ANES2217;  ANEW3411;  ANEZ0930;  ANDQ5229;  ANDS1325;  ANDS3979;  ANDS4783;  ANDW4348;  ANDX2423;  ANDX3070;  ANDX5326;  ANDX5326;  ANDX5326;  ANDX5388;  ANDX5388;  ANDZ3831;  ANDZ3831;  ANEN1066;  ANER1539;  ANES4092;  ANEW3411;  ANEZ3418;  ANFN2792;  ANEN1066;  ANDX3070;  ANEP2422;  ANDS4783;  ANDX0755;  ANDY1215;  ANDS3979;  ANFN2792;  ANFP2356;  ANDS4783;  ANDT1684;  ANDZ3831;  ANEP2422;  ANEQ1364;  ANDS3979;  ANFN2792;  ANDX2423;  ANDX3070;  ANCS1570;  ANDX2037;  ANEQ2744;  ANDQ3051;  ANDV2513;  ANDX3070;  ANEN4887;  ANDS4783;  ANDS4783;  ANDX2037;  ANEW3411;  ANDQ3886;  ANDQ3886;  ANDZ3831;  ANEQ2744;  ANDQ3051;  ANDS4783;  ANDS4783;  ANDS4783;  ANDX5388;  ANEZ2105;  ANDQ3051;  ANDX5326;  ANEQ1364;  ANEZ3418;  ANDS4783;  ANFP2356;  ANDX5388;  ANDQ3886;  ANEQ2744;  ANDQ3051;  ANDX5388;  ANDQ3051;  ANDS3979;  ANDS4783;  ANFP2356;  ANDQ3051;  ANDS1325;  ANFP2356;  ANDT1684;  ANDX2037;  ANDX2037;  ANDX3070;  ANEZ0930;  ANCU2730;  ANDS1325;  AND",
      "center_classification_date": "20210617",
      "distribution_pattern": "Worldwide distribution. US Nationwide, ALGERIA,   ANDORRA,  ARGENTINA,  AUSTRALIA,  AUSTRIA,  AZERBAIJAN,  BAHRAIN,  BELGIUM,  BRAZIL,  BRUNEI,  BULGARIA,  CHILE,  COSTA RICA,  CYPRUS,  CZECH REPUBLIC,  DENMARK,  DEUTSCHLAND,  EGYPT,  FINLAND,  FRANCE,  GEORGIA,  GREECE,  HONG KONG,  HUNGARY,  INDIA,  IRAN,  IRAQ,  IRELAND,  ISRAEL,  ITALY,  JORDAN,  KAZAKHSTAN,  MALAYSIA,  MEXICO,  NETHERLANDS,  NORWAY,  OMAN,  PANAMA,  POLAND,  PORTUGAL,  ROMANIA,  RUSSIA,  SAUDIA ARABIA,  SINGAPORE,  SLOVAKIA (SLOVAK REPUBLIC),  SLOVENIA,  SOUTH AFRICA,  SOUTH KOREA,  SPAIN,  SWEDEN,  SWITZERLAND,  TAIWAN,  TURKEY,  UNITED ARAB EMIRATES (UAE),  UNITED KINGDOM,  VIETNAM",
      "state": "AZ",
      "product_description": "VENOVO Venous Stent System 10F    Product Description(REF)/dimension:  VENEL16040\tVenovo 10F 16/40/1200mm OUS  VENEL16060\tVenovo 10F 16/60/1200mm OUS  VENEL16080\tVenovo 10F 16/80/1200mm OUS  VENEL16100\tVenovo 10F 16/100/1200mm OUS  VENEL16120\tVenovo 10F 16/120/1200mm OUS  VENEL16140\tVenovo 10F 16/140/1200mm OUS  VENEL16160\tVenovo 10F 16/160/1200mm OUS  VENEL18040\tVenovo 10F 18/40/1200mm OUS  VENEL18060\tVenovo 10F 18/60/1200mm OUS  VENEL18080\tVenovo 10F 18/80/1200mm OUS  VENEL18100\tVenovo 10F 18/100/1200mm OUS  VENEL18120\tVenovo 10F 18/120/1200mm OUS  VENEL18140\tVenovo 10F 18/140/1200mm OUS  VENEL18160\tVenovo 10F 18/160/1200mm OUS  VENEL20040\tVenovo 10F 20/40/1200mm OUS  VENEL20060\tVenovo 10F 20/60/1200mm OUS  VENEL20080\tVenovo 10F 20/80/1200mm OUS  VENEL20100\tVenovo 10F 20/100/1200mm OUS  VENEL20120\tVenovo 10F 20/120/1200mm OUS  VENEL20140\tVenovo 10F 20/140/1200mm OUS  VENEL20160\tVenovo 10F 20/160/1200mm OUS  VENEM16040\tVenovo 10F 16/40/800mm OUS  VENEM16060\tVenovo 10F 16/60/800mm OUS  VENEM16080\tVenovo 10F 16/80/800mm OUS  VENEM16100\tVenovo 10F 16/100/800mm OUS  VENEM16120\tVenovo 10F 16/120/800mm OUS  VENEM16140\tVenovo 10F 16/140/800mm OUS  VENEM16160\tVenovo 10F 16/160/800mm OUS  VENEM18040\tVenovo 10F 18/40/800mm OUS  VENEM18060\tVenovo 10F 18/60/800mm OUS  VENEM18080\tVenovo 10F 18/80/800mm OUS  VENEM18100\tVenovo 10F 18/100/800mm OUS  VENEM18120\tVenovo 10F 18/120/800mm OUS  VENEM18140\tVenovo 10F 18/140/800mm OUS  VENEM18160\tVenovo 10F 18/160/800mm OUS  VENEM20040\tVenovo 10F 20/40/800mm OUS  VENEM20060\tVenovo 10F 20/60/800mm OUS  VENEM20080\tVenovo 10F 20/80/800mm OUS  VENEM20100\tVenovo 10F 20/100/800mm OUS  VENEM20120\tVenovo 10F 20/120/800mm OUS  VENEM20140\tVenovo 10F 20/140/800mm OUS  VENEM20160\tVenovo 10F 20/160/800mm OUS  VENUL16040\tVenovo 10F 16/40/1200mm US  VENUL16060\tVenovo 10F 16/60/1200mm US  VENUL16080\tVenovo 10F 16/80/1200mm US  VENUL16100\tVenovo 10F 16/100/1200mm US  VENUL16120\tVenovo 10F 16/120/1200mm US  VENUL16140\tVenovo 10F 16/140/1200mm US  VENUL16160\tVenovo 10F 16/160/1200mm US  VENUL18040\tVenovo 10F 18/40/1200mm US  VENUL18060\tVenovo 10F 18/60/1200mm US  VENUL18080\tVenovo 10F 18/80/1200mm US  VENUL18100\tVenovo 10F 18/100/1200mm US  VENUL18120\tVenovo 10F 18/120/1200mm US  VENUL18140\tVenovo 10F 18/140/1200mm US  VENUL18160\tVenovo 10F 18/160/1200mm US  VENUL20040\tVenovo 10F 20/40/1200mm US  VENUL20060\tVenovo 10F 20/60/1200mm US  VENUL20080\tVenovo 10F 20/80/1200mm US  VENUL20100\tVenovo 10F 20/100/1200mm US  VENUL20120\tVenovo 10F 20/120/1200mm US  VENUL20140\tVenovo 10F 20/140/1200mm US  VENUL20160\tVenovo 10F 20/160/1200mm US  VENUM16040\tVenovo 10F 16/40/800mm US  VENUM16060\tVenovo 10F 16/60/800mm US  VENUM16080\tVenovo 10F 16/80/800mm US  VENUM16100\tVenovo 10F 16/100/800mm US  VENUM16120\tVenovo 10F 16/120/800mm US  VENUM16140\tVenovo 10F 16/140/800mm US  VENUM16160\tVenovo 10F 16/160/800mm US  VENUM18040\tVenovo 10F 18/40/800mm US  VENUM18060\tVenovo 10F 18/60/800mm US  VENUM18080\tVenovo 10F 18/80/800mm US  VENUM18100\tVenovo 10F 18/100/800mm US  VENUM18120\tVenovo 10F 18/120/800mm US  VENUM18140\tVenovo 10F 18/140/800mm US  VENUM18160\tVenovo 10F 18/160/800mm US  VENUM20040\tVenovo 10F 20/40/800mm US  VENUM20060\tVenovo 10F 20/60/800mm US  VENUM20080\tVenovo 10F 20/80/800mm US  VENUM20100\tVenovo 10F 20/100/800mm US  VENUM20120\tVenovo 10F 20/120/800mm US  VENUM20140\tVenovo 10F 20/140/800mm US  VENUM20160\tVenovo 10F 20/160/800mm US",
      "report_date": "20210623",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Bard Peripheral Vascular Inc",
      "recall_number": "Z-1897-2021",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "87208",
      "more_code_info": "X2037;  ANEZ0930;  ANDS4783;  ANDX0755;  ANDX2423;  ANDX2423;  ANDX3070;  ANDZ3831;  ANEP2422;  ANEP2422;  ANEP2422;  ANEP2422;  ANEP2422;  ANEQ1364;  ANEQ2744;  ANER1539;  ANER1539;  ANES2217;  ANEW3411;  ANEX0751;  ANEZ2105;  ANEZ3418;  ANFN1448;  ANFN1448;  ANFN2792;  ANDX2037;  ANER1539;  ANDQ3051;  ANDQ3051;  ANDS4783;  ANDX2037;  ANDX2423;  ANDX3070;  ANDZ1355;  ANDZ3831;  ANER1539;  ANES2217;  ANFN1448;  ANEQ2744;  ANEP3928;  ANEQ2744;  ANFN3983;  ANEN4887;  ANDS4783;  ANDX2423;  ANDX3070;  ANDX5326;  ANEP2422;  ANEP3928;  ANEP3928;  ANFP2356;  ANDX5388;  ANEZ2105;  ANEZ3418;  ANFN1448;  ANDS1325;  ANEP2422;  ANFN1448;  ANDX5326;  ANDX5388;  ANEP3928;  ANEP3928;  ANEQ2744;  ANEX0751;  ANDS1325;  ANCU2730;  ANDX3070;  ANDX3070;  ANEP2422;  ANDX5388;  ANDS1325;  ANDX3070;  ANDX3070;  ANDX3070;  ANEZ3418;  ANEQ2744;  ANEZ2105;  ANDV0708;  ANDX0755;  ANDX3070;  ANEN2135;  ANEQ2744;  ANEQ2744;  ANDS4783;  ANDS4783;  ANDS4783;  ANDX3070;  ANDX5326;  ANEN1066;  ANEP3928;  ANEP3928;  ANEQ1364;  ANFP2356;  ANDX5326;  ANDX5388;  ANEP2422;  ANDQ3051;  ANDV0708;  ANDX5326;  ANDS3979;  ANDQ3051;  ANDT1684;  ANEN2135;  ANEQ2744;  ANDX0755;  ANDS4783;  ANER1539;  ANEQ2744;  ANFN1448;  ANEP2422;  ANDQ3051;  ANDS4783;  ANEQ2744;  ANDV2513;  ANEP3928;  ANFN2792;  ANDV0708;  ANDX3070;  ANEN4887;  ANFP2356;  ANCS1570;  ANEP2422;  ANDV1984;  ANEQ2744;  ANEN1066;  ANEN1066;  ANEN2135;  ANDS4783;  ANDV0708;  ANDX5388;  ANDT2944;  ANEP3928;  ANEQ1364;  ANFN2792;  ANCY4319;  ANEP2422;  ANDP3967;  ANDQ3051;  ANDQ3051;  ANDS1325;  ANEP3928;  ANEQ2744;  ANER1539;  ANES1046;  ANEW3411;  ANEZ0930;  ANEZ2105;  ANEN2135;  ANER1539;  ANER1539;  ANFN1448;  ANDP3967;  ANDP3967;  ANDQ3051;  ANDS1325;  ANDS4783;  ANDT1684;  ANDT2944;  ANDW4348;  ANDX0755;  ANDX2037;  ANDX2423;  ANDX2423;  ANDX3070;  ANDY1215;  ANDZ3831;  ANEN2135;  ANEN4887;  ANES1046;  ANES1046;  ANES1046;  ANES2217;  ANEW3411;  ANEW3411;  ANEX0751;  ANEX0751;  ANEZ2105;  ANEZ3418;  ANEZ3418;  ANFN2792;  ANFN3983;  ANFN3983;  ANFP0884;  ANEN1066;  ANEN2135;  ANEN2135;  ANEP2422;  ANEP3928;  ANFP0884;  ANDQ3051;  ANDQ3886;  ANDQ5229;  ANDS4783;  ANDT1684;  ANDT2944;  ANDV2513;  ANDX2423;  ANDX5326;  ANEN2135;  ANEN4887;  ANEP2422;  ANEP3928;  ANEQ1364;  ANEQ2744;  ANEQ2744;  ANER1539;  ANEZ0930;  ANEZ0930;  ANEZ2105;  ANEZ3418;  ANDT2944;  ANDV0708;  ANDQ5229;  ANDX5326;  ANDX5388;  ANDS1325;  ANDX2423;  ANDZ1355;  ANDZ3831;  ANDQ3886;  ANEN2135;  ANDX0755;  ANDQ3886;  ANDS4783;  ANDT1684;  ANFP0884;  ANDS1325;  ANDS4783;  ANEN2135;  ANEN4887;  ANEP3928;  ANFP2356;  ANEP3928;  ANEQ1364;  ANES4092;  ANDT2944;  ANDY1215;  ANCS0657;  ANEZ0930;  ANEZ2105;  ANDS4783;  ANES4092;  ANEX0751;  ANDQ3051;  ANEP2422;  ANEQ2744;  ANER1539;  ANDQ3051;  ANDX2423;  ANDX2423;  ANEN2135;  ANEP2422;  ANEQ2744;  ANER1539;  ANEZ0930;  ANDS4783;  ANER1539;  ANEQ1364;  ANEZ3418;  ANDY1215;  ANDS4783;  ANDT2944;  ANDV0252;  ANDV0708;  ANDV1247;  ANDV1984;  ANDV2513;  ANDW4348;  ANDX0755;  ANDX0755;  ANDX2037;  ANDX2037;  ANDZ1355;  ANDS4783;  ANDS4783;  ANEP2422;  ANFP0884;  ANDV1984;  ANDX2423;  ANDX2423;  ANDY1215;  ANEQ2744;  ANEQ2744;  ANFP2356;  ANDS4783;  ANDY1215;  ANDX2423;  ANDY1215;  ANEN1066;  ANEN1066;  ANEP2422;  ANES4092;  ANDX5388;  ANEQ1364;  ANDX5326;  ANDZ3831;  ANDN3156;  ANEQ2744;  ANCT2427;  ANDX3070;  ANDX0755;  ANDX5326;  ANES4092;  ANES4092;  ANEX0751;  ANEZ2105;  ANDQ3051;  ANDS1325;  ANDQ3886;  ANDS3979;  ANDX5326;  ANDZ1355;  ANDQ3051;  ANDX2037;  ANDX3070;  ANEN2135;  ANEN4887;  ANEQ2744;  ANEQ2744;  ANEZ0930;  ANFN1448;  ANFN1448;  ANFN2792;  ANDN1921;  ANDS1325;  ANDX2037;  ANDX5326;  ANDX5388;  ANDY1215;  ANDZ1355;  ANDZ3831;  ANFN1448;  ANEZ0930;  ANEQ2744;  ANEN1066;  ANFP0884;  ANFP2356;  ANDV2513;  ANFP3400;  ANDX5326;  ANDQ3891;  ANDQ5239;  ANDS3981;  ANDS3981;  ANDS4784;  ANDS4784;  ANDT1685;  ANDV3571;  ANDW1485;  ANDX2424;  ANDX2424;  ANDX2424;  ANDZ1341;  ANDZ1341;  ANDZ2289;  ANDZ3833;  ANDZ3833;  ANEP0869;  ANES1032;  ANDZ3833;  ANDQ3891;  ANDS0760;  ANDS3981;  ANDS4784;  ANDX0707;  ANDX2424;  ANDX5318;  ANDX5318;  ANDY1224;  ANDZ1341;  ANDZ3833;  ANEN1067;  ANEN2137;  ANER1542;  ANES1032;  ANES4093;  ANEZ0931;  ANEZ0931;  ANEZ3419;  ANDV0694;  ANEW2004;  ANEP0869;  ANDX2424;  ANEQ1365;  ANCQ1444;  ANCS3356;  ANDT1685;  ANDZ3833;  ANEP3940;  ANEV3070;  ANEZ0931;  ANEN1067;  ANEP2431;  ANEZ0931;  ANEZ0931;  ANDS3981;  ANEP2431;  ANER1542;  ANDX5318;  ANDV3571;  ANES4093;  ANEV3070;  ANDY1224;  ANES1032;  ANES4093;  ANDS0760;  ANDX2424;  ANEQ1365;  ANEZ3937;  ANDX2041;  ANDX2424;  ANEP0869;  ANEQ1365;  ANEZ3937;  ANDS3981;  ANDW1485;  ANDX2041;  ANDX2424;  ANEZ0931;  ANDS4784;  ANDT1685;  ANDS1272;  ANDW2016;  ANFP2339;  ANCY1576;  ANDY1224;  ANCT3626;  ANDX3074;  ANDX5318;  ANER1542;  ANDQ0331;  ANES1032;  ANFP2339;  ANDX2424;  ANDW1485;  ANDX5318;  ANDQ0331;  ANDT1685;  ANDX2041;  ANDX5318;  ANDY1224;  ANDZ3833;  ANDX3074;  ANES4093;  ANDS3981;  ANDX0707;  ANES1032;  ANES4093;  ANDS4784;  ANDV3571;  ANDX2424;  ANDX5318;  ANEP0869;  ANES1032;  ANEW2004;  ANDZ3833;  ANEN2137;  ANEZ3937;  ANEP0869;  ANDS4784;  ANEX6015;  ANFP2339;  ANEZ0931;  ANES1032;  ANDQ0331;  ANEW3377;  ANEP2431;  ANES1032;  ANDX5318;  ANEP0869;  ANEP0869;  ANER1542;  ANES1032;  ANEZ0931;  ANDX3074;  ANEX6015;  ANDV4540;  ANEQ2745;  ANDP3992;  ANDQ0331;  ANDX5318;  ANDX5318;  ANES1032;  ANEX6015;  ANFN1495;  ANEZ0931;  ANEZ0931;  ANDX5318;  ANEP0869;  ANEP0869;  ANES1032;  ANES4093;  ANDX0707;  ANDX2424;  ANDX5318;  ANDS4784;  ANEP2431;  ANER1542;  ANFP2339;  ANDQ3891;  ANDS3981;  ANDW1485;  ANEP2431;  ANEZ0931;  ANDQ0331;  ANDQ0331;  ANDP3971;  ANDQ5239;  ANEN4874;  ANDX5318;  ANEP0869;  ANES1032;  ANEW2004;  ANER1542;  ANDV0694;  ANEP0869;  ANEV3070;  ANEP2431;  ANDZ2289;  ANEN4874;  ANEP0869;  ANEW3377;  ANEZ3937;  ANDX5318;  ANDV3571;  ANEP2431;  ANDT1685;  ANEN4874;  ANEP0869;  ANDT1685;  ANDX0707;  ANDX5318;  ANDT1685;  ANDX2424;  ANDY1224;  ANES1032;  ANEZ0931;  ANCN2261;  ANEP0869;  ANDQ0331;  ANEN1067;  ANEP2431;  ANEQ1365;  ANEV3070;  ANEX6015;  ANDZ3833;  ANER1542;  ANEV3070;  ANCY0524;  ANCY3327;  ANEN4874;  ANEP2431;  ANEZ0931;  ANDQ0331;  ANDQ5239;  ANDS0760;  ANDS1272;  ANDV0694;  ANDV3571;  ANDV4540;  ANDW1485;  ANDW1485;  ANDX0707;  ANDX0707;  ANDX2041;  ANDX3074;  ANES1032;  ANES1032;  ANES1032;  ANEW2004;  ANEW3377;  ANEX6015;  ANEX6015;  ANEX6015;  ANEZ0931;  ANEZ3419;  ANEZ3937;  ANFN1495;  ANFN1495;  ANFN2793;  ANEN4874;  ANEP0869;  ANEP2431;  ANEP3940;  ANEQ1365;  ANDQ0331;  ANDQ0331;  ANDQ3891;  ANDS4784;  ANDT1685;  ANDV0694;  ANDX2424;  ANEN1067;  ANEN2137;  ANEP0869;  ANEP0869;  ANES1032;  ANES1032;  ANES1032;  ANEZ3937;  ANEZ3937;  ANFP2339;  ANDS3981;  ANEZ3937;  ANDQ5239;  ANEP2431;  ANEN4874;  ANDX2041;  ANDX2424;  ANDT1685;  ANEP0869;  ANEQ1365;  ANFP2339;  ANEP0869;  ANES1032;  ANCQ3790;  ANER1542;  ANEZ3937;  ANDX2041;  ANDX2424;  ANEP0869;  ANEZ3937;  ANEQ1365;  ANDS3981;  ANDW1485;  ANDX0707;  ANDX2041;  ANDX2041;  ANDZ2289;  ANEP0869;  ANEP2431;  ANEP2431;  ANER1542;  ANER1542;  ANEW2004;  ANEZ3937;  ANCP0224;  ANCW2296;  ANEN4874;  ANEZ0931;  ANDS4784;  ANDT1685;  ANER1542;  ANES1032;  ANEX6015;  ANDY1224;  ANDT1685;  ANDW1485;  ANDW2016;  ANDW3579;  ANDW4364;  ANDX2424;  ANDX5318;  ANDX5318;  ANDZ1341;  ANEN1067;  ANCY1574;  ANCY3327;  ANDQ0331;  ANDX2424;  ANDQ0331;  ANDS1272;  ANDV3571;  ANDV4540;  ANDX2424;  ANDX2424;  ANES4093;  ANEV3070;  ANEW2004;  ANFN3984;  ANFP2339;  ANDS4784;  ANDS4784;  ANDS1272;  ANDX2424;  ANDZ3833;  ANEQ1365;  ANEZ0931;  ANDX2424;  ANES4093;  ANER1542;  ANDS1272;  ANDY1224;  ANDZ1341;  ANEN2137;  ANEN2137;  ANEZ3937;  ANCY1574;  ANDS0760;  ANEQ2745;  ANER1542;  ANFN1495;  ANDZ2289;  ANDZ3833;  ANDS4784;  ANEP0869;  ANDS4784;  ANDS3981;  ANDS3981;  ANES4093;  ANES1032;  ANES1032;  ANDQ0331;  ANDQ5239;  ANDQ3891;  ANDT1685;  ANEN2137;  ANDX5318;  ANEP3940;  ANES4093;  ANEP0869;  ANEQ1365;  ANER1542;  ANES1032;  ANEV3070;  ANEZ0931;  ANEZ0931;  ANEZ3937;  ANEZ3937;  ANEQ2745;  ANDZ1341;  ANDS4784;  ANEN2137;  ANEN4874;  ANEP0869;  ANEQ1365;  ANEQ2745;  ANER1542;  ANFN2793;  ANFN3984;  ANFP0885;  ANFP2339;  ANEQ2745;  ANES1032;  ANEZ3419;  ANDW1485;  ANEQ1365;  ANDV4540;  ANEQ1365;  ANES4093;  ANEZ3937;  ANDU1855;  ANDV4530;  ANDV4530;  ANDW4381;  ANDX0735;  ANDY3294;  ANDY3294;  ANDZ3830;  ANDZ3830;  ANEN4882;  ANDZ3830;  ANEZ3433;  ANFN2847;  ANEQ1366;  ANDX0735;  ANDZ3830;  ANEP3926;  ANDU1855;  ANEN3388;  ANEQ1366;  ANES1047;  ANDZ3830;  ANEP3926;  ANEN4882;  ANEP3926;  ANEQ1366;  ANES1047;  ANDQ5254;  ANEN4882;  ANEP3926;  ANDZ1334;  ANEP3926;  ANDX0735;  ANDX5386;  ANDW4381;  ANDX0735;  ANEP3926;  ANEQ1366;  ANES1047;  ANES1047;  ANDZ3830;  ANDR2275;  ANEQ1366;  ANDW4381;  ANEQ1366;  ANES1047;  ANEQ1366;  ANES1047;  ANDP3965;  ANDV4530;  ANDQ5254;  ANDQ5254;  ANDR2275;  ANCX2731;  ANDV4530;  ANEN4882;  ANES1047;  ANEQ1366;  ANDZ1334;  ANEP3926;  ANES1047;  ANEQ1366;  ANDP3965;  ANDV4530;  ANDY3294;  ANEQ2746;  ANDZ1334;  ANDX0735;  ANDX5386;  ANDY3294;  ANEN4882;  ANEP3926;  ANEQ1366;  ANDU1855;  ANDU1855;  ANEN4882;  ANDX5386;  ANDY3294;  ANEQ1366;  ANDP3965;  ANEQ1366;  ANEQ1366;  ANEQ1366;  ANDN0605;  ANEN4882;  ANEP3926;  ANEQ1366;  ANEP3926;  ANDU1855;  ANEN4882;  ANEQ1366;  ANEQ1366;  ANES1047;  ANEQ2746;  ANEN3388;  ANEP3926;  ANEQ1366;  ANFN2847;  ANDZ3830;  ANDY3294;  ANEP3926;  ANEP3926;  ANEZ3938;  ANDZ3830;  ANEQ1366;  ANDP3965;  ANDQ3884;  ANDR2275;  ANDR2275;  ANDU1855;  ANDU3138;  ANDU3138;  ANDV4530;  ANDW4381;  ANDX3067;  ANDX5325;  ANDX5325;  ANDX5386;  ANDY3294;  ANDZ1334;  ANDZ1334;  ANDZ3830;  ANDZ3830;  ANEN3388;  ANEQ1366;  ANEZ3938;  ANEZ3938;  ANFN2847;  ANFN2847;  ANFP0886;  ANFP0886;  ANEN3388;  ANEN4882;  ANEN4882;  ANEP2416;  ANEP3926;  ANDP3965;  ANDQ5254;  ANDV4530;  ANDX0735;  ANDX5386;  ANEP3926;  ANEQ1366;  ANEQ1366;  ANEQ2746;  ANEQ2746;  ANDU1855;  ANEZ3938;  ANDQ3884;  ANDZ3830;  ANEP3926;  ANDX5386;  ANDZ3830;  ANDU1855;  ANES1047;  ANEQ1366;  ANEQ1366;  ANDQ5254;  ANDZ1334;  ANDX5386;  ANDQ5254;  ANDV4530;  ANDX3067;  ANDZ3830;  ANEP3926;  ANEP3926;  ANEQ1366;  ANEQ1366;  ANBZ2217;  ANDX0735;  ANES1047;  ANEQ1366;  ANEQ1366;  ANEQ1366;  ANDV4530;  ANCP0219;  ANCY3321;  ANDZ3830;  ANDW4381;  ANDU1855;  ANDV4530;  ANDX0735;  ANDX0735;  ANDX5325;  ANDX5386;  ANDY3294;  ANDZ1334;  ANDZ3830;  ANEN1061;  ANDY3294;  ANDZ1334;  ANDQ3884;  ANDR2275;  ANDR2275;  ANDU1855;  ANDV4530;  ANEP3926;  ANEQ1366;  ANEQ2746;  ANES1047;  ANES1047;  ANES1047;  ANFP4501;  ANDZ1334;  ANDZ1334;  ANFP4501;  ANDQ3884;  ANEQ1366;  ANES1047;  ANDZ3830;  ANDZ3830;  ANDR2275;  ANDZ1334;  ANDZ1334;  ANEQ1366;  ANEP3926;  ANEQ1366;  ANDR2275;  ANDR2275;  ANEQ1366;  ANES1047;  ANDU1855;  ANDZ1334;  ANDU3138;  ANEQ1366;  ANEZ3938;  ANDZ3830;  ANDQ3884;  ANDR2275;  ANDR2275;  ANDU1855;  ANDX5386;  ANDU1855;  ANEP3926;  ANFP4501;  ANDY3294;  ANDV4530;  ANEN1061;  ANEN3388;  ANEQ1366;  ANEQ1366;  ANEN3388;  ANEZ3938;  ANFP0886;  ANFP2380;  ANFP4501;  ANEQ1366;  ANEQ2746;  ANEZ3938;  ANDZ3830;  ANEQ1366;  ANDV4530;  ANES1047;  ANDQ1725;  ANDQ3044;  ANDQ3044;  ANDV0713;  ANDW2008;  ANDW2008;  ANDX5365;  ANDY2453;  ANDY2453;  ANDY3279;  ANDY3279;  ANEN3383;  ANEN4885;  ANDQ1725;  ANDQ3044;  ANDX0747;  ANDX5365;  ANDY2453;  ANDY3279;  ANDY3279;  ANEN3383;  ANEQ1367;  ANEQ2747;  ANER1525;  ANEX3089;  ANEZ3420;  ANFN1449;  ANFN3985;  ANFP0887;  ANEQ4073;  ANDX5365;  ANDW2008;  ANDY2453;  ANDY2453;  ANEQ4073;  ANDY2453;  ANEQ4073;  ANEQ2747;  ANEN4885;  ANDQ3044;  ANDW2008;  ANEQ2747;  ANEQ1367;  ANDQ3044;  ANDW2008;  ANES2218;  ANEQ4073;  ANDQ3044;  ANEQ2747;  ANEQ4073;  ANEQ1367;  ANES2218;  ANEQ4073;  ANDQ3044;  ANDX5365;  ANDY2453;  ANEQ4073;  ANDY2453;  ANEQ2747;  ANES2218;  ANDQ1725;  ANEQ4073;  ANDQ1725;  ANDQ3044;  ANDQ3044;  ANDX5365;  ANEQ2747;  ANES2218;  ANDV0713;  ANEQ4073;  ANDQ1725;  ANEQ4073;  ANER1525;  ANES4094;  ANEZ0983;  ANFN1449;  ANCU2732;  ANDQ1725;  ANDQ1725;  ANDQ3044;  ANDW1481;  ANDW2008;  ANDY2453;  ANEQ1367;  ANEQ1367;  ANEQ2747;  ANEQ4073;  ANEQ4073;  ANDQ3044;  ANDQ3044;  ANDQ3044;  ANDQ3044;  ANDY2453;  ANCX2730;  ANDV0713;  ANEQ1367;  ANER1525;  ANET0559;  ANDQ1725;  ANDX5365;  ANES4094;  ANDV0713;  ANDY2453;  ANDY2453;  ANEQ4970;  ANDX5365;  ANDQ3044;  ANDV0713;  ANDW1481;  ANDW2008;  ANDX0747;  ANDX5365;  ANDQ3044;  ANDV0713;  ANDW1481;  ANDX5365;  ANDW1481;  ANEQ4073;  ANET0559;  ANDQ3044;  ANDQ3044;  ANDQ3044;  ANDV0713;  ANDX0747;  ANDY2453;  ANDY2453;  ANDY3279;  ANEN3383;  ANEN4885;  ANEQ4073;  ANEQ4970;  ANEQ4970;  ANES4094;  ANES4094;  ANET0559;  ANET0559;  ANET0559;  ANEX3089;  ANEY4326;  ANEY4326;  ANFN1449;  ANFP0887;  ANDQ3044;  ANEN3383;  ANEN4885;  ANEQ4073;  ANER1525;  ANDV0713;  ANDY2453;  ANEN4885;  ANEQ2747;  ANDW1481;  ANDV0713;  ANDV0713;  ANEQ4970;  ANES2218;  ANDQ3044;  ANDX5365;  ANDY2453;  ANEQ4073;  ANDY2453;  ANEQ4970;  ANDY3279;  ANFP2334;  ANEX3089;  ANEY4326;  ANEY4326;  ANEZ0983;  ANEZ3420;  ANFN1449;  ANFP0887;  ANEZ3420;  ANEN4885;  ANEP3913;  ANEP3913;  ANDQ3897;  ANDX2786;  ANDX2786;  ANDS2462;  ANEN4891;  ANDQ5230;  ANEP3913;  ANDX2786;  ANDQ3897;  ANDS2462;  ANDQ5230;  ANDX2786;  ANDQ5230;  ANDX2786;  ANDX2786;  ANEP3913;  ANDQ5230;  ANDQ5230;  ANDX2786;  ANER1547;  ANEN4891;  ANDS2462;  ANDQ5230;  ANDS2462;  ANDS2462;  ANEN4891;  ANEP3913;  ANDQ3897;  ANDQ5230;  ANCP3380;  ANEP3913;  ANER1547;  ANEQ1374;  ANEQ1374;  ANER1547;  ANDS2462;  ANDX5371;  ANEP3913;  ANDN0839;  ANDX2786;  ANDQ3897;  ANDQ5230;  ANEP3913;  ANEP3913;  ANDX2786;  ANDX2786;  ANDX2786;  ANDX2786;  ANDQ3897;  ANDQ3897;  ANDQ5230;  ANEQ1374;  ANDN1924;  ANDQ3066;  ANDQ5230;  ANDS2462;  ANDS4841;  ANDX2786;  ANDX2786;  ANDX5371;  ANDZ3836;  ANEN4891;  ANEP3913;  ANDS4841;  ANEQ1374;  ANER1547;  ANER1547;  ANER1547;  ANDQ3897;  ANDS2462;  ANDS4841;  ANDS4841;  ANDX2786;  ANDX2786;  ANDX2786;  ANDX5371;  ANDX5371;  ANEQ1374;  ANEQ1374;  ANEQ1374;  ANER1547;  ANER1547;  ANER1547;  ANER1547;  ANDQ5230;  ANCU3710;  ANEN4891;  ANEP3913;  ANCW4229;  ANCX1891;  ANDQ5230;  ANDQ5230;  ANDQ3897;  ANDS2462;  ANER1547;  ANER1547;  ANDX2786;  ANDQ3897;  ANDQ5230;  ANDQ5230;  ANDS2462;  ANER1547;  ANDQ3897;  ANEN4891;  ANEP3913;  ANEP3913;  ANDV1258;  ANDV1258;  ANDX0728;  ANDY2454;  ANDY3300;  ANDY3300;  ANDY3300;  ANEN4878;  ANEN4878;  ANEP0858;  ANEP0858;  ANEP0858;  ANEP0858;  ANEP2410;  ANEP5313;  ANEP5313;  ANEP5313;  ANEQ4971;  ANEQ4971;  ANER0524;  ANER0524;  ANDY2454;  ANEN1068;  ANDV1258;  ANEP5313;  ANDQ3046;  ANDQ3046;  ANDQ3046;  ANDU4050;  ANEP2410;  ANEP3915;  ANEP5313;  ANDP0504;  ANCW4228;  ANDY2454;  ANEN1068;  ANEP3915;  ANEP5313;  ANER0524;  ANEP5313;  ANEP0858;  ANEP5313;  ANEP5313;  ANEP3915;  ANDU4050;  ANDY2454;  ANDQ3046;  ANEP3915;  ANEP5313;  ANER0524;  ANDQ3046;  ANDU4050;  ANDY2454;  ANEQ4971;  ANDP0504;  ANDQ3046;  ANDU4050;  ANDX0728;  ANDX5366;  ANDY2454;  ANEP5313;  ANEP5313;  ANER0524;  ANEP5313;  ANER1510;  ANEP3915;  ANDQ3046;  ANEP2410;  ANCX1885;  ANDX5366;  ANDU4050;  ANDV0695;  ANDV1258;  ANDX0728;  ANDX5366;  ANDY2454;  ANDY3300;  ANDY3300;  ANEN1068;  ANDP0504;  ANDV0695;  ANEP3915;  ANEP5313;  ANER1510;  ANEN4878;  ANER1510;  ANES4095;  ANDQ3046;  ANDV0695;  ANDV1258;  ANDY2454;  ANDY3300;  ANEN4878;  ANEQ4971;  ANES4095;  ANES4095;  ANES4095;  ANEP5313;  ANDU4050;  ANDU4050;  ANEN1068;  ANDV1258;  ANER1510;  ANDY3300;  ANEP0858;  ANFN1450;  ANDP0504;  ANEP0858;  ANEP2410;  ANEP2410;  ANEP3915;  ANDV1258;  ANDP3964;  ANDP3964;  ANDP3964;  ANDQ3045;  ANDR4201;  ANDS4829;  ANDU1854;  ANDX5303;  ANDX5303;  ANDY3290;  ANEN1076;  ANEN3916;  ANEP3923;  ANES4072;  ANEW4224;  ANEW4224;  ANEX6013;  ANEZ3432;  ANDR4201;  ANEN3916;  ANEQ2723;  ANEQ4954;  ANEX6013;  ANEN1076;  ANDQ3045;  ANDR4201;  ANCP2415;  ANEZ3432;  ANEZ3432;  ANCW3272;  ANDP3964;  ANDU1854;  ANEQ4049;  ANEQ4954;  ANEP3923;  ANCW3272;  ANEX6013;  ANEQ2723;  ANEX3084;  ANEQ4049;  ANEQ2723;  ANEQ4049;  ANEX3084;  ANEQ4954;  ANEY2036;  ANEP3923;  ANDP3964;  ANEX6013;  ANDP2701;  ANDP3964;  ANDX5303;  ANDP3964;  ANDX5303;  ANDU1854;  ANDU1854;  ANEX6013;  ANEQ4049;  ANEX6013;  ANDU3109;  ANDR4201;  ANFP0909;  ANEN3916;  ANDU1854;  ANEN3916;  ANDR4201;  ANER5134;  ANEX6013;  ANEX6013;  ANDP3964;  ANDQ3045;  ANDR4201;  ANDY3290;  ANDY4966;  ANDP3964;  ANDQ3045;  ANDR4201;  ANDR4201;  ANDU1854;  ANDU1854;  ANDU1854;  ANDX5303;  ANDY4966;  ANDY4966;  ANEP3923;  ANEP3923;  ANEP3923;  ANEP3923;  ANEQ4049;  ANEQ4049;  ANEQ4954;  ANES4072;  ANEX6013;  ANEX6013;  ANEX6013;  ANEX6013;  ANEX6013;  ANEY2036;  ANDU3109;  ANDU1854;  ANDU3109;  ANDX5303;  ANDY4966;  ANEN3916;  ANEY2036;  ANDP3964;  ANDS4829;  ANDU1854;  ANEY2036;  ANEQ2723;  ANFP0909;  ANDR4201;  ANDY3290;  ANER5134;  ANEQ2723;  ANDQ3045;  ANDX5303;  ANDU1854;  ANDX5303;  ANDN0826;  ANDP3964;  ANDU3109;  ANDY4966;  ANEX6013;  ANFP0909;  ANDY4966;  ANEW4224;  ANEP3923;  ANDP3964;  ANCR3115;  ANDR4201;  ANDR4201;  ANEQ4954;  ANEQ4049;  ANDQ3045;  ANDR4201;  ANES4072;  ANEN3916;  ANCR3115;  ANDU1854;  ANFP0909;  ANEX6013;  ANDS4829;  ANEN1076;  ANEQ4954;  ANDP3964;  ANEX3084;  ANDY3290;  ANDU3109;  ANDY4966;  ANEQ4954;  ANES4072;  ANEY2036;  ANFP0909;  ANDR4201;  ANDR4201;  ANDS4829;  ANEQ4049;  ANFP0909;  ANDU3109;  ANDU1854;  ANDU3109;  ANDX5303;  ANEN1076;  ANEQ2723;  ANDS4829;  ANEQ4954;  ANFP2371;  ANDU1854;  ANDR2272;  ANDR3257;  ANDR3257;  ANDS4830;  ANDW4373;  ANDX1407;  ANDX5321;  ANDZ1336;  ANDZ2305;  ANEN1077;  ANEN2141;  ANEN4898;  ANEP0855;  ANER1518;  ANES4073;  ANEZ0915;  ANDP2691;  ANDP3961;  ANEQ4955;  ANCY0521;  ANDT1060;  ANER1518;  ANEX3085;  ANEW4225;  ANDX2422;  ANEX1701;  ANEP0855;  ANCQ0138;  ANCX0890;  ANDR3257;  ANDQ5249;  ANEP2413;  ANEZ3404;  ANCY0521;  ANEZ3404;  ANEZ3404;  ANDP3961;  ANEQ4050;  ANER0473;  ANEZ3404;  ANEZ2090;  ANDS4830;  ANDX1375;  ANER1518;  ANEQ4050;  ANDP1360;  ANEZ3404;  ANEP2413;  ANEX3085;  ANDX1375;  ANDX2422;  ANDT1060;  ANEQ4050;  ANEN4898;  ANCT4316;  ANEQ4050;  ANEP0855;  ANER0473;  ANEQ4955;  ANER0473;  ANEY2028;  ANDQ5249;  ANEP0855;  ANEP2413;  ANDT1682;  ANDX2422;  ANDX2422;  ANEP0855;  ANDW4373;  ANDP3961;  ANCT4316;  ANDX3066;  ANEZ2090;  ANEZ2090;  ANEP0855;  ANER0473;  ANEX3085;  ANDN0824;  ANDX2422;  ANEN4898;  ANEP0855;  ANDX5321;  ANDX5321;  ANEZ3404;  ANCY1568;  ANDP3961;  ANEP0855;  ANER0473;  ANEP0855;  ANDX2422;  ANDQ5249;  ANEY2028;  ANEZ3404;  ANDX5321;  ANEW4225;  ANEW4225;  ANEY2028;  ANEZ0915;  ANEN2141;  ANDV0703;  ANDR4199;  ANDX3066;  ANEP0855;  ANEQ4050;  ANER0473;  ANEN1077;  ANDQ3882;  ANDQ3882;  ANDR4199;  ANDS4830;  ANDT1682;  ANDU3122;  ANDU3122;  ANEP0855;  ANER1518;  ANEW4225;  ANEX1701;  ANEX3085;  ANDN0824;  ANDP1360;  ANDP2691;  ANDP3961;  ANDQ3043;  ANDQ3882;  ANDR3257;  ANDR4199;  ANDY1227;  ANCV1586;  ANCX3881;  ANCY0521;  ANDP3961;  ANDQ3882;  ANDR2272;  ANDR4199;  ANDT1682;  ANDU4071;  ANDV0703;  ANDV0703;  ANDW4373;  ANDX1375;  ANDX1407;  ANDX2422;  ANDX3066;  ANDY1227;  ANDZ1336;  ANDZ2305;  ANDZ2305;  ANEN2141;  ANEN2141;  ANEP0855;  ANEP0855;  ANEP2413;  ANEP2413;  ANEQ4050;  ANEQ4050;  ANEQ4050;  ANES2197;  ANEX0675;  ANEX0675;  ANEX1701;  ANEX3085;  ANEY2028;  ANEZ0915;  ANEZ2090;  ANEZ3404;  ANEZ3404;  ANEZ3404;  ANDX2422;  ANEN4898;  ANEP2413;  ANER1518;  ANER1518;  ANEN1077;  ANEP0855;  ANEZ3404;  ANDT1682;  ANEX3085;  ANDZ1336;  ANES4073;  ANEX3085;  ANDT1682;  ANEP2413;  ANEQ4955;  ANEX3085;  ANDX3066;  ANET0549;  ANEZ3404;  ANDR3257;  ANDR3257;  ANDX5321;  ANDZ1336;  ANEP0855;  ANES2197;  ANER1518;  ANDQ5249;  ANDT1682;  ANDZ1336;  ANER0473;  ANDX1375;  ANEZ0915;  ANDR4199;  ANDT1060;  ANDT1060;  ANDU3122;  ANDU4071;  ANDX5321;  ANDY1227;  ANEN1077;  ANEP0855;  ANET0549;  ANEX3085;  ANEY2028;  ANEZ3404;  ANDT1682;  ANER1518;  ANES2197;  ANEX0675;  ANDT1060;  ANCX0890;  ANEN2141;  ANES2197;  ANEX3085;  ANCT4316;  ANDP3961;  ANEX0675;  ANDT1060;  ANDX3066;  ANEZ2090;  ANDP3961;  ANCX3881;  ANDR4199;  ANDU4071;  ANES4073;  ANDT1682;  ANDX1407;  ANDX2422;  ANEP0855;  ANEP2413;  ANEQ4050;  ANDX2422;  ANEP2413;  ANEP2413;  ANEZ3404;  ANDX2422;  ANCY1568;  ANEZ2090;  ANER0473;  ANDQ5249;  ANDR4199;  ANDS4830;  ANDW4373;  ANEN2141;  ANEP0855;  ANEQ4050;  ANEZ3404;  ANET0549;  ANDR4199;  ANDP3961;  ANEY2028;  ANDP3961;  ANDR4199;  ANDS4830;  ANDV0703;  ANDV0703;  ANDW4373;  ANDX2422;  ANDX5321;  ANEQ4050;  ANEQ4955;  ANEP0855;  ANDV0703;  ANDP3961;  ANDP3961;  ANDP3961;  ANDQ5249;  ANDX3066;  ANES2197;  ANEX3085;  ANDU4071;  ANET0549;  ANEQ4050;  ANER0473;  ANEP2413;  ANDR4199;  ANDT1682;  ANDX2422;  ANDX5321;  ANEP0855;  ANEP2413;  ANDW4373;  ANDZ1336;  ANDT1682;  ANDX1375;  ANDX2422;  ANDZ1336;  ANET0549;  ANFN1436;  ANFP0869;  ANER1518;  ANCX0890;  ANDP3961;  ANES2197;  ANEW4225;  ANEN2141;  ANEP0855;  ANDX2422;  ANDX3066;  ANEQ4050;  ANEX3085;  ANDX2422;  ANEP2413;  ANET0549;  ANEN2141;  ANEZ3404;  ANEQ4955;  ANET0549;  ANDR3264;  ANDR3264;  ANDR3264;  ANDR4204;  ANDU1111;  ANDW4390;  ANDX2784;  ANDX5379;  ANDY4959;  ANDZ3840;  ANEN3403;  ANEN4900;  ANET0550;  ANET3790;  ANEW4226;  ANEY1226;  ANEY4327;  ANFN1437;  ANDP2697;  ANDZ1346;  ANDZ2285;  ANEP3566;  ANEP3566;  ANDR4204;  ANEZ3405;  ANCT4317;  ANDX2784;  ANCQ2604;  ANDP3974;  ANDW4390;  ANFN3963;  ANDP2697;  ANEY4327;  ANDY4959;  ANDP2697;  ANEX0676;  ANDR3264;  ANDV0707;  ANDV2514;  ANEX3086;  ANEX6014;  ANDP2697;  ANDR4204;  ANDP2697;  ANDU3136;  ANDX2784;  ANEP3941;  ANEW4226;  ANDU3136;  ANEP3566;  ANEW4226;  ANEX3086;  ANEQ4051;  ANEX3086;  ANDW4390;  ANDZ1346;  ANEP0864;  ANEP3941;  ANEZ3405;  ANEZ3405;  ANDR4204;  ANEQ4051;  ANET3790;  ANFN3963;  ANDR4204;  ANCT1142;  ANEX3086;  ANEX3086;  ANER1544;  ANDX2784;  ANEP0864;  ANEP3566;  ANES2198;  ANEX6014;  ANEX6014;  ANCT1142;  ANEW3371;  ANEY1226;  ANEX0676;  ANEX3086;  ANEX4717;  ANDX3076;  ANDW4390;  ANER1544;  ANEW3371;  ANEX6014;  ANDY1235;  ANCT1142;  ANDV0707;  ANEZ3405;  ANER1544;  ANEX6014;  ANEP3566;  ANEP0864;  ANDY1235;  ANEP0864;  ANEY1226;  ANEY4327;  ANDZ2285;  ANDV2514;  ANDR3264;  ANDR4204;  ANDV0286;  ANDV0286;  ANDV0707;  ANDV0707;  ANDX1417;  ANET3790;  ANES2198;  ANES4074;  ANET0550;  ANEX6014;  ANDP1364;  ANDP2697;  ANDP3974;  ANDR3264;  ANDR3264;  ANDR3264;  ANDY2457;  ANCT4317;  ANCY2191;  ANCY2191;  ANDR3264;  ANDR4204;  ANDR4204;  ANDU3136;  ANDV0286;  ANDV2514;  ANDV2514;  ANDW4390;  ANDX1417;  ANDX2784;  ANDX3076;  ANDX5379;  ANDY1235;  ANDY2457;  ANDY4959;  ANDY4959;  ANEN3403;  ANEP3566;  ANEP3566;  ANEP3566;  ANEP3941;  ANEQ4051;  ANEQ4051;  ANEQ4051;  ANET3790;  ANET3790;  ANEW3371;  ANEW4226;  ANEW4226;  ANEX0676;  ANEX3086;  ANEX4717;  ANEY1226;  ANEZ3405;  ANEZ3405;  ANFN1437;  ANEY4327;  ANDZ1346;  ANEW4226;  ANEX3086;  ANDR4204;  ANDZ3840;  ANEX6014;  ANDP2697;  ANDR4204;  ANDY4959;  ANEQ4956;  ANER1544;  ANES2198;  ANET3790;  ANDP3974;  ANDR3264;  ANDW4390;  ANDX3076;  ANDX5379;  ANEP3566;  ANEQ4956;  ANCY4320;  ANDP2697;  ANDY4959;  ANDZ1346;  ANEP0864;  ANEY1226;  ANEP3566;  ANDP3974;  ANDV0286;  ANEY1226;  ANFN3963;  ANEW3371;  ANEX3086;  ANFN3963;  ANCP1865;  ANDR4204;  ANFN1437;  ANER1544;  ANDR3264;  ANDR3264;  ANDX1417;  ANDX2784;  ANDX5379;  ANDY2457;  ANEN4900;  ANES2198;  ANEX4717;  ANEX6014;  ANEY1226;  ANDX5379;  ANDX2784;  ANEW4226;  ANET3790;  ANEX0676;  ANDZ3840;  ANEP3566;  ANEX0676;  ANEX6014;  ANCR0698;  ANDV0707;  ANDX2784;  ANDX3076;  ANDZ1346;  ANEY1226;  ANEZ3405;  ANER1544;  ANDV0707;  ANDY4959;  ANEP3566;  ANEX3086;  ANES2198;  ANEP3566;  ANEX4717;  ANFN1437;  ANFN1437;  ANEW4226;  ANFN1437;  ANDX2784;  ANEP3566;  ANDU1111;  ANDV2514;  ANDZ3840;  ANEQ4956;  ANEX6014;  ANEY4327;  ANDU3136;  ANEP3566;  ANER1544;  ANEY4327;  ANFN1437;  ANDR3264;  ANDR4204;  ANDR4204;  ANDV0707;  ANDV0707;  ANDX3076;  ANEP3566;  ANEP3566;  ANEP3566;  ANEP3566;  ANET0550;  ANET3790;  ANEY4327;  ANDY4959;  ANEX3086;  ANDV0286;  ANDW4390;  ANDZ1346;  ANDZ2285;  ANEQ4956;  ANFN1437;  ANER1544;  ANET3790;  ANDR4204;  ANDR4204;  ANDV0286;  ANDX5379;  ANDZ1346;  ANEN4900;  ANEP3566;  ANEQ4956;  ANER1544;  ANET0550;  ANEX3086;  ANEZ3405;  ANDV0286;  ANDX3076;  ANEP0864;  ANEP3566;  ANEP3566;  ANER1544;  ANEW3371;  ANDZ1346;  ANDP3974;  ANEW3371;  ANEX0676;  ANDX2784;  ANDY2457;  ANEX6014;  ANDR4204;  ANDZ1346;  ANDU3136;  ANEQ4956;  ANDP3974;  ANEP3566;  ANES2198;  ANEP3566;  ANDZ1346;  ANBY1142;  ANDP3974;  ANDR4204;  ANDU1111;  ANDY1235;  ANDY2457;  ANEP3566;  ANEP3941;  ANEW4226;  ANEP0864;  ANDR4204;  ANDR3264;  ANDX2784;  ANEP3941;  ANEQ4956;  ANEZ3405;  ANDP2697;  ANDR3264;  ANDY4959;  ANES2198;  ANES4074;  ANEY4327;  ANFN3963;  ANCX3882;  ANDX5379;  ANEQ4051;  ANEX4717;  ANEX6014;  ANDX2784;  ANEP3566;  ANDX2784;  ANDX3076;  ANET0550;  ANDZ3840;  ANFN1437;  ANDP1367;  ANDP1367;  ANDR0760;  ANDX2043;  ANDX5329;  ANDX5359;  ANDY1217;  ANDZ1353;  ANDZ1353;  ANCX2727;  ANDY1217;  ANEY1227;  ANDX2043;  ANDX3080;  ANEQ4957;  ANDR3265;  ANEQ4052;  ANDR0760;  ANDX2043;  ANDY4975;  ANEQ4052;  ANES2199;  ANEY4329;  ANEZ3406;  ANDP2698;  ANDR3265;  ANDX3080;  ANES2199;  ANES4075;  ANES4075;  ANDX0750;  ANEY1227;  ANEZ3406;  ANDR3265;  ANDX5329;  ANDS2480;  ANDY4975;  ANEX1706;  ANDX5359;  ANEQ4052;  ANDT1229;  ANCV1588;  ANCW3266;  ANDR3265;  ANDY1217;  ANDZ1353;  ANER1519;  ANET0551;  ANDX5359;  ANEN4892;  ANEN4892;  ANEQ2724;  ANEQ4052;  ANEQ4957;  ANET0551;  ANEY1227;  ANCP0215;  ANEQ4052;  ANEY1227;  ANEY1227;  ANEP2427;  ANEQ4052;  ANEY1227;  ANDX0750;  ANER1519;  ANEQ4052;  ANDR3265;  ANDX0750;  ANDX2043;  ANCX2727;  ANDX3080;  ANDP2698;  ANDP2698;  ANDX2043;  ANEY4329;  ANDR3265;  ANDY4975;  ANDP2698;  ANDX5359;  ANEQ4052;  ANEX1706;  ANDY4975;  ANDQ5250;  ANDY4975;  ANDT1229;  ANDS2480;  ANDQ3894;  ANDR3265;  ANDX3080;  ANDQ3063;  ANDQ3894;  ANDQ3894;  ANDR0760;  ANDR3265;  ANDS2480;  ANDX5329;  ANDY1217;  ANDY4975;  ANER1519;  ANEX1706;  ANEX1706;  ANDQ1728;  ANDQ3063;  ANDX5359;  ANCW3266;  ANCX1877;  ANDN3159;  ANDP1367;  ANDQ3063;  ANDQ3894;  ANDQ5250;  ANDQ5250;  ANDR0760;  ANDR3265;  ANDR3265;  ANDR3265;  ANDR3265;  ANDS1323;  ANDS4831;  ANDV0691;  ANDX0750;  ANDX0750;  ANDX1387;  ANDX1387;  ANDX1387;  ANDX1422;  ANDX2043;  ANDX3080;  ANDX5359;  ANDX5359;  ANEN4892;  ANEP2427;  ANEP2427;  ANEP2427;  ANEQ2724;  ANEQ4052;  ANEQ4052;  ANEQ4957;  ANES2199;  ANET0551;  ANEZ3406;  ANDT1229;  ANDX1387;  ANEY4329;  ANDX5359;  ANDT1229;  ANDT1229;  ANDT1229;  ANDV0691;  ANDY1217;  ANDS4831;  ANDY4975;  ANDS2480;  ANEQ4052;  ANER1519;  ANER1519;  ANEZ3406;  ANDR3265;  ANDS4831;  ANDT1229;  ANDS1323;  ANDQ5250;  ANDS2480;  ANDS4831;  ANDX1387;  ANDX2043;  ANDX5329;  ANDX5359;  ANDY1217;  ANDY4975;  ANDY4975;  ANEX1706;  ANEX3087;  ANEY1227;  ANDX5359;  ANEN4892;  ANES4075;  ANCP0215;  ANDX1387;  ANDX5329;  ANER1519;  ANEQ4052;  ANES4075;  ANEQ4957;  ANEY4329;  ANDX1422;  ANEX1706;  ANDS4831;  ANDX1387;  ANDX1387;  ANDX3080;  ANDQ1728;  ANDS4831;  ANDV0691;  ANDX1387;  ANDZ1353;  ANEQ4052;  ANDP2698;  ANDP2698;  ANDX3080;  ANEY4329;  ANEX1706;  ANEY1227;  ANEY4329;  ANDR3265;  ANDR3265;  ANDS2480;  ANDS2480;  ANDY4975;  ANEN4892;  ANER1519;  ANER1519;  ANET0551;  ANET0551;  ANEY1227;  ANEP2427;  ANCR0803;  ANDS2480;  ANEQ4052;  ANDZ1353;  ANEQ4052;  ANCV1588;  ANDP2698;  ANDV0691;  ANDX0750;  ANDX3080;  ANDP1367;  ANDR3265;  ANDV0691;  ANEN4892;  ANEQ4052;  ANEX3087;  ANEX3087;  ANEX3087;  ANDX1422;  ANDX2043;  ANDX5329;  ANER1519;  ANES4075;  ANES4075;  ANDP2698;  ANEQ4052;  ANET0551;  ANDX3080;  ANET0551;  ANDQ1728;  ANCR3121;  ANDS4831;  ANDX2043;  ANDY1217;  ANDY4975;  ANDY4975;  ANES4075;  ANDR3265;  ANDX0750;  ANET0551;  ANET0551;  ANEQ4052;  ANDY1217;  ANDQ5250;  ANDR3265;  ANDR3265;  ANDX5359;  ANDY1217;  ANDZ1353;  ANDX5329;  ANEZ3406;  ANDQ3063;  ANER1519;  ANDZ1353;  ANDS4831;  ANER1519;  ANDR3265;  ANDV0691;  ANDX5359;  ANET0551;  ANEY4329;  ANDS2480;  ANEQ4052;  ANEZ3406;  ANDR3265;  ANCW3266;  ANDT1229;  ANDX2043;  ANDY4975;  ANES2199;  ANES2199;  ANET0551;  ANFN2851;  ANEZ3406;  ANDZ1353;  ANDX3080;  ANEQ2724;  ANDY1217;  ANER1519;  ANDX0750;  ANDX1422;  ANDX5329;  ANEZ3406;  ANFN1438;  ANFN1438;  ANCV4705;  ANCV4705;  ANDN3158;  ANDN3158;  ANDP0516;  ANDP0516;  ANDP0516;  ANDP0516;  ANDY1223;  ANDY4961;  ANEN1063;  ANEP0865;  ANDV2516;  ANCQ1443;  ANCY2193;  ANDS0759;  ANER1541;  ANER1541;  ANDW4409;  ANDX1381;  ANEQ4053;  ANDP0516;  ANDX1381;  ANDX2040;  ANET0552;  ANDP0516;  ANDS0759;  ANDV2516;  ANEQ4053;  ANDY4961;  ANDV1232;  ANEP2414;  ANEQ4053;  ANER1541;  ANER1541;  ANCY3312;  ANDP0516;  ANDY1223;  ANDV0712;  ANEP2414;  ANER1541;  ANCV4705;  ANDW4409;  ANDW4409;  ANDX1381;  ANEN1063;  ANEP2414;  ANEP2414;  ANEQ1350;  ANEQ4958;  ANEQ1350;  ANDV2516;  ANBZ1289;  ANDR3262;  ANEP0865;  ANEQ4053;  ANDS4832;  ANDX2040;  ANDP0516;  ANDT2447;  ANET0552;  ANDX1381;  ANEQ4958;  ANER1541;  ANDS4832;  ANDV2516;  ANDX1381;  ANDX3073;  ANEQ1350;  ANEQ2725;  ANES2200;  ANDT2447;  ANDV0263;  ANDV0712;  ANDV2516;  ANDX2040;  ANDY1223;  ANEN4881;  ANES2200;  ANES4076;  ANDP0516;  ANDR3261;  ANDR3261;  ANDR3262;  ANDX3073;  ANDX5312;  ANCV4705;  ANDP0516;  ANDR3261;  ANDR3262;  ANDR3262;  ANDS4832;  ANDS4832;  ANDS4832;  ANDV0712;  ANDV1232;  ANDV1232;  ANDV2516;  ANDV2516;  ANDW4409;  ANDX2040;  ANDX5312;  ANDY1223;  ANDY4961;  ANDY4961;  ANEN1063;  ANEP2414;  ANEP2414;  ANEP2414;  ANEQ1350;  ANEQ1350;  ANEQ4053;  ANEQ4053;  ANEQ4958;  ANEQ4958;  ANER1541;  ANER1541;  ANER1541;  ANDS4832;  ANDT2447;  ANDS4832;  ANDT2447;  ANEX1711;  ANER1541;  ANDP0516;  ANDP0516;  ANDV1232;  ANDV2516;  ANDX5312;  ANEQ2725;  ANEQ4958;  ANEQ4053;  ANEQ4053;  ANDV2516;  ANCQ1443;  ANDS0759;  ANER1541;  ANEQ4053;  ANEQ4958;  ANDP0516;  ANDR3262;  ANDV0263;  ANEW3420;  ANEX1711;  ANDS0759;  ANDT2447;  ANDV0263;  ANDV1232;  ANDW4409;  ANDX3073;  ANDY1223;  ANDY4961;  ANEN4881;  ANEP0865;  ANEX0753;  ANEP2414;  ANER1541;  ANEQ2725;  ANEQ4958;  ANER1541;  ANDV2516;  ANDP0516;  ANDR3261;  ANEP2414;  ANDW4409;  ANEP2414;  ANCQ4291;  ANDX2040;  ANDY1223;  ANEP2414;  ANEW3420;  ANEW3420;  ANEX0753;  ANDR3261;  ANDT2447;  ANDX1415;  ANEN1063;  ANEP2414;  ANEQ4958;  ANER1541;  ANES2200;  ANET0552;  ANEP2414;  ANDS4832;  ANDX2040;  ANDW4409;  ANDV2516;  ANDX5312;  ANEQ2725;  ANEZ2091;  ANET0552;  ANCY3312;  ANDR3261;  ANDV0712;  ANDV2516;  ANDV2516;  ANEP2414;  ANEQ1350;  ANEQ2725;  ANEQ4053;  ANDS0759;  ANDS0759;  ANDV0263;  ANDV0263;  ANDV2516;  ANEN1063;  ANEP2414;  ANES2200;  ANEW3420;  ANEW3420;  ANDS0759;  ANDS4832;  ANDV1232;  ANEN1063;  ANEN4881;  ANES4076;  ANCQ1443;  ANDX3073;  ANDV0263;  ANDX1381;  ANCY3312;  ANDR3262;  ANDR3262;  ANDS0759;  ANEQ4958;  ANES4076;  ANDV2516;  ANES2200;  ANEN1063;  ANDX2040;  ANDR3261;  ANDR3261;  ANDS0759;  ANDS4832;  ANDX2040;  ANDY4961;  ANEN1063;  ANEP2414;  ANEQ4958;  ANER1541;  ANDR3261;  ANDX2040;  ANDT2447;  ANDX1381;  ANEN1063;  ANET0552;  ANFN3965;  ANEQ4958;  ANCX1873;  ANDV0712;  ANDV2516;  ANDX1415;  ANEQ1350;  ANEQ2725;  ANEX1711;  ANDX1381;  ANEW3420;  ANEZ2091;  ANDN0594;  ANDN0594;  ANDN0594;  ANDN0594;  ANDS1270;  ANDX2030;  ANDX5354;  ANDY1214;  ANDY1214;  ANDX3069;  ANDY1214;  ANDS1270;  ANDX2030;  ANDY1214;  ANDY1214;  ANEN3404;  ANEN3404;  ANEP2436;  ANEQ4054;  ANEQ4054;  ANCN3586;  ANDV1246;  ANDX2030;  ANDS1270;  ANEN3404;  ANDR4202;  ANDS4833;  ANDU3083;  ANDX2030;  ANDX3069;  ANDV1246;  ANDX1379;  ANDX3069;  ANDY1214;  ANDS1270;  ANDX3069;  ANCV1595;  ANDN0594;  ANDS1270;  ANDU3083;  ANDV1246;  ANDV1246;  ANDX2030;  ANDX3069;  ANEN3404;  ANEP2436;  ANEP2436;  ANEQ4054;  ANDR4202;  ANDR4202;  ANDS1270;  ANDS1270;  ANDT2448;  ANDX1379;  ANDX2030;  ANES2201;  ANEQ4054;  ANDS4833;  ANDT2448;  ANDU3083;  ANDU3083;  ANEQ4054;  ANEN3404;  ANDT2448;  ANDU3083;  ANDU3083;  ANDX2030;  ANES2201;  ANES2201;  ANDX2030;  ANDX3069;  ANDN0594;  ANEQ4054;  ANES2201;  ANDT2448;  ANEQ4054;  ANDY1214;  ANDN0594;  ANDU3083;  ANEP2436;  ANEQ4054;  ANDX3069;  ANDX5354;  ANFN3966;  ANEN3404;  ANDR4202;  ANEW3415;  ANEW3415;  ANDY1214;  ANDX1379;  ANEW3415;  ANEX0677;  ANDV4544;  ANDQ1727;  ANDQ1727;  ANDV4544;  ANDX0764;  ANDX1370;  ANDX1421;  ANDX5358;  ANDY1228;  ANCR1659;  ANDX0764;  ANDQ5228;  ANDV3574;  ANDW1488;  ANDX1370;  ANDX1421;  ANDQ1727;  ANDX1385;  ANDV3574;  ANDX0764;  ANDX1370;  ANDX1370;  ANEQ4055;  ANEX0677;  ANEX0677;  ANDV4544;  ANDQ1727;  ANDX0764;  ANDS2450;  ANCV3483;  ANDW1488;  ANDQ1727;  ANDQ3061;  ANDR2280;  ANDS1284;  ANDX1421;  ANCU1845;  ANCV1589;  ANDQ5228;  ANDX1370;  ANDX1370;  ANDX1370;  ANDX1370;  ANDX1385;  ANDX1385;  ANDX1421;  ANDY1228;  ANEZ0916;  ANEZ2092;  ANEZ3408;  ANDS4786;  ANEX0677;  ANDX0764;  ANDX0764;  ANDX5358;  ANDY1228;  ANEZ0916;  ANEX0677;  ANDQ1727;  ANDQ1727;  ANDQ3061;  ANDQ5228;  ANDR2280;  ANDS2450;  ANDS4786;  ANDS4834;  ANDT4133;  ANDV0688;  ANDV3574;  ANDV3574;  ANEZ0916;  ANEZ3408;  ANEX0677;  ANEQ4055;  ANDX5358;  ANEQ4055;  ANFP2372;  ANDS2450;  ANDX1421;  ANDS2450;  ANDX5358;  ANDV3574;  ANDX1385;  ANEQ4055;  ANEZ2092;  ANDX1385;  ANDX0764;  ANCV3483;  ANDV4544;  ANDX0764;  ANDX1370;  ANDX5358;  ANEQ4055;  ANES2202;  ANEZ3408;  ANDS4834;  ANDX1370;  ANDR2280;  ANDS2450;  ANDV3574;   ANDV4544;  ANDV4544;  ANDX5358;  ANDY1228;  ANDY1228;  ANEQ4055;  ANEQ4055;  ANES2202;  ANDQ5228;  ANDX1370;  ANDS4834;  ANDV3574;  ANDX1385;  ANEQ4055;  ANES2202;  ANDQ1727;  ANDS2450;  ANDV4544;  ANDX1370;  ANDX1370;  ANDX1385;  ANDX1421;  ANFP2372;  ANDX1385;  ANDX5358;  ANDX1421;  ANDX0764;  ANDX1370;  ANEZ0916;  ANFN3967;  ANDS2450;  ANDX1370;  ANES2202;  ANDX5358;  ANEQ4055;  ANEZ3408;  ANDP3983;  ANDP3983;  ANDT1056;  ANDT1056;  ANDX5360;  ANDZ1349;  ANDZ2302;  ANEN4890;  ANEN4890;  ANEP3917;  ANER1526;  ANEW4261;  ANEW4261;  ANEZ3421;  ANFP3414;  ANEQ4074;  ANDT1056;  ANEP5321;  ANEN4890;  ANEW4261;  ANEQ4074;  ANCX1881;  ANEP5321;  ANER1526;  ANEQ4074;  ANEQ4074;  ANER1526;  ANDX5360;  ANDX5360;  ANDP3983;  ANEQ4074;  ANDT1056;  ANEP5321;  ANEQ4074;  ANER1526;  ANFP3414;  ANDX5360;  ANDN0838;  ANDZ2302;  ANCX1881;  ANDT1056;  ANDX5360;  ANDX5360;  ANDX5360;  ANER1526;  ANDX5360;  ANER1526;  ANEP5321;  ANEP5321;  ANDN0838;  ANDT1056;  ANEQ4074;  ANER1526;  ANDN0838;  ANDT1056;  ANDT1056;  ANDW4398;  ANDX5360;  ANDZ1349;  ANDZ2302;  ANDZ2302;  ANDT1056;  ANEQ4074;  ANDT1056;  ANDZ2302;  ANDT1056;  ANEP5321;  ANFP3414;  ANEP5321;  ANEQ4074;  ANEQ4074;  ANDN0838;  ANEP3917;  ANFP3414;  ANFP3414;  ANDP3995;  ANDS4787;  ANDX0761;  ANDZ1351;  ANDZ2280;  ANDZ2280;  ANDZ2280;  ANEN2143;  ANEN3394;  ANER1546;  ANEZ0984;  ANFN3986;  ANFP0888;  ANEN3394;  ANDZ2280;  ANDV2136;  ANDV2518;  ANEP2411;  ANEQ1368;  ANDP3981;  ANDX5350;  ANEQ1368;  ANER1546;  ANDS4787;  ANCX2734;  ANEQ2748;  ANDX5350;  ANEP2411;  ANEQ4972;  ANEQ4972;  ANDP3981;  ANDU4073;  ANDV0700;  ANDX0761;  ANES1033;  ANDV2518;  ANDX5350;  ANEQ1368;  ANER1546;  ANEQ4972;  ANDP3981;  ANDP3981;  ANDP3995;  ANDU4073;  ANDX5350;  ANDV2518;  ANCX2734;  ANDS4787;  ANDU4073;  ANDV2136;  ANEN2143;  ANES1033;  ANDS4787;  ANES1033;  ANDV2518;  ANEQ4972;  ANEQ1368;  ANDS4787;  ANEP2411;  ANDP3981;  ANEQ2748;  ANDP3995;  ANDV2136;  ANDX5350;  ANEP2411;  ANEQ1368;  ANDX5350;  ANEQ1368;  ANEQ4972;  ANDX0761;  ANDX5350;  ANDX5350;  ANDX5350;  ANEP2411;  ANEQ1368;  ANEQ1368;  ANEQ1368;  ANEQ1368;  ANER1546;  ANDX0761;  ANDV0700;  ANEN4875;  ANEQ1368;  ANDV1211;  ANDZ1351;  ANEN2143;  ANEQ2748;  ANDP3995;  ANDX5350;  ANEQ2748;  ANEN3394;  ANDS4787;  ANDV2136;  ANEQ1368;  ANCY3330;  ANEP2411;  ANDV0700;  ANEQ2748;  ANEN3394;  ANEQ1368;  ANER1546;  ANEQ2748;  ANEQ4972;  ANDP3981;  ANDV2136;  ANEQ1368;  ANEN4875;  ANDV1211;  ANEQ1368;  ANEQ2748;  ANDP3995;  ANEQ1368;  ANEQ1368;  ANEQ1368;  ANEQ2748;  ANEQ4972;  ANEQ4972;  ANER1546;  ANER1546;  ANES1033;  ANEZ0984;  ANFP0888;  ANEQ2748;  ANDS4787;  ANDN0837;  ANDP3981;  ANDP3981;  ANDP3981;  ANDP3995;  ANDP3995;  ANDP3995;  ANDV0700;  ANDV2136;  ANDV2136;  ANDX0761;  ANDX5350;  ANFN1451;  ANFP0888;  ANFP0888;  ANFP0888;  ANFP0888;  ANEN4875;  ANEP2411;  ANEP2411;  ANEQ1368;  ANCX2734;  ANDP3995;  ANDV2518;  ANDX0761;  ANER1546;  ANES1033;  ANEZ0984;  ANDP3981;  ANDX5350;  ANDV2136;  ANDV2518;  ANEP2411;  ANDP2699;  ANDU4073;  ANDS4787;  ANDV2136;  ANEP2411;  ANDZ2280;  ANCS3352;  ANDS4787;  ANDS4787;  ANES1033;  ANES1033;  ANDP3995;  ANDV2136;  ANDX0761;  ANEQ4972;  ANDX5350;  ANDV1211;  ANDV2518;  ANDZ1351;  ANEN2143;  ANDX5350;  ANDX5350;  ANEN3394;  ANDZ1351;  ANDZ1351;  ANDZ2280;  ANEQ1368;  ANES1033;  ANES1033;  ANDS4787;  ANDX5350;  ANDP3981;  ANEQ1368;  ANES1033;  ANEN2143;  ANDV2136;  ANES1033;  ANDV2136;  ANCY3330;  ANDP3981;  ANDU4073;  ANDV2136;  ANEN3394;  ANEQ1368;  ANEQ1368;  ANCW1438;  ANDU4073;  ANDX5350;  ANEP2411;  ANEQ1368;  ANEQ1368;  ANER1546;  ANES1033;  ANDV1211;  ANDX5350;  ANES1033;  ANFN1451;  ANES1033;  ANDX5350;  ANEN3394;  ANEZ0984;  ANEZ0984;  ANFN1451;  ANFN1451;  ANFP0888;  ANES1033;  ANDV2136;  ANDZ2280;  ANFP3413;  ANEQ1368;  ANCS3351;  ANCT0405;  ANCY0522;  ANDS2471;  ANDS2471;  ANDS2471;  ANDS4781;  ANDS4781;  ANDS4781;  ANDX5328;  ANDX5328;  ANDX5328;  ANDX5328;  ANDX5368;  ANDX5368;  ANDX5368;  ANDY1233;  ANDY1233;  ANEP3916;  ANEQ1369;  ANEQ2749;  ANEQ2749;  ANER0525;  ANES1048;  ANDZ1356;  ANDQ0328;  ANDS2471;  ANDW4416;  ANDX0719;  ANDX5368;  ANDZ1356;  ANEN2138;  ANEN2138;  ANER1538;  ANEZ0932;  ANEZ2106;  ANCY0522;  ANEP3916;  ANEQ1369;  ANEQ2749;  ANEN2138;  ANES1048;  ANES1048;  ANEQ2749;  ANEQ2749;  ANES1048;  ANCW1442;  ANEP3916;  ANCQ3793;  ANDX5328;  ANDX0719;  ANDX5328;  ANDZ1356;  ANER1538;  ANER1538;  ANDS2471;  ANDS2471;  ANDX0719;  ANDX5368;  ANDS2471;  ANDS4781;  ANES2220;  ANEQ2749;  ANDW2011;  ANEP3916;  ANEQ2749;  ANES1048;  ANES1048;  ANDX5328;  ANDW2011;  ANDS2471;  ANDY1233;  ANEN2138;  ANES2220;  ANDX5368;  ANEQ2749;  ANDS2471;  ANER0525;  ANDX5368;  ANEQ2749;  ANEQ2749;  ANDX5328;  ANEN2138;  ANEQ1369;  ANDW1483;  ANDW4416;  ANDX3068;  ANDS4781;  ANDV0696;  ANDX5328;  ANEN2138;  ANEQ2749;  ANER0525;  ANCT2435;  ANDQ0328;  ANDQ0328;  ANDS4781;  ANER0525;  ANDP3989;  ANDZ1356;  ANER0525;  ANDX5368;  ANDS4781;  ANDX5328;  ANEQ2749;  ANDS2471;  ANEP3916;  ANES1048;  ANDX5328;  ANEN2138;  ANDS4781;  ANDW2011;  ANDZ1356;  ANER1538;  ANDP3989;  ANDQ0328;  ANEQ4973;  ANER1538;  ANDS2471;  ANEP3916;  ANEZ3422;  ANDP3989;  ANDQ0328;  ANDQ0328;  ANDQ0328;  ANDQ0328;  ANDV0696;  ANDW1483;  ANDW4416;  ANDW4416;  ANEZ0932;  ANEZ2106;  ANEZ3422;  ANEZ3422;  ANEZ3422;  ANEN2138;  ANEN2138;  ANEP2425;  ANEP3916;  ANEP3916;  ANEQ1369;  ANDQ0328;  ANDS4781;  ANDX3068;  ANDX5328;  ANDX5328;  ANEQ1369;  ANER0525;  ANER1538;  ANER1538;  ANES1048;  ANDQ0328;  ANEQ4973;  ANDS2471;  ANDS4781;  ANDX5328;  ANDS4781;  ANDS4781;  ANEP3916;  ANEP3916;  ANDX5368;  ANDZ1356;  ANDX5368;  ANES1048;  ANCN2262;  ANCP0220;  ANEN2138;  ANER0525;  ANDS4781;  ANDW2011;  ANDX0719;  ANDX5328;  ANDX5368;  ANDY1233;  ANEN2138;  ANEN2138;  ANEN2138;  ANET0560;  ANET0560;  ANEZ3422;  ANCS3351;  ANDS4781;  ANDX5328;  ANDZ1356;  ANER0525;  ANDX5328;  ANDS4781;  ANEQ4973;  ANET0560;  ANDW2011;  ANET0560;  ANDX5328;  ANDY1233;  ANEN2138;  ANEP3916;  ANEQ4973;  ANER0525;  ANES1048;  ANES1048;  ANES2220;  ANDX5368;  ANCT2435;  ANDZ1356;  ANES1048;  ANDX5368;  ANDZ1356;  ANEQ2749;  ANEQ4973;  ANER0525;  ANER0525;  ANER1538;  ANET0560;  ANEZ0932;  ANEZ3422;  ANFN1452;  ANDW4416;  ANDZ1356;  ANFP2342;  ANEN2138;  ANDQ0327;  ANDU4076;  ANDX0712;  ANDZ2299;  ANEN3381;  ANEP3914;  ANER1511;  ANES4096;  ANFP3415;  ANER0476;  ANER0476;  ANEP2437;  ANER0476;  ANER0476;  ANDX0712;  ANEP2437;  ANER0476;  ANDX0712;  ANDQ0327;  ANER0476;  ANES1049;  ANEP2437;  ANDQ3042;  ANDW2006;  ANDQ0327;  ANDX5364;  ANES4096;  ANEN2154;  ANES1049;  ANDX5364;  ANEP2437;  ANDX0712;  ANDX0712;  ANDU4076;  ANDU4076;  ANCX1870;  ANDY1232;  ANEP2437;  ANDU4076;  ANDZ2299;  ANER0476;  ANEP5316;  ANDQ0327;  ANDQ0327;  ANES4096;  ANES4096;  ANDN0823;  ANDN3151;  ANDQ0327;  ANDQ0327;  ANDQ0327;  ANDQ3042;  ANDU4076;  ANDU4076;  ANDW1479;  ANDW2006;  ANDX0712;  ANDX5364;  ANDX5364;  ANER0476;  ANFP3415;  ANFP3415;  ANFQ1015;  ANEP5316;  ANEP5316;  ANEQ1370;  ANEQ2750;  ANCX2844;  ANDX0712;  ANDX2785;  ANDX5364;  ANER0476;  ANER0476;  ANDQ0327;  ANDU4076;  ANCS0658;  ANDX0712;  ANDU4076;  ANEP5316;  ANES4096;  ANCP1862;  ANER0476;  ANEP2437;  ANDU4076;  ANDW2006;  ANDW2006;  ANDX2785;  ANDX5364;  ANDY1232;  ANDY1232;  ANDY1232;  ANDZ2299;  ANDZ2299;  ANEN2154;  ANEN3381;  ANES1049;  ANES1049;  ANES4096;  ANDX0712;  ANDX0712;  ANDZ2299;  ANER1511;  ANFP3415;  ANDX0712;  ANER0476;  ANDZ2299;  ANDX0712;  ANDZ2299;  ANER1511;  ANDQ3042;  ANDU4076;  ANDX0712;  ANDX5364;  ANDX5364;  ANDZ2299;  ANEP3914;  ANEP5316;  ANEP5316;  ANEQ1370;  ANEQ2750;  ANER0476;  ANER0526;  ANER1511;  ANES1049;  ANFP2343;  ANDQ3062;  ANEN1069;  ANEN1069;  ANEN1069;  ANEN1069;  ANDQ3062;  ANCY4321;  ANEN1069;  ANEN1069;  ANDY3278;  ANCY4321;  ANDQ3062;  ANDQ3062;  ANDQ3062;  ANDY3278;  ANDQ3062;  ANDQ3062;  ANEN1069;  ANDQ3062;  ANEN3401;  ANEP2419;  ANDQ3062;  ANDQ3062;  ANEN1069;  ANDQ3062;  ANEN3401;  ANDQ3062;  ANDQ3062;  ANFP2343;  ANDQ3062;  ANDQ3062;  ANEP2419;  ANDQ3062;  ANEP2419;  ANDQ3062;  ANDQ3062;  ANDW2022;  ANDW4350;  ANDW4350;  ANDW4350;  ANDY3278;  ANDY3278;  ANDY3278;  ANEN1069;  ANEP2419;  ANDQ3062;  ANCR0707;  ANCY3329;  ANEN1069;  ANDQ3062;  ANDY3278;  ANCY4321;  ANDY3278;  ANEN1069;  ANDQ3895;  ANEQ4974;  ANEQ4974;  ANES4097;  ANDX0740;  ANER0527;  ANER0527;  ANCW1445;  ANEP5314;  ANCX2733;  ANDQ3895;  ANDW2025;  ANEQ4974;  ANER0527;  ANEN1085;  ANES4097;  ANDX0740;  ANEN1085;  ANDQ3895;  ANCX2733;  ANDQ3895;  ANCW1445;  ANER0527;  ANDQ3895;  ANDW2025;  ANDX0740;  ANDY2466;  ANDY3281;  ANEN1085;  ANEP5314;  ANDQ3895;  ANDQ3895;  ANEP5314;  ANEP5314;  ANEQ1371;  ANEQ2751;  ANEQ4974;  ANER0527;  ANDY3307;  ANEN4889;  ANCW1446;  ANDP0517;  ANDY3307;  ANCQ3795;  ANDX5380;  ANDY3307;  ANDY3307;  ANDP0517;  ANDQ3893;  ANDQ3893;  ANDQ3893;  ANDQ3893;  ANEQ4075;  ANFN1453;  ANDP0517;  ANDQ3893;  ANDY3307;  ANDX5380;  ANCQ3795;  ANDY3307;  ANDP0517;  ANDX0752;  ANDX5380;  ANCQ3795;  ANEN4889;  ANEQ4075;  ANEQ4075;  ANEQ4075;  ANDQ3893;  ANDQ3893;  ANDY3307;  ANEN4889;  ANDX5380;  ANFN1453;  ANDQ3893;  ANEQ4075;  ANDP3991;  ANDQ3054;  ANDS1297;  ANDX1414;  ANDX1414;  ANDX5356;  ANDX5356;  ANDZ1340;  ANDZ1340;  ANEN1078;  ANEN1078;  ANEN4872;  ANEN4872;  ANEQ2726;  ANES2203;  ANEU0742;  ANEW3372;  ANEW3372;  ANEZ0917;  ANDP3991;  ANDP3991;  ANEP5320;  ANEQ4056;  ANCY1573;  ANEQ4056;  ANCW3267;  ANES2203;  ANCY1573;  ANEP5320;  ANEQ4056;  ANES2203;  ANEP5320;  ANDS1297;  ANDS1297;  ANCY1573;  ANDS1297;  ANES2203;  ANES2203;  ANEN1078;  ANDP3991;  ANDQ3054;  ANDR3260;  ANDR3260;  ANDS1297;  ANDX5356;  ANDZ1340;  ANEP5320;  ANEU0742;  ANEX4718;  ANEX4718;  ANEZ0917;  ANEZ0917;  ANDS4835;  ANDS4835;  ANEZ0917;  ANDS1297;  ANDX5356;  ANEN4872;  ANEP5320;  ANEP5320;  ANEQ2726;  ANES2203;  ANEP5320;  ANDX5356;  ANDZ1340;  ANEW3372;  ANEN1078;  ANEZ0917;  ANCW3267;  ANEP5320;  ANEU0742;  ANDX5356;  ANEQ2726;  ANDS1297;  ANEN1078;  ANEP5320;  ANEQ4056;  ANEX4718;  ANDS4835;  ANDZ1340;  ANEN1078;  ANDS1297;  ANEX4718;  ANDS4835;  ANCY1573;  ANDQ3054;  ANDP2702;  ANDS4835;  ANDS4835;  ANDS1297;  ANES2203;  ANEU0742;  ANEQ2726;  ANDS1297;  ANEN1078;  ANDX5356;  ANEX4718;  ANDS4835;  ANES2203;  ANDR0756;  ANDS4780;  ANDV2512;  ANDV4539;  ANDX1377;  ANDX1377;  ANDX1409;  ANDY3297;  ANEN2148;  ANET3791;  ANEW3419;  ANEX6158;  ANEY2029;  ANDX1377;  ANEQ4057;  ANDP0514;  ANEX3088;  ANCX0895;  ANEQ1351;  ANEQ4959;  ANDV2512;  ANEN4893;  ANEQ4959;  ANDR0756;  ANDV4539;  ANEQ4959;  ANDV2512;  ANDP0514;  ANDR0756;  ANDS4780;  ANDS4780;  ANEX3088;  ANEY2029;  ANDV2512;  ANDV0704;  ANCX0895;  ANDR0756;  ANEQ4959;  ANEQ4959;  ANEQ4959;  ANEQ4959;  ANDX1377;  ANDX1377;  ANDX1409;  ANDS0755;  ANDX1409;  ANDS4780;  ANDS4780;  ANDX1409;  ANEQ4959;  ANES1036;  ANDT1230;  ANDV4539;  ANEY2029;  ANDY3297;  ANDN3154;  ANDZ1337;  ANDW4342;  ANES1036;  ANES2204;  ANDP0514;  ANDQ3047;  ANDX1377;  ANDX5367;  ANDN3154;  ANDN3154;  ANDQ3047;  ANDR0756;  ANDS0755;  ANDS0755;  ANDS4780;  ANDS4780;  ANDU3097;  ANDV0264;  ANDV0704;  ANDV0704;  ANDV2512;  ANDV2512;  ANDV4539;  ANDW4342;  ANDX1409;  ANDX5367;  ANDY3297;  ANDY3297;  ANDY3297;  ANDZ1337;  ANDZ1337;  ANEN2148;  ANEN2148;  ANEQ1351;  ANEQ1351;  ANEQ1351;  ANEQ4057;  ANEQ4057;  ANEQ4959;  ANES2204;  ANEU3304;  ANEU3304;  ANEX3088;  ANEX6158;  ANEQ1351;  ANEX3088;  ANDS0755;  ANDX5367;  ANEX6158;  ANDT1230;  ANDX1409;  ANEQ4959;  ANDR0756;  ANDW4342;  ANEN2148;  ANEX3088;  ANEY2029;  ANDT1230;  ANEX3088;  ANEX3088;  ANDV2512;  ANDR0756;  ANDR0756;  ANDS0755;  ANDS4780;  ANDV0264;  ANDV0704;  ANDW4342;  ANDX1409;  ANEQ1351;  ANEU3304;  ANDR0756;  ANDW4342;  ANDZ1337;  ANEN2148;  ANEN4893;  ANDS4780;  ANDT1230;  ANEU3304;  ANEN4893;  ANDP0514;  ANDZ1337;  ANEQ4959;  ANEN2148;  ANEU3304;  ANEU3304;  ANEY2029;  ANDY3297;  ANDY3297;  ANDP0514;  ANDV0264;  ANEN2148;  ANEN4893;  ANEN4893;  ANET3791;  ANEU3304;  ANEY2029;  ANEU3304;  ANDS4780;  ANDZ1337;  ANEN4893;  ANEU3304;  ANEX6158;  ANEY2029;  ANDW4342;  ANDN3154;  ANDQ3047;  ANDQ3047;  ANDS4780;  ANEN2148;  ANEY2029;  ANEY2029;  ANDT1230;  ANDR0756;  ANDV4539;  ANDT1230;  ANDZ1337;  ANES1036;  ANES2204;  ANES1036;  ANDV2512;  ANES2204;  ANEW3419;  ANEY2029;  ANEZ3434;  ANEN2148;  ANDS0755;  ANDZ1337;  ANDY3297;  ANCX1878;  ANDN3154;  ANDX1377;  ANET3791;  ANDT1230;  ANEY2029;  ANEU3304;  ANDP3977;  ANDR2279;  ANDS1308;  ANDS1308;  ANDS4836;  ANDW4358;  ANDX1384;  ANDX3078;  ANDX5381;  ANDZ1347;  ANDZ2293;  ANEN4873;  ANES4077;  ANET0553;  ANEV1713;  ANEW3373;  ANEX0748;  ANCS0652;  ANDP3977;  ANDP3977;  ANDY1236;  ANDZ2293;  ANEN2146;  ANDS1308;  ANEN4873;  ANCS0652;  ANDS1308;  ANEQ4058;  ANDS1308;  ANCY2195;  ANCS0652;  ANDR2279;  ANDY1236;  ANER1520;  ANDS1308;  ANDV0299;  ANDV3573;  ANDV4542;  ANDX1384;  ANDV0299;  ANDS4836;  ANDX1420;  ANEN4873;  ANEN4873;  ANEW3373;  ANEX0748;  ANEX0748;  ANEX0748;  ANDR2279;  ANDZ1347;  ANEN4873;  ANDV4542;  ANEW3373;  ANER1520;  ANDP3977;  ANEX0748;  ANDS4836;  ANEN4873;  ANES2205;  ANDV0299;  ANDV3573;  ANCV2917;  ANDP3977;  ANEQ4058;  ANDV0299;  ANER1520;  ANER5166;  ANES2205;  ANES4077;  ANDP3994;  ANDQ3060;  ANDR2279;  ANDQ3060;  ANDR2279;  ANDS1308;  ANDS1308;  ANDS4836;  ANDV0299;  ANDV0701;  ANDV3573;  ANDV3573;  ANDV4542;  ANDW4358;  ANDX1384;  ANDX1420;  ANDY1236;  ANDZ1347;  ANDZ1347;  ANDZ1347;  ANEN2146;  ANEP2423;  ANEQ4058;  ANEQ4058;  ANER1520;  ANER1520;  ANER1520;  ANEW3373;  ANEW4227;  ANEW4227;  ANEW4227;  ANEW4227;  ANEX0748;  ANDX1384;  ANDX3078;  ANDV3573;  ANDS4836;  ANDX1384;  ANCV2917;  ANDS4836;  ANDY1236;  ANDY1236;  ANEV1713;  ANEX0748;  ANDS1308;  ANDR2279;  ANDR2279;  ANDZ1347;  ANEN4873;  ANEQ4058;  ANEQ4058;  ANEW3373;  ANDV3573;  ANEX0748;  ANDV0299;  ANDV4542;  ANDR2279;  ANDV0701;  ANDV4542;  ANER1520;  ANEW3373;  ANEX0748;  ANDV4542;  ANDX1420;  ANEN4873;  ANEQ4058;  ANEX0748;  ANCS0652;  ANDX1384;  ANDX1420;  ANEX0748;  ANEW3373;  ANEX0748;  ANDX1420;  ANDX1384;  ANDX3078;  ANER1520;  ANEX0748;  ANDV0299;  ANET0553;  ANDV0299;  ANDV4542;  ANDX1420;  ANDX3078;  ANES4077;  ANEV1713;  ANEW4227;  ANEW4227;  ANEW4227;  ANDS1308;  ANDV0701;  ANDZ1347;  ANEN4873;  ANER1520;  ANEV1713;  ANEX0748;  ANDZ1347;  ANCV2917;  ANCV2917;  ANDR2279;  ANDW4358;  ANDX3078;  ANDX3078;  ANDY1236;  ANEQ4058;  ANEX0748;  ANDV0299;  ANDV0299;  ANEQ4058;  ANES4077;  ANDP3994;  ANES2205;  ANDX1384;  ANDZ1347;  ANDX1384;  ANDX1420;  ANDX3078;  ANDZ1347;  ANDZ1347;  ANEQ4058;  ANDZ2293;  ANDS1308;  ANDS4836;  ANDZ1347;  ANEN2146;  ANER5166;  ANEV1713;  ANFP0875;  ANEN2146;  ANEN4873;  ANDV4542;  ANDP3994;  ANDX1420;  ANEQ4058;  ANEV1713;  ANEW4227;  ANEX0748;  ANDX5304;  ANDQ5238;  ANDX5304;  ANCP1855;  ANCT4319;  ANES4078;  ANET3792;  ANDQ0329;  ANCV1590;  ANDV3570;  ANDX5304;  ANDX5387;  ANDX5387;  ANET3792;  ANCV3486;  ANDX1378;  ANEN2151;  ANEP3920;  ANDX1378;  ANDX5304;  ANDV3570;  ANEW3414;  ANEW3414;  ANDX5304;  ANDX5304;  ANES4078;  ANDV3570;  ANDP3966;  ANDX5304;  ANER4198;  ANDX5304;  ANDX5387;  ANEN3380;  ANDV3570;  ANEW2045;  ANDU1109;  ANCX3899;  ANDX5304;  ANEN3380;  ANDQ5238;  ANDX5304;  ANES1037;  ANDX5387;  ANER4198;  ANES1037;  ANDP3966;  ANDQ0329;  ANDQ0329;  ANDQ3885;  ANDQ3885;  ANDQ5238;  ANDP3966;  ANDQ0329;  ANDQ0329;  ANDU1109;  ANDV0692;  ANDV3570;  ANDV3570;  ANDW4375;  ANDX1378;  ANDX1378;  ANDX5304;  ANDX5304;  ANDX5387;  ANDX5387;  ANEN2151;  ANEN3380;  ANEN4888;  ANEN4888;  ANEP3920;  ANEP3920;  ANEP3920;  ANEQ4059;  ANEQ4059;  ANEW2045;  ANEW2045;  ANEW3414;  ANDX1378;  ANEQ4059;  ANDX5387;  ANEP3920;  ANET3792;  ANET3792;  ANEW2045;  ANDV0692;  ANDX1378;  ANDQ0329;  ANDQ0329;  ANDV0692;  ANDW4375;  ANDX1378;  ANEN2151;  ANES4078;  ANET3792;  ANEW2045;  ANEW3414;  ANES4078;  ANEN4888;  ANEP3920;  ANEW2045;  ANDW4375;  ANEN3380;  ANEN4888;  ANDW4375;  ANDW4375;  ANEN2151;  ANEN3380;  ANDQ0329;  ANEN4888;  ANEX6021;  ANEW2045;  ANEX6021;  ANEX6021;  ANER4198;  ANDX5387;  ANET3792;  ANCP1855;  ANDX5387;  ANDX5304;  ANDX5387;  ANEP3920;  ANES1037;  ANEW2045;  ANEW2045;  ANEW3414;  ANDW4375;  ANDX5387;  ANER4198;  ANES4078;  ANDX1378;  ANDX5304;  ANDU1109;  ANDV0692;  ANDV3570;  ANEW2045;  ANDW4375;  ANES4078;  ANEQ4059;  ANDV3570;  ANEW3414;  ANCV3486;  ANES1037;  ANDX5304;  ANDX5304;  ANDX5304;  ANEP3920;  ANEW3414;  ANEW3414;  ANCV3486;  ANEN2151;  ANFN1439;  ANDX5304;  ANDX5387;  ANEN2151;  ANEN2151;  ANEN3380;  ANEN4888;  ANEQ4059;  ANEW3414;  ANEX6021;  ANCX1879;  ANCX1879;  ANDR3259;  ANDY3286;  ANDZ1328;  ANEQ2727;  ANCN1541;  ANDY2467;  ANES1038;  ANDW4392;  ANDW4392;  ANBU2413;  ANES4079;  ANDY2467;  ANDY2467;  ANCQ2605;  ANEQ4060;  ANCQ2605;  ANDX5317;  ANCN1541;  ANCX1879;  ANEQ2727;  ANEQ4060;  ANCQ2605;  ANDQ1715;  ANCR3118;  ANET0554;  ANDX5317;  ANDS2460;  ANDQ1715;  ANDY2467;  ANDQ3048;  ANDR3259;  ANDS2460;  ANDZ1328;  ANDQ1715;  ANDQ3048;  ANDQ3048;  ANDR3259;  ANDX5317;  ANCX1879;  ANCX1879;  ANDQ3048;  ANDQ3048;  ANDR3259;  ANDS2460;  ANDS2460;  ANDV1257;  ANDW4392;  ANDX5317;  ANDX5317;  ANDY3286;  ANEP5318;  ANEQ2727;  ANEQ2727;  ANEQ4060;  ANEQ4060;  ANES1038;  ANES1038;  ANES1038;  ANES4079;  ANES4079;  ANES4079;  ANET0554;  ANET0554;  ANDS4837;  ANDY2467;  ANDS2460;  ANDS2460;  ANES1038;  ANES1038;  ANES4079;  ANDS2460;  ANDV1257;  ANEX6020;  ANDQ1715;  ANDS2460;  ANDS2460;  ANDS2460;  ANDV1257;  ANDV1257;  ANDW4392;  ANEW3413;  ANEX6020;  ANEX6020;  ANDV1257;  ANDX5317;  ANEW3413;  ANDX5317;  ANDX5317;  ANDX5317;  ANEQ4060;  ANEW3413;  ANDS4837;  ANDV1257;  ANDW4392;  ANDZ1328;  ANEP5318;  ANEP5318;  ANCQ2605;  ANDX5317;  ANDS2460;  ANEQ4060;  ANEQ4060;  ANES1038;  ANDV1257;  ANEQ4060;  ANDZ1328;  ANDX5317;  ANEQ2727;  ANEX6020;  ANEX6020;  ANDS2460;  ANCX1879;  ANDQ1715;  ANDS4837;  ANDY2467;  ANET0554;  ANEW3413;  ANFP0910;  ANDS2460;  ANEW3413;  ANEX6020;  ANCX3900;  ANCX3900;  ANDP3963  ANDP3963;  ANDP3963;  ANDP3963;  ANDS4838;  ANDV0686;  ANEQ2728;  ANCV2919;  ANDS4779;  ANEP5319;  ANDS4779;  ANES2206;  ANCV2919;  ANEP5319;  ANDP3963;  ANDR2274;  ANDR4200;  ANDW1482;  ANDW4410;  ANDX1376;  ANCV2919;  ANDR2274;  ANDR4200;  ANDS4779;  ANDS4838;  ANDV0686;  ANEP5319;  ANEQ2728;  ANEQ2728;  ANES2206;  ANES2206;  ANES2206;  ANEQ2728;  ANEQ2728;  ANES2206;  ANDP3963;  ANDP3963;  ANDR4200;  ANDR4200;  ANDR4200;  ANEP5319;  ANFN3970;  ANFN3970;  ANFN3970;  ANDW1482;  ANDS4838;  ANCV2919;  ANDS4838;  ANDV0686;  ANEP5319;  ANEQ2728;  ANDR2274;  ANCV2919;  ANDS4779;  ANDS4779;  ANDS4779;  ANEP5319;  ANES2206;  ANES2206;  ANDS4779;  ANDV0686;  ANDS4779;  ANDS4779;  ANCX1887;  ANCX1887;  ANCY2197;  ANCY2197;  ANDQ1718;  ANDQ1718;  ANDX1410;  ANDX5377;  ANDX5377;  ANCT0396;  ANER5150;  ANDW1484;  ANDW4344;  ANDQ3049;  ANCT0396;  ANDX5377;  ANCT0396;  ANDX1410;  ANDQ3049;  ANDW4344;  ANCT0396;  ANDW4344;  ANDW4344;  ANDX1410;  ANDX5377;  ANDX5377;  ANER5150;  ANEX0750;  ANDX5377;  ANCY2197;  ANCY2197;  ANDQ1718;  ANDQ3049;  ANDS3978;  ANDS3978;  ANDS4782;  ANDV0705;  ANDV0705;  ANDW1484;  ANEX1710;  ANEX3095;  ANDW4344;  ANDX5377;  ANDX1410;  ANER5150;  ANDX1410;  ANCN3585;  ANDW4344;  ANDS3978;  ANDW1484;  ANDW1484;  ANDW1484;  ANDX5377;  ANER5150;  ANES4081;  ANDV0705;  ANDX5377;  ANER5150;  ANDQ3049;  ANDX5377;  ANEX1710;  ANDW1484;  ANDX1410;  ANDW1484;  ANDW1484;  ANDX5377;  ANEX0750;  ANEX1710;  ANEX1710;  ANEX3095;  ANEQ4076;  ANDQ3056;  ANDQ3056;  ANDQ3892;  ANDQ3892;  ANDQ3892;  ANDY4970;  ANDZ1344;  ANDZ1344;  ANEP3921;  ANEQ4076;  ANES4098;  ANES4098;  ANEZ0933;  ANEP3921;  ANER0507;  ANDY2451;  ANEQ4975;  ANCX3889;  ANEQ4076;  ANDZ1344;  ANER0507;  ANCR1291;  ANDV0697;  ANCR1291;  ANDQ3056;  ANDQ3892;  ANDY2451;  ANEQ4975;  ANEP3921;  ANDY2451;  ANDZ1344;  ANEP3921;  ANDQ3056;  ANDQ3056;  ANDQ3892;  ANDV0697;  ANDV0697;  ANDY2451;  ANEZ0933;  ANEZ0933;  ANDQ3056;  ANDQ3892;  ANDV0697;  ANDZ1344;  ANEQ4975;  ANER0507;  ANDQ3892;  ANDV0697;  ANES4098;  ANCR1291;  ANEQ4975;  ANDV0697;  ANDY2451;  ANDY4970;  ANDY4970;  ANDY2451;  ANDY4970;  ANDZ1344;  ANEQ4975;  ANDZ1344;  ANEQ4975;  ANEZ0933;  ANEZ0933;  ANFP0914;  ANDQ1729;  ANDQ3064;  ANDV4532;  ANDX2783;  ANDX5370;  ANDZ1350;  ANDZ1350;  ANDZ2303;  ANDZ2303;  ANEN3384;  ANEP2440;  ANEP3918;  ANES4099;  ANES4099;  ANDX5370;  ANEQ4976;  ANEQ4976;  ANEQ4976;  ANDQ3064;  ANDQ1729;  ANDX5370;  ANEN3384;  ANDQ3064;  ANDN0836;  ANDW2026;  ANDZ2303;  ANEN3384;  ANEQ4976;  ANEQ4976;  ANCX3888;  ANDQ3064;  ANDV4532;  ANDZ1350;  ANEN3384;  ANEP3918;  ANEQ4976;  ANDQ3064;  ANDQ3064;  ANDX5370;  ANDZ1350;  ANEQ4976;  ANCX3888;  ANDQ3064;  ANDQ3064;  ANDX5370;  ANDX2783;  ANEQ4976;  ANEQ4976;  ANER0508;  ANEQ4976;  ANEN3384;  ANDX5370;  ANEQ4976;  ANDX5370;  ANEQ4976;  ANDQ3064;  ANDQ3064;  ANDQ3064;  ANDV4532;  ANDV4532;  ANDW2026;  ANDX0766;  ANDX5370;  ANDX5370;  ANFP3412;  ANEP3918;  ANEP5330;  ANEP5330;  ANEQ1372;  ANEQ2752;  ANEQ4976;  ANDQ1729;  ANCX3888;  ANDQ3064;  ANDQ3064;  ANDQ3064;  ANDQ3064;  ANDX0766;  ANDX5370;  ANDQ3064;  ANER0509;  ANDX5370;  ANDZ1350;  ANDZ2303;  ANER0508;  ANDQ3064;  ANER0508;  ANDQ3064;  ANER0508;  ANES4099;  ANEN3384;  ANEQ4976;  ANDW2026;  ANDX0766;  ANDX2783;  ANDX5370;  ANDZ1350;  ANEQ4976;  ANDZ2303;  ANDQ1729;  ANDZ2303;  ANDZ2303;  ANDQ3064;  ANDQ3064;  ANEQ4976;  ANER0509;  ANDQ3064;  ANDW2026;  ANER0508;  ANDX5370;  ANDQ3064;  ANEQ4976;  ANEQ4976;  ANDZ2303;  ANDZ1350;  ANEQ4976;  ANES4099;  ANES4099;  ANDQ3064;  ANDQ3064;  ANEQ4976;  ANEQ4976;  ANEQ4976;  ANDZ2303;  ANER0508;  ANDZ2277;  ANDP3982;  ANDQ3065;  ANDQ3065;  ANDV4527;  ANDX0724;  ANDX0724;  ANDX5351;  ANDY3292;  ANDY3292;  ANDZ2277;  ANDZ2277;  ANEP3933;  ANER1513;  ANER1513;  ANES4100;  ANES4100;  ANFP2344;  ANDQ3065;  ANDZ2277;  ANEP2439;  ANER0506;  ANDZ2277;  ANDQ3065;  ANCP0221;  ANDP3982;  ANDZ2277;  ANDX5351;  ANDQ3065;  ANDY2468;  ANER0506;  ANDP3982;  ANDQ3065;  ANDZ1333;  ANDZ1333;  ANER0506;  ANER0506;  ANEQ2753;  ANDQ3065;  ANER0506;  ANFP2344;  ANEQ1373;  ANEQ2753;  ANER1513;  ANDZ1333;  ANDZ2277;  ANDQ3065;  ANER0506;  ANDZ2277;  ANEP2439;  ANEQ2753;  ANEQ4977;  ANDP3982;  ANER0506;  ANDQ3065;  ANDQ3065;  ANDY2468;  ANEP2439;  ANEQ2753;  ANEQ1373;  ANER0506;  ANER0506;  ANER0506;  ANDV4527;  ANDW4366;  ANDX0724;  ANDZ1333;  ANFP2344;  ANEP3933;  ANEQ1373;  ANEQ1373;  ANEQ2753;  ANDP3982;  ANDP3982;  ANDQ3065;  ANDY2468;  ANDQ3065;  ANES4100;  ANEQ4977;  ANEP2439;  ANER0506;  ANDZ1333;  ANDZ1333;  ANER1513;  ANDV4527;  ANDV4527;  ANDQ3065;  ANDZ2277;  ANER0506;  ANDV4527;  ANDW4366;  ANDX0724;  ANDX5351;  ANDY3292;  ANDZ1333;  ANER0506;  ANDY3292;  ANDZ1333;  ANER1513;  ANDY2468;  ANDZ1333;  ANES4100;  ANDZ1333;  ANDZ2277;  ANEP2439;  ANEP2439;  ANDQ3065;  ANER0506;  ANER1513;  ANES4100;  ANDZ2277;  ANDY2468;  ANDP3982;  ANEQ4077;  ANEQ4977;  ANER0506;  ANFP2344;  ANCW1448;  ANDY2455;  ANDY3285;  ANDY3285;  ANEP5310;  ANCW1448;  ANEP5310;  ANER1527;  ANER1527;  ANDY3285;  ANCX2735;  ANDY2455;  ANER1527;  ANDY2455;  ANEU0743;  ANEP5310;  ANEP5310;  ANEP5310;  ANDY2455;  ANDY2455;  ANCW1448;  ANCP0222;  ANCX2735;  ANDQ3896;  ANER1527;  ANEP5310;  ANER1527;  ANER1527;  ANER1527;  ANDQ3896;  ANDW1490;  ANDW1490;  ANDY2455;  ANDY2455;  ANDY3285;  ANER1527;  ANET2532;  ANET2532;  ANET2532;  ANDQ3896;  ANDY2455;  ANER1527;  ANER1527;  ANER1527;  ANCX2735;  ANDQ3896;  ANDY3285;  ANDY3285;  ANCP0222;  ANCR3435;  ANCX2735;  ANCW1448;  ANEU0743;  ANER1527;  ANCX2735;  ANDW1490;  ANEP5310;  ANDQ3896;  ANDW1490;  ANDW2027;  ANDY2455;  ANDY3285;  ANDY3285;  ANEP5310;  ANDW1490;  ANET2532;  ANDW1490;  ANDY2455;  ANEP0874;  ANER1527;  ANDQ3896;  ANDY2455;  ANEP0874;  ANEP5310;  ANEP5310;  ANEP5310;  ANEP5310;  ANEP5310;  ANDZ2294;  ANES2221;  ANDZ2294;  ANDW2024;  ANDZ2294;  ANCY3341;  ANDW2024;  ANCR1293;  ANEP0876;  ANDZ2294;  ANES2221;  ANES2221;  ANDZ2294;  ANEP0876;  ANES2221;  ANCY3341;  ANDW2024;  ANDZ2294;  ANEP5331;  ANEP5331;  ANES2221;  ANDW2024;  ANEP5331;  ANCY3341;  ANEP0876;  ANDY3296;  ANEN3399;  ANEN3399;  ANEN3399;  ANEN3399;  ANEP2438;  ANDY3296;  ANDY3296;  ANDY3296;  ANDU4029;  ANDU4029;  ANDU4029;  ANDY2450;  ANDY2450;  ANDU4029;  ANDY2450;  ANDY2450;  ANDY3296;  ANEP2438;  ANFP2345;  ANDY2450;  ANDY3296;  ANDQ3881;  ANEP0875;  ANEP0875;  ANDV0253;  ANEP0875;  ANEP5315;  ANDQ3881;  ANDQ3881;  ANEP5315;  ANEP5315;  ANFN3988;  ANDQ3881;  ANEP0875;  ANDV0253;  ANCU2731;  ANDQ3881;  ANCU2731;  ANDQ3881;  ANEP0875;  ANDW2005;  ANDQ3881;  ANDQ3881;  ANDW2005;  ANDW2005;  ANES2222;  ANDV0253;  ANDW2005;  ANEP0875;  ANEQ2754;  ANDQ3881;  ANDV0253;  ANEP0875;  ANDR3255;  ANDU3353;  ANDU3353;  ANDU3353;  ANDU3353;  ANDX5313;  ANDX5313;  ANDZ1343;  ANDZ1343;  ANDZ1343;  ANDZ1343;  ANDZ2300;  ANET1858;  ANET1859;  ANET1859;  ANET1859;  ANEW3374;  ANEW3374;  ANEX6017;  ANEX6017;  ANCT4327;  ANCT4327;  ANCT4327;  ANCW4223;  ANDP3980;  ANDP3980;  ANDZ1343;  ANDZ2300;  ANEW3374;  ANCT4327;  ANCW4223;  ANEZ3439;  ANET1859;  ANET1859;  ANCT4327;  ANCT3617;  ANDZ1343;  ANEW3374;  ANDR3255;  ANDZ1343;  ANDZ1343;  ANDX5313;  ANDX5313;  ANDX5313;  ANET1859;  ANET1859;  ANEX6017;  ANEZ3439;  ANDU3353;  ANDZ1343;  ANEZ3439;  ANDR3255;  ANET1859;  ANEZ3439;  ANBZ0424;  ANDR3255;  ANDU3353;  ANDR3255;  ANET1858;  ANDR3255;  ANET1859;  ANCW4223;  ANEZ3439;  ANDZ2300;  ANEZ3439;  ANET1859;  ANEZ3439;  ANDQ3059;  ANDS1324;  ANDS1324;  ANDT2896;  ANDT2896;  ANDT2896;  ANDV0699;  ANDX1419;  ANDX5369;  ANDY4967;  ANEN3395;  ANEP5327;  ANER1521;  ANES4082;  ANEX0755;  ANEX3096;  ANCV1591;  ANCX3901;  ANDP3976;  ANDQ3059;  ANDQ3059;  ANEN3395;  ANCV1591;  ANER1521;  ANDZ2295;  ANEQ1352;  ANER5119;  ANCQ4292;  ANDU1112;  ANER5119;  ANEW3417;  ANEQ4061;  ANDT2896;  ANEQ4061;  ANEQ4061;  ANCQ4292;  ANDX5369;  ANDU1112;  ANEQ4061;  ANDU1112;  ANCT1143;  ANDU1112;  ANCU1840;  ANDT2896;  ANEQ1352;  ANER1521;  ANDX5369;  ANDY4967;  ANDP3976;  ANDT2896;  ANDT2896;  ANDX1383;  ANDX1383;  ANDX1383;  ANDX1383;  ANDX1419;  ANDX5369;  ANDX5369;  ANDX5369;  ANDY4967;  ANDZ2295;  ANEQ4061;  ANEQ4061;  ANEQ4061;  ANEQ4061;  ANER5119;  ANDT2896;  ANDX1383;  ANDS1324;  ANDT2896;  ANEQ4061;  ANDZ2295;  ANDY4967;  ANDU1112;  ANEQ4061;  ANER1521;  ANER5119;  ANES4082;  ANDQ3059;  ANDY4967;  ANEQ4061;  ANDT2896;  ANDX1419;  ANDX5369;  ANDZ2295;  ANEQ1352;  ANEX0755;  ANDX1383;  ANER1521;  ANDQ3059;  ANDS1324;  ANEP5327;  ANEQ1352;  ANEW3417;  ANEX0755;  ANEP5327;  ANEX0755;  ANCS2429;  ANCY3336;  ANEQ4061;  ANER1521;  ANEX3096;  ANES4082;  ANDS1324;  ANDY4967;  ANDY4967;  ANDT2896;  ANDX5369;  ANEX0755;  ANDS1324;  ANDZ2295;  ANEQ1352;  ANEQ4061;  ANES4082;  ANDT2896;  ANDU1112;  ANDV0699;  ANDY4967;  ANEQ4061;  ANDT2896;  ANER5119;  ANDQ3059;  ANDY4967;  ANDS1324;  ANDU1112;  ANER5119;  ANDT2896;  ANER1521;  ANDS1324;  ANER1521;  ANEN3395;  ANEX3096;  ANDX5369;  ANEX0755;  ANER1521;  ANFP2321;  ANER5119;  ANDS0758;  ANDT4132;  ANDU4060;  ANDX1413;  ANDX1413;  ANDX1413;  ANDX5355;  ANDX5355;  ANDZ2298;  ANEN3406;  ANEP3922;  ANER2793;  ANER5135;  ANES1039;  ANES4083;  ANET3793;  ANEX1703;  ANDS0758;  ANDY3293;  ANDV1210;  ANEU3274;  ANDV1210;  ANDU4060;  ANDV1210;  ANDX5355;  ANDX5355;  ANDX5355;  ANDX5355;  ANDX1413;  ANER1522;  ANEX4719;  ANDS0758;  ANEQ4062;  ANEQ4062;  ANER1522;  ANEX1703;  ANEY2030;  ANEU3274;  ANDX5355;  ANES4083;  ANDX5355;  ANDY3293;  ANDV1210;  ANDV0693;  ANDS0758;  ANDX5355;  ANEP3922;  ANDV1210;  ANEQ4062;  ANER2793;  ANDP3968;  ANDQ3053;  ANDS0758;  ANDQ3053;  ANDQ3889;  ANDQ3889;  ANDU4060;  ANDV0693;  ANDV0693;  ANDV0693;  ANDV1210;  ANDX5355;  ANDY3293;  ANDZ2298;  ANEP0859;  ANEP0859;  ANEP0859;  ANEP3922;  ANEQ4062;  ANER1522;  ANER1522;  ANES1039;  ANET3793;  ANET3793;  ANEU3274;  ANEU3274;  ANEU3274;  ANDV1210;  ANDV0693;  ANEY2030;  ANDZ2298;  ANDT4132;  ANES1039;  ANEX4719;  ANDV1210;  ANDY3293;  ANDY3293;  ANER2793;  ANES4083;  ANDQ3889;  ANDX5355;  ANDZ2298;  ANER1522;  ANES1039;  ANDV1210;  ANDS0758;  ANEX4719;  ANDV0693;  ANES1039;  ANET3793;  ANEU3274;  ANEY2030;  ANES4083;  ANDX5355;  ANDS0758;  ANDX5355;  ANDX5355;  ANDP3968;  ANEP3922;  ANES1039;  ANEU3274;  ANEY2030;  ANDZ2298;  ANEX4719;  ANER1522;  ANDX5355;  ANDP3968;  ANDS0758;  ANDV0693;  ANDY3293;  ANDT4132;  ANEP3922;  ANEQ4062;  ANDT4132;  ANEX4719;  ANDP3968;  ANDV1210;  ANES4083;  ANDP3968;  ANDV0693;  ANER2793;  ANEQ4062;  ANER1522;  ANDP3968;  ANES1039;  ANEQ4062;  ANEX1703;  ANEY2030;  ANEY2030;  ANDY3293;  ANEN3406;  ANES1039;  ANEZ0918;  ANEZ2093;  ANDP3968;  ANDY3293;  ANEP0859;  ANER1522;  ANEX4719;  ANDQ3050;  ANDQ3050;  ANDS1271;  ANDS1271;  ANDS1271;  ANDS1271;  ANDV4531;  ANDW4359;  ANDW4359;  ANDX5310;  ANDX5310;  ANDY3302;  ANDY3302;  ANDZ2306;  ANER1523;  ANES4084;  ANEZ0919;  ANCS0653;  ANCW3269;  ANDY3302;  ANCS0653;  ANER1523;  ANDS1271;  ANDV4531;  ANDX5310;  ANDZ2306;  ANEP3929;  ANER1523;  ANDV4531;  ANDX1411;  ANEP5333;  ANEQ2729;  ANEQ2729;  ANCS0653;  ANEQ2729;  ANDX1411;  ANDX5310;  ANDP0515;  ANDW4359;  ANDV4531;  ANEP5333;  ANDX1411;  ANEN3382;  ANDQ1721;  ANDQ3050;  ANDQ3050;  ANCU1849;  ANCW3269;  ANDQ1721;  ANDQ3050;  ANDV4531;  ANDV4531;  ANDW4359;  ANDX1411;  ANEN3382;  ANEP3929;  ANEP5333;  ANEP5333;  ANEQ2729;  ANDV4531;  ANCV1594;  ANDZ2306;  ANEP5333;  ANDY2459;  ANDZ2306;  ANDV4531;  ANDY3302;  ANDY3302;  ANER1523;  ANEW3375;  ANEZ0919;  ANDQ3050;  ANDX5310;  ANDY2459;  ANER5151;  ANEQ2729;  ANDS1271;  ANER5151;  ANDQ1721;  ANDQ3050;  ANDQ3050;  ANDV4531;  ANDV4531;  ANDV4531;  ANEP5333;  ANER5151;  ANEZ0919;  ANDP0515;  ANDS1271;  ANDS1271;  ANEP5333;  ANEW3375;  ANEW3375;  ANEP3929;  ANER5151;  ANER1523;  ANER1523;  ANCS0653;  ANEQ2729;  ANDY2459;  ANEQ2729;  ANEW3375;  ANDV4531;  ANDV4531;  ANDY3302;  ANER1523;  ANDX1411;  ANEQ2729;  ANES4084;  ANEN3382;  ANEQ2729;  ANER5151;  ANDV4531;  ANDS1271;  ANDX1411;  ANER1523;  ANEW3375;  ANEZ0919;  ANDZ2306;  ANEN3382;  ANEP3929;  ANEP3929;  ANDY3302;  ANEZ0919;  ANDS1285;  ANDS1285;  ANDT1231;  ANDV0689;  ANDX5357;  ANDZ2309;  ANDZ2309;  ANEP0851;  ANEP0851;  ANEP0851;  ANEP5322;  ANER1524;  ANEZ2094;  ANEZ2094;  ANDS1285;  ANEQ4063;  ANER1524;  ANDX5357;  ANDX5357;  ANDW4376;  ANDT1231;  ANEQ4063;  ANDP3993;  ANDR2277;  ANDS1285;  ANDX5357;  ANDP3972;  ANDT1231;  ANDV0689;  ANDW4376;  ANDX5357;  ANDX5357;  ANDY3306;  ANEP0851;  ANEQ4063;  ANEQ4063;  ANEQ4063;  ANDV0689;  ANEQ4063;  ANDW4376;  ANDW4376;  ANDW4376;  ANEP0851;  ANES1040;  ANDW4376;  ANEQ4063;  ANDX5357;  ANDT1231;  ANES1040;  ANDX5357;  ANEQ4063;  ANEQ4063;  ANDT1231;  ANEP0851;  ANEQ4063;  ANCW3313;  ANDT1231;  ANEQ4063;  ANEQ4063;  ANDY3306;  ANEQ4063;  ANEQ4063;  ANEQ4063;  ANEQ4063;  ANDY3306;  ANDX5357;  ANEZ2094;  ANDS2469;  ANDS2469;  ANDX5378;  ANDS2469;  ANDX5378;  ANES2207;  ANDS2469;  ANDS2469;  ANDN0607;  ANDS2469;  ANES2207;  ANDW4393;  ANDS2469;  ANDX5378;  ANES2207;  ANDW4393;  ANDN0607;  ANDN0607;  ANDX5378;  ANDN0607;  ANDN0607;  ANDS2469;  ANDS2469;;  ANDX5378;  ANDS2469;  ANDS2469;  ANDW4393;  ANES2207;  ANDS2469;  ANES2207;  ANDX5378;  ANDX5378;  ANCU0856;  ANDS2469;  ANDS2469;  ANES2207;  ANDX5378;  ANDQ3057;  ANDS2481;  ANDU1110;  ANDU1866;  ANDX1369;  ANDX1369;  ANDY3295;  ANDY3295;  ANDZ1335;  ANEN3386;  ANEP0863;  ANES1041;  ANES4085;  ANEW3376;  ANEW4228;  ANEX0678;  ANEX1698;  ANEZ0920;  ANDS2481;  ANDU1110;  ANDU1866;  ANDU1866;  ANEX1698;  ANEP2418;  ANDS2481;  ANDS2481;  ANEP0863;  ANDY3295;  ANDY3295;  ANDQ0332;  ANDU1866;  ANDY3295;  ANEZ0920;  ANEP2418;  ANDU1866;  ANDQ0332;  ANDQ3057;  ANDQ3057;  ANDS2481;  ANDU1866;  ANDX1369;  ANDX1369;  ANEP2418;  ANEQ4960;  ANEW3376;  ANEW4228;  ANEX1698;  ANEZ0920;  ANEZ0920;  ANDU1110;  ANDU1866;  ANEZ0920;  ANEX0678;  ANEX1698;  ANES4085;  ANDS2481;  ANDU1110;  ANDS2481;  ANEX1698;  ANEP0863;  ANEW3376;  ANDQ0332;  ANDX1369;  ANES4085;  ANEZ0920;  ANEZ0920;  ANEP2418;  ANEX0678;  ANEN3386;  ANDX1369;  ANDZ1335;  ANEQ4064;  ANEW3376;  ANEW4228;",
      "recall_initiation_date": "20201214",
      "postal_code": "85281-2438",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}