{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Portage",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65486",
      "recalling_firm": "Stryker Instruments Div. of Stryker Corporation",
      "address_1": "4100 E Milham Ave",
      "address_2": "N/A",
      "postal_code": "49002-9704",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - USA Nationwide including the states of CA, CO, FL, HI, IA, IL, KS, KY, LA, MD, MI, MN, MS, NC, NH, NJ, NV, NY, OH, PA, TX, VA, ITALY, and the countries of AUSTRALIA, CANADA, UNITED KINGDOM, NETHERLANDS, INDIA, JAPAN, NEW ZEALAND, SWITZERLAND, HONG KONG, and TAIWAN.",
      "recall_number": "Z-1878-2013",
      "product_description": "Sonopet Ultrasonic Aspirator Console  (Console 110V with Pedal & Pole)    Product Usage:  The Sonopet Console is indicated for use in surgical procedures where fragmentation, emulsification, and aspiration of soft and hard tissue is desirable, including neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, plastic and reconstructive surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and thoracoscopic surgery.",
      "product_quantity": "215 units total",
      "reason_for_recall": "A supplied component exhibiting corrosion was built into the power supply boards of the Sonopet Consoles listed above. The  issue is limited to 2 lots of the supplied component. The presence of the corrosion on the relay of the power supply board can  potentially lead to a conduction defect which could cause the irrigation pump to stop working. The purpose of the irrigation  pump is to supply sali",
      "recall_initiation_date": "20130624",
      "center_classification_date": "20130805",
      "termination_date": "20140130",
      "report_date": "20130814",
      "code_info": "5450-850-000 (Console 110V with Pedal & Pole)  S/N  1131300134, 1131900074, 1200500134, 1203200244, 1210100114, 1131300144, 1131900084, 1200500144, 1203200254, 1210100124, 1131300154, 1131900094, 1200500154, 1203200274, 1210100134, 1131300164, 1131900104, 1200500164, 1203700954, 1210100144, 1131300174, 1131900114, 1200500174, 1203700964, 1210100154, 1131300184, 1131900124, 1200500184, 1203700974, 1210700014, 1131300194, 1132800114, 1200500194, 1203900024, 1210700024, 1131300204, 1132800124, 1200500204, 1204000044, 1210700034, 1131300214, 1132800134, 1201300494, 1204000054, 1210700044, 1131300224, 1132800144, 1201300544, 1207400154, 1210700054, 1131300234, 1132800154, 1201300754, 1207400164, 1213800114, 1131300244, 1132800164, 1201300764, 1207400174, 1213800124, 1131900014, 1132800174, 1201300774, 1207400184, 1213800134, 1131900024, 1132800184, 1201300784, 1208300014, 1213800144, 1131900034, 1132800194, 1201300794, 1208300024, 1213800154, 1131900044, 1132800204, 1203200214, 1208300034, 1131900054, 1200500114, 1203200224, 1208300044, 1131900064, 1200500124, 1203200234, 1208300054."
    }
  ]
}