  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Costa Mesa",
      "address_1": "1620 Sunflower Ave",
      "reason_for_recall": "Medtronic is issuing a voluntary field corrective action for all its Newport\" HT70 and Newport\" HT70 Plus ventilators because they may shutdown spontaneously during normal operation without an accompanying alarm.",
      "address_2": "",
      "product_quantity": "12,966",
      "code_info": "All Newport Medical HT70 and HT70- Plus Ventilators",
      "center_classification_date": "20170501",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide) Internationally to   Canada, Australia, Brazil, Japan, Korea, Mexico, Bahrain, Egypt, Greece  Hungary, Ireland, Israel, Jordan, Kazakhstan, Lithuania, Mauritius  Namibia, Norway,  Pakistan, Poland, Qatar, Romania,  Russian Federation  South Africa, Spain, Turkey, United Arab Emirates, United Kingdom  Algeria, Armenia, Austria, Bahrain, Belarus, Belgium, Botswana  Bulgaria, Cyprus,  Czech Republic, Denmark, Egypt, Finland, France  Germany, Greece, Hungary, Iran, Ireland, Israel, Italy, Kenya, Lebanon  Lithuania, Macedonia, Malawi, Malta, Mauritius,  Mozambique, Netherlands  Norway, Oman, Poland, Portugal,  Qatar, Republic of Moldova, Romania  Saudi Arabia, Serbia, Slovakia, Sudan, Sweden,  Switzerland, Tunisia  Turkey, Uganda, Ukraine,  United Kingdom, Yemen, and Zimbabwe",
      "state": "CA",
      "product_description": "Newport Medical HT70 and HT70- Plus Ventilators, Rx Only  Product Usage:    The Newport HT70 family of ventilators is intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces. Specifically, the Newport HT70 family of ventilators is applicable for infant, pediatric and adult patients greater than or equal to 5 kg (11 lbs). The Newport HT70 is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. The Newport HT70 is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.",
      "report_date": "20170510",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Newport Medical Instruments Inc",
      "recall_number": "Z-1874-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "76891",
      "termination_date": "20200718",
      "more_code_info": "",
      "recall_initiation_date": "20170330",
      "postal_code": "92626-1513",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}