  "results": [
    {
      "status": "Ongoing",
      "city": "Morrisville",
      "state": "NC",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92373",
      "recalling_firm": "TELEFLEX LLC",
      "address_1": "3015 Carrington Mill Blvd",
      "address_2": "N/A",
      "postal_code": "27560-5437",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including Puerto Rico.",
      "recall_number": "Z-1870-2023",
      "product_description": "Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100181030",
      "product_quantity": "2874 units",
      "reason_for_recall": "Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.",
      "recall_initiation_date": "20230525",
      "center_classification_date": "20230629",
      "report_date": "20230705",
      "code_info": "UDI/DI 14026704340528, Batch Numbers:  18GG23, 18HG17, 18HG21, 18IG17, 18LG18, 19AG25, 19BG19, 19BG32, 19GT16, 19IT53, 19LT21, KME22F2174, KME22G1753",
      "more_code_info": ""
    }
  ]
}