{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94508",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada,  China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan,  Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia,  Spain, Sweden, Switzerland, Thailand, United Kingdom.",
      "recall_number": "Z-1855-2024",
      "product_description": "Panorama HFO- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device.  Model Number (REF):  781350",
      "product_quantity": "118 units",
      "reason_for_recall": "The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's  technical room, resulting in Smoke and/or fire",
      "recall_initiation_date": "20240503",
      "center_classification_date": "20240517",
      "report_date": "20240529",
      "code_info": "UDI-DI:  (1) 782152 N/A (2) 782110 (01)00884838099364(21)  Serial Numbers:  Serial Number 37019 37067 37149 37188 37184 37034 37036 37058 37167 37173 37183 37195 37197 37009 37011 37013 37020 37027 37031 37050 37055 37080 37093 37100 37121 37127 37132 37137 37141 37143 37150 37157 37170 37180 37193 37194 37023 37044 37066 37071 37106 37124 37160 37175 37182 37192 37022 37068 37082 37129 37134 37145 37156 37171 37181 37187 19083 37161 37178 37088 37024 37098 37135 37139 37176 37153 37060 37116 37196 37015 37073 37109 37122 37133 37162 37185 37026 37096 37177 37002 37012 37017 37029 37039 37041 37047 37048 37052 37053 37056 37077 37083 37084 37086 37087 37095 37102 37104 37107 37110 37113 37120 37123 37125 37128 37130 37136 37142 37158 37159 37163 37165 37168 37189 37069 37174 37126",
      "more_code_info": ""
    }
  ]
}