  "results": [
    {
      "status": "Terminated",
      "city": "Munich",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87984",
      "recalling_firm": "Brainlab AG",
      "address_1": "Olof-palme-str. 9",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AZ, AR, CA, FL, IN, NJ, NY, NC, OR, PA, and TX.  There was no government/military distribution.",
      "recall_number": "Z-1846-2021",
      "product_description": "ExacTrac Dynamic software, Model 20910-01B ETD Positioning and Monitoring Installer 1.0.2 - Product Usage: intended to position patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, to monitor the patient position, and to provide a beam hold signal in case of a deviations in order to treat lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.",
      "product_quantity": "22 systems in the U.S.",
      "reason_for_recall": "Display of potential patient movement might be delayed to the user for high dose treatments.",
      "recall_initiation_date": "20210514",
      "center_classification_date": "20210610",
      "termination_date": "20230320",
      "report_date": "20210616",
      "code_info": "Version 1.02, GTIN 04056481142315"
    }
  ]
}