  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "94508",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada,  China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan,  Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia,  Spain, Sweden, Switzerland, Thailand, United Kingdom.",
      "recall_number": "Z-1844-2024",
      "product_description": "Evolution upgrade 3.0T   Model Number (REF):   (1)782117;  (2) 782143",
      "product_quantity": "",
      "reason_for_recall": "The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's  technical room, resulting in Smoke and/or fire",
      "recall_initiation_date": "20240503",
      "center_classification_date": "20240517",
      "report_date": "20240529",
      "code_info": "UDI-DI: N/A  Serial Numbers:  42246 85713 42029 71774 85668 85680 85685 85694 85705 42023 42024 42375 42217 42402 71044 85698 85700 85146 85677 85699 71595 71815 85649 42320 85684 71836 85679 85682 85697 85701 85708 85714",
      "more_code_info": ""
    }
  ]
}