  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Trumbull",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82821",
      "recalling_firm": "CooperSurgical, Inc.",
      "address_1": "95 Corporate Dr",
      "address_2": "N/A",
      "postal_code": "06611-1350",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution in the states of CA, CT, IN, MA, MI, MN, NC, NV, NY, OH, TX, WA and WI.",
      "recall_number": "Z-1840-2019",
      "product_description": "LEEP Precision Integrated System 120V, Model LP-10-120    Product Usage:  The LEEP PRECISION INTEGRATED SYSTEM l 20V combines three components - LEEP PRECISION Cart, LEEP PRECISION Generator, and the LEEP PRECISION Smoke Evacuator for loop electrosurgical excision procedures. The loop electrosurgical excision procedure (LEEP) utilizes a very thin wire in the shape of a loop heated by electrical current generated by a power supply to efficiently remove abnormal cells. The LEEP PRECISION  INTEGRATED SYSTEM has been designed to provide an efficient workstation that can efficiently meet all the unique LEEP requirements of the clinic and practice-based facilities .",
      "product_quantity": "19",
      "reason_for_recall": "2.5A slow blow fuses may have been installed in the LEEP RPECISION Generator rather than the appropriate 3.15A slow blow fuses.",
      "recall_initiation_date": "20190423",
      "center_classification_date": "20190618",
      "termination_date": "20200424",
      "report_date": "20190626",
      "code_info": "2018110001,2018110002, 2018110003, 2018110004, 2018110005,2018110006, 2018110007, 2018110008,2018110011, 2018110012,2018110013, 2018110014,  2018110015,2018110016, 2018110017,  2018110019, 2018110020,2018110021,&  2018110022"
    }
  ]
}