  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "North Haven",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82950",
      "recalling_firm": "Covidien LLC",
      "address_1": "60 Middletown Ave",
      "address_2": "N/A",
      "postal_code": "06473-3908",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide domestic distribution, worldwide foreign distribution.",
      "recall_number": "Z-1833-2019",
      "product_description": "SIGNIA TRI-STAPLE 2.0 BLACK 45MM EXTRA THICK RELOA, SIG45AXT",
      "product_quantity": "3,113,280",
      "reason_for_recall": "The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the potential for infection and/or sepsis.",
      "recall_initiation_date": "20190517",
      "center_classification_date": "20190615",
      "report_date": "20190626",
      "code_info": "N6E0176UX  N6E0278UX  N6E0584UX  N6E0587UX  N6E0589UX  N6E0590UX  N6E0975UX  N6F0141UX  N6F0142UX  N6F0143UX  N6F0144UX  N6F0340UX  N6F0341UX  N6F0342UX  N6F0343UX  N6F0436UX  N6G0033UX  N6G0034UX  N6G0035UX  N6G0036UX  N6G0037UX  N6G0038UX  N6G0125UX  N6G0744UX  N6H0757UX  N6J0436UX  N6K0010UX  N6K0372UX  N6K0415UX  N6K0990UX  N6K0991UX  N6K1110UX  N7C0111X  N7C0381X  N7C0499X  N7C0611X  N7D0303X"
    }
  ]
}