{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Richmond",
      "state": "VA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96752",
      "recalling_firm": "Mckesson Medical-Surgical Inc. Corporate Office",
      "address_1": "9954 Maryland Drive",
      "address_2": "Deep Run Iii Ste. 4000",
      "postal_code": "23233",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AL, AR, IL, KS, MT, NC, NJ, OR, SC, TN & WV.",
      "recall_number": "Z-1832-2025",
      "product_description": "ORTHO CLINICAL DIAGNOSTICS INC VITROS PHOSPHORUS (60/CR 5CR/PK) ORCLNL REAGENT.",
      "product_quantity": "10 units",
      "reason_for_recall": "transit delays of certain cold chain products due to severe thunderstorms which may caused the products to be stored outside the recommended labeled storage conditions. This failure could increase the probability of an incorrect test result or a delayed test result.",
      "recall_initiation_date": "20250424",
      "center_classification_date": "20250527",
      "report_date": "20250604",
      "code_info": "Catalog No. 1513209; GTIN: 10758750009220; Order No. 46094482."
    }
  ]
}