      "status": "Terminated",
      "city": "Saint Charles",
      "state": "MO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85352",
      "recalling_firm": "Elekta Inc",
      "address_1": "1450 Beale St Ste 205",
      "address_2": "N/A",
      "postal_code": "63303",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US nationwide distributions in the states of MA, NH, NC, UT, OH, FL, MN, NY, LA and MT.",
      "recall_number": "Z-1826-2020",
      "product_description": "Leksell Vantage Stereotactic System. Instrument Carrier, Part of Leksell Vantage Arc (1053958). Neurological stereotaxic Instrument",
      "product_quantity": "25 units.",
      "reason_for_recall": "A faulty locking piece of the Instrument Carrier that does not fulfil the requirement of locking force to the Arc has been found during an internal check of a Leksell Vantage Stereotactic Arc System.",
      "recall_initiation_date": "20200330",
      "center_classification_date": "20200429",
      "termination_date": "20240410",
      "report_date": "20200506",
      "code_info": "Serial Numbers:SH00106, SH00120, SH00149, SH00160, SH00165, SH00110, SH00143, SH00150, SH00161, SH00166, SH00111, SH00144, SH00157, SH00162, SH00167, SH00112, SH00147, SH00158, SH00163, SH00168, SH00114, SH00148, SH00159, SH00164 and SH00178. UDI 7340048304887."
    }
  ]
}